Specific Process Knowledge/Characterization/AFM: Atomic Force Microscopy: Difference between revisions

Jesyup (talk | contribs)
No edit summary
Jesyup (talk | contribs)
No edit summary
Line 203: Line 203:




{| class="wikitable" style="max-width:1100px"
{| class="wikitable" style="max-width:800px"
|+ Comparison between images of the Tip Checker sample
|+ Comparison between images of the Tip Checker sample
|-
|-
! !! SEM Image !! Worn Probe !! New probe
! !! Worn Probe !! New probe
|-
|-
! Image
! Image
| [[File:AFMWornTip.jpg|thumb|left|alt=SEM image of Tip checker sample| SEM image of the tip checker sample (placeholder)]]
| [[File:AFMWornTip.jpg|thumb|left|alt=Image taken using a worn probe| AFM image of the tip checker sample taken