Specific Process Knowledge/Lithography/Resist/UVresist

From LabAdviser

UV resist comparison table

Comparison of specifications and feature space of UV photoresists.

Resist AZ 5214E AZ MiR 701 AZ nLOF 2020 AZ 4562 SU-8 TI Spray
Resist tone
  • Positive
  • Negative (image reversal)
Positive Negative Positive Negative
  • Positive
  • Negative (image reversal)
Thickness range 1.5 - 4.2 µm 1.5 - 4 µm 1.5 - 4 µm 5 - 10 µm 1 - 200 µm 0.5 - 5 µm
Coating tool

Automatic coaters:

  • Spin coater: Gamma UV lithography
  • Spin coater: Gamma ebeam & UV

Manual coaters:

  • Spin coater: Labspin 2
  • Spin coater: Labspin 3
  • Spin coater: RCD8
  • Spray coater

Automatic coaters:

  • Spin coater: Gamma UV lithography
  • Spin coater: Gamma ebeam & UV

Manual coaters:

  • Spin coater: Labspin 2
  • Spin coater: Labspin 3
  • Spin coater: RCD8
  • Spray coater

Automatic coaters:

  • Spin coater: Gamma UV lithography

Manual coaters:

  • Spin coater: Labspin 2
  • Spin coater: Labspin 3
  • Spin coater: RCD8
  • Spray coater

Automatic coaters:

  • Spin coater: Gamma ebeam & UV

Manual coaters:

  • Spin coater: Labspin 2
  • Spin coater: Labspin 3
  • Spin coater: RCD8

Manual coaters:

  • Spin coater: RCD8

Spray coater

Spectral sensitivity 310 - 420 nm 310 - 445 nm 310 - 380 nm 310 - 445 nm 300 - 375 nm 310 - 440 nm
Exposure tool Mask aligner or Maskless aligner
Developer
  • AZ 351B
  • AZ 726 MIF
  • AZ 351B
  • AZ 726 MIF
AZ 726 MIF AZ 726 MIF mr-DEV 600 (PGMEA) AZ 726 MIF
Rinse DIW DIW DIW DIW IPA DIW
Remover
  • Acetone
  • Remover 1165
  • Acetone
  • Remover 1165
  • Acetone
  • Remover 1165

Remover 1165

Cured SU-8 is practically insoluble

Plasma ashing can remove crosslinked SU-8

  • Acetone
  • Remover 1165
Comments
  • Can be used for both positive and image reversed (negative) processes with resist thickness between 1 and 4 µm.
  • Good adhesion for wet etch.

High selectivity for dry etch

Negative sidewalls for lift-off

For processes with resist thickness between 6 µm and 25 µm

  • High aspect ratio
  • Resist thickness 1 µm to several 100 µm

Spray coater specific resist

Process flow examples

Mask aligner: Process flow AZ 5214E positive‎ Process flow AZ 5214E image reversal

Maskless aligner: Process flow AZ 5214E positive‎ Process flow AZ 5214E image reversal

Mask aligner: Process_Flow_AZ_MiR701.docx‎

Maskless aligner: Process_Flow_AZ_MiR701_MLA.docx‎

Mask aligner: Process_Flow_AZ_nLOF_2020.docx‎

Maskless aligner: Process_Flow_AZ_nLOF_2020_MLA.docx‎

Mask aligner: Process_Flow_thick_AZ4562.docx‎

Maskless aligner: Process_Flow_thick_AZ4562_MLA.docx‎

Process_Flow_SU8_70um.docx‎

Most of the process knowledge on SU-8 is based in research groups


Other process flows:

Chip on carrier: A procedure for UV lithography on a chip using automatic coater and developer.

Exposure dose

Spectral sensitivity of AZ resists represented as optical absorption coefficient.

During exposure of the resist, the photoinitiator, or photo-active component, reacts with the exposure light, and starts the reaction which makes the resist develop in the developer.

In a positive resist, it makes the resist become soluble in the developer. In a negative resist, usually assisted by thermal energy in the post-exposure bake (PEB), it makes the resist insoluble in the developer. The amount of light required to fully develop the resist in the development process, is the exposure dose.

The optimal exposure dose is a function of many parameters, including the type of resist, the resist thickness, and the sensitivity of the resist.

Resist sensitivity
The resist sensitivity is a measure of how efficiently it reacts to the exposure light. Spectral sensitivity is the sensitivity of the resist as a function of wavelength. It is usually given simply as the range from the wavelength below which absorption in the resist material makes lithography impractical to the wavelength at which the photoinitiator is no longer efficiently activated.

Within the sensitivity range, the optical absorption is commonly used as a measure of sensitivity. A high absorption coefficient signifies a high sensitivity, as the light is absorbed by the photoinitiator. Because of spectral sensitivity, the optimal dose of a given resist type and thickness is also a function of the spectral distribution of the exposure light, i.e. the equipment used for the exposure. Using a combination of experience, calculation and assumptions, it may be possible to estimate the dose for an exposure equipment, if the exposure dose is already known on another equipment.

Due to reflection and refraction at the interface between the resist and the substrate, the optimal dose may also be a function of the type of substrate used. Unless otherwise stated, the exposure doses given below are for standard silicon wafers.

Apart from the already mentioned factors, the optimal dose also depends on the developer chemistry and the parameters used in the development process. Finally, the requirements to the lithographic process in terms of resolution, bias (line broadening), etch selectivity, side wall angle, etc. may narrow down, or widen, the process window. The exposure doses given in the sections below should be used as a starting point for individual fabrication process development.

Due to the process of bleaching, where the absorption of the resist changes during exposure, the exposure dose is unfortunately not always constant at different intensities of the exposure light. The exposure time is thus not always a linear function of the exposure intensity.

Calculate exposure time

In the maskless aligners, the dose is set directly as a process parameter in the job. In mask aligners, on the other hand, the parameter that is set is the exposure time, i.e. how long the shutter is open during the exposure.

The exposure dose, D [J/m2], in terms if exposure light intensity I [W/m2] and exposure time t [s], is given by:

Failed to parse (SVG (MathML can be enabled via browser plugin): Invalid response ("Math extension cannot connect to Restbase.") from server "https://wikimedia.org/api/rest_v1/":): {\displaystyle D=I \sdot t}

Since the intensity is specific to the spectral sensitivity of the sensor used to measure the exposure light, and the exposure time is specific to the spectral distribution of the exposure light (cf. spectral sensitivity), this dose is specific to the combination of exposure source and optical sensor.

Given an exposure dose, the exposure time, t, is calculated as:

Failed to parse (SVG (MathML can be enabled via browser plugin): Invalid response ("Math extension cannot connect to Restbase.") from server "https://wikimedia.org/api/rest_v1/":): {\displaystyle t = D \sdot I^{-1}}

It is important to keep in mind that this exposure time is valid only for a specific combination of exposure source and optical sensor, as well as for a specific development process.

Exposure dose for mask aligners

The exposure doses listed below are for generic good exposure results, and can be a compromise between getting good lines, as well as good dots, in both clear field and dark field exposures. The optimal dose for any given specific project, could be different from the listed values.

All doses are for standard silicon wafers, unless otherwise stated. Development is done using 2.38% TMAH.


KS Aligner

The KS Aligner has an i-line notch filter installed. This results in an exposure light peak around 365 nm with a FWHM of 7 nm.

Date Thickness Dose Development Comments
AZ 5214E
Old German version
Long ago 1.5 µm 72 mJ/cm2 Single puddle, 60 s Positive process
Long ago 2.2 µm 80 mJ/cm2
Long ago 4.2 µm 160 mJ/cm2
AZ 4562
Old German version
Long ago 10 µm 510 mJ/cm2 Multiple puddle, 4 x 60 s Multiple exposure with 10-15 s pauses is recommended.
AZ MiR 701
Old PFOA containing version
Long ago 1 µm 180 mJ/cm2 Single puddle, 60 s PEB: 60 s at 110°C
Long ago 2 µm 200 mJ/cm2 Single puddle, 60 s
Long ago 4 µm 400 mJ/cm2 Single puddle, 60 s PEB: 90 s at 110°C
AZ nLOF 2020 Long ago 2 µm 110 mJ/cm2 Single puddle, 60 s PEB: 60 s at 110°C

Side wall angle ~15°

For smaller angle (~5°), develop 30 seconds instead

Aligner: MA6 - 2

The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak around 365 nm with a FWHM of 7 nm.

Date Thickness Dose Development Comments
AZ 5214E
Old German version
Long ago 1.5 µm 72 mJ/cm2 Single puddle, 60 s Positive process
Long ago 2.2 µm 90 mJ/cm2 Single puddle, 60 s
AZ 5214E Image Reversal
Old German version
Long ago 1.5 µm 22 mJ/cm2 Single puddle, 60 s Image reversal process.
Reversal bake: 120 s at 110°C.
Flood exposure: 200 mJ/cm2
Long ago 2.2 µm 25 mJ/cm2 Single puddle, 60 s
AZ MiR 701
Old PFOA containing version
Long ago 1.5 µm 169 mJ/cm2 Single puddle, 60 s PEB: 60 s at 110°C
Long ago 2 µm 200 mJ/cm2 Single puddle, 60 s
Long ago 4 µm 280 mJ/cm2 Single puddle, 60 s PEB: 60 s at 110°C
Process adopted from process logs
AZ nLOF 2020 Long ago 1.5 µm 104 mJ/cm2 Single puddle, 30 s PEB: 60 s at 110°C
Use 60 s development for lift-off
AZ 5214E
New Japanese version
2023-01-11
jehem
1.5 µm 70 mJ/cm2 Single puddle, 60 s
AZ 5214E Image Reversal
New Japanese version
2023-01-11
jehem
2.2 µm 22 mJ/cm2 Single puddle, 60 s Image reversal process.
Reversal bake: 60 s at 110°C.
Flood exposure: 500 mJ/cm2
AZ 4562
New Japanese version
2021-12-08
jehem
10 µm 550 mJ/cm2 Multiple puddles, 5 x 60 s Priming: HMDS
Rehydration after SB: 1 hour (may not be necessary)
Exposure: Multiple exposures with pauses, 5 x (10 s exposure + 10 s pause)
Degassing after exposure: 1 hour (may not be necessary)
AZ MiR 701
New PFOA free version
2021-06-23
elkh
1.5 µm 150 mJ/cm2 Single puddle, 60 s PEB: 60 s at 110°C


Exposure dose for maskless aligners

The exposure doses listed below are for generic good exposure results, and can be a compromise between getting good lines, as well as good dots, in both clear field and dark field exposures. The optimal dose for any given specific project, could be different from the listed values.

All doses are for standard silicon wafers, unless otherwise stated. Development is done using 2.38% TMAH.


Aligner: Maskless 01

The Aligner: Maskless 01 has a 365 nm LED light source with a FWHM of 8 nm.

Date Thickness Exposure mode Dose Defoc Resolution Comments
AZ 5214E
Old German version
2021-08-19 1.5 µm Fast 80 mJ/cm2 -4 2 µm Dev: SP60s
2021-08-19 1.5 µm Quality 80 mJ/cm2 -4 1.25 µm Dev: SP60s
AZ 5214E image reversal
Old German version
2020-03-01 2.2 µm Fast 43 mJ/cm2 0 >2 µm (a lot of stitching) Reversal bake: 120s@110°C, Flood exposure: 200mJ/cm2, Dev: SP60s
2021-08-25 2.2 µm Quality 26 mJ/cm2 -4 1.75 µm Reversal bake: 120s@110°C, Flood exposure: 200mJ/cm2, Dev: SP60s
AZ MiR 701
Old PFOA containing version
2021-08-25 1.5 µm Fast 225 mJ/cm2 -4 1.25 µm
Tested using dehydration reducing measures
PEB: 60s@110°C, Dev: SP60s
AZ nLOF 2020 2020-02-01 2.0 µm Fast 300 mJ/cm2 0 2 µm (due to stitching) PEB: 60s@110°C, Dev: SP60s
2020-03-01 2.0 µm Quality 180-200 mJ/cm2 0 1 µm PEB: 60s@110°C, Dev: SP60s
AZ 5214E
New Japanese version
2023-01-06 1.5 µm Fast 80 mJ/cm2 -2 1.75 µm Dev: SP60s
AZ 5214E image reversal
New Japanese version
2023-01-11 2.2 µm Fast 42 mJ/cm2 -2 1.75 µm Reversal bake: 60s@110°C, Flood exposure: 500mJ/cm2, Dev: SP60s
AZ MiR 701
New PFOA free version
2021-08-25 1.5 µm Fast 225 mJ/cm2 -4 1.25 µm (due to stitching)
Tested using dehydration reducing measures
PEB: 60s@110°C, Dev: SP60s
AZ 4562
New Japanese version
2021-12-07 10 µm Fast 750 mJ/cm2 0 ≤5 µm Priming: HMDS
Rehydration after SB: 1 hour (may not be necessary)
Exposure: Multiple exposures with pauses, 5 x (10 s exposure + 10 s pause)
Degassing after exposure: 1 hour (may not be necessary)
Development: Multiple puddles, 5 x 60 s


Dehydration reducing measures used for testing AZ MiR 701:
The CDA used for the pneumatic autofocus will dehydrate the resist. To reduce this effect, the writehead is parked far away from the write area while setting up the job for at least a few minutes, before starting the exposure.


Aligner: Maskless 02

The Aligner: Maskless 02 has a 375 nm laser light source, and a 405 nm laser light source, each with a FWHM of ~1 nm.

Date Thickness Laser Autofocus Exposure mode Dose Defoc Resolution Comments
AZ 5206E
(AZ5214E diluted with PGMEA 1:1 per volume)
Old German version
Long ago 0.5 µm 375 nm Optical Fast 60 mJ/cm2 -6 1 µm (not optimized) Dev: 2xSP30s
Long ago Optical Quality 60 mJ/cm2 -6 ~750 nm (not optimized) Dev: 2xSP30s
AZ 5214E
Old German version
Long ago 1.5 µm 405 nm Optical Fast 90 mJ/cm2 -2 1-2 µm Dev: SP60s
Long ago 375 nm Optical Fast 65 mJ/cm2 2 ~1 µm Dev: SP60s
Long ago Optical Quality 65 mJ/cm2 2 ~750 nm Dev: SP60s
AZ MiR 701
Old PFOA containing version
Long ago 1.5 µm 405 nm Optical Fast 200 mJ/cm2 -5 ~1 µm (not optimized) PEB: 60s@110°C, Dev: SP60s
Long ago 375 nm Optical Fast 170 mJ/cm2 -5 1 µm PEB: 60s@110°C, Dev: SP60s
Long ago Optical Quality 180 mJ/cm2 -6 (Feb 2019)
-2 (Apr 2019)
<750 nm PEB: 60s@110°C, Dev: SP60s
Large structures probably over-exposed
AZ nLOF 2020 2021-02-22
jehem
2 µm 375 nm Optical Fast 350 mJ/cm2 0 1 µm PEB: 120s@110°C, Dev: SP60s
2021-02-22
jehem
Optical Quality 350 mJ/cm2 0 1 µm PEB: 120s@110°C, Dev: SP60s
AZ 5214E
New Japanese version
2021-08-20
jehem
1.5 µm 375 Optical Fast 70 mJ/cm2 -4 1 µm Dev: SP60s
AZ 5214E image reversal
New Japanese version
2021-11-25
jehem
2.2 µm 375 Optical Fast 16 mJ/cm2 -4 1.75 µm Reversal bake: 60s@110°C, Flood exposure: 500mJ/cm2, Dev: SP60s
AZ MiR 701
New PFOA free version
2021-08-20
jehem
1.5 µm 375 Optical Fast 200 mJ/cm2 -6 1 µm
Tested using dehydration reducing measures
PEB: 60s@110°C, Dev: SP60s
AZ 4562
New Japanese version
2021-12-08
jehem
10 µm 375 Optical Fast 750 mJ/cm2 0 ≤5 µm Priming: HMDS
Rehydration after SB: 1 hour (may not be necessary)
Exposure: Multiple exposures with pauses, 5 x (10 s exposure + 10 s pause)
Degassing after exposure: 1 hour (may not be necessary)
Development: Multiple puddles, 5 x 60 s


Dehydration reducing measures used for testing AZ MiR 701:
The CDA used for the pneumatic autofocus will dehydrate the resist. To reduce this effect, the writehead is parked far away from the write area while setting up the job for at least a few minutes, before starting the exposure.


Aligner: Maskless 03

The Aligner: Maskless 03 has a 405 nm laser light source with a FWHM of ~1 nm.

Date Thickness Laser Autofocus Exposure mode Dose Defoc Resolution Comments
AZ 5214E
Old German version
2020-10-01 1.5 µm 405 nm Pneumatic Fast 70 mJ/cm2 0 2 µm (due to stitching) Dev: SP60s
2021-03-23 405 nm Pneumatic Quality 65 mJ/cm2 -2 1 µm Dev: SP60s
AZ MiR 701
Old PFOA containing version
2020-02-01 1.5 µm 405 nm Pneumatic Fast 300 mJ/cm2 0 2 µm (due to stitching)
Non optimized dehydration reduction
PEB: 60s@110°C, Dev: SP60s
2021-08-25 405 nm Pneumatic Quality 175 mJ/cm2 -2 1 µm
Tested using dehydration reducing measures
PEB: 60s@110°C, Dev: SP60s
AZ 5214E image reversal
Old German version
2020-10-01 2.2 µm 405 nm Pneumatic Fast 43 mJ/cm2 0 >2 µm (a lot of stitching) Reversal bake: 120s@110°C, Flood exposure: 200mJ/cm2, Dev: SP60s
2021-08-25 405 nm Pneumatic Quality 32 mJ/cm2 -1 1 µm Reversal bake: 120s@110°C, Flood exposure: 200mJ/cm2, Dev: SP60s
AZ nLOF 2020 2023-01-13 2.0 µm 405 Pneumatic Quality 9000 mJ/cm2 0 1 µm PEB: 60s @ 100°C,
Dev: SP60s
Due to the high dose required on this tool, it is recommended to process nLOF on tools with a more appropriate exposure light source
AZ 5214E
New Japanese version
2023-01-06 1.5 µm 405 Pneumatic Quality 70 mJ/cm2 -2 1 µm Dev: SP60s
AZ 5214E image reversal
New Japanese version
2023-01-06 2.2 µm 405 Pneumatic Quality 50 mJ/cm2 -2 1 µm Reversal bake: 60s@110°C, Flood exposure: 500mJ/cm2, Dev: SP60s
AZ MiR 701
New PFOA free version
2021-08-25 1.5 µm 405 Pneumatic Quality 175 mJ/cm2 -2 1 µm
Tested using dehydration reducing measures
PEB: 60s@110°C, Dev: SP60s
AZ 4562
New Japanese version
2021-12-08 10 µm 405 Pneumatic Quality 550 mJ/cm2 1 ≤5 µm Priming: HMDS
Rehydration after SB: 1 hour (may not be necessary)
Exposure: Multiple exposures with pauses, 5 x (10 s exposure + 10 s pause)
Degassing after exposure: 1 hour (may not be necessary)
Development: Multiple puddles, 5 x 60 s


Dehydration reducing measures used for testing AZ MiR 701:
The CDA used for the pneumatic autofocus will dehydrate the resist. To reduce this effect, the writehead is parked far away from the write area while setting up the job for at least a few minutes, before starting the exposure.


Exposure dose when using AZ 351B developer (NaOH)

The exposure doses listed below are for generic good exposure results, and can be a compromise between getting good lines, as well as good dots, in both clear field and dark field exposures. The optimal dose for any given specific project, could be different from the listed values.

All doses are for standard silicon wafers, unless otherwise stated. Development is done using AZ 351B developer (NaOH).

KS Aligner (351B)

The KS Aligner has an i-line notch filter installed. This results in an exposure light peak around 365 nm with a FWHM of 7 nm.

Date Thickness Dose Development] Comments
AZ 5214E Long ago 1.5 µm 70mJ/cm2 60 s Positive process
Long ago 2.2 µm 72 mJ/cm2 70 s
Long ago 4.2 µm 160 mJ/cm2 180 s
AZ 5214E Long ago 1.5 µm 30 mJ/cm2 60 s Image reversal process.

Reversal bake: 100s at 110°C.
Flood exposure after reversal bake: 210 mJ/cm2

Long ago 2.2 µm 35 mJ/cm2 70 s
AZ 4562 Long ago 10 µm 320 mJ/cm2 5 minutes Multiple exposure with 10-15 s pauses is recommended.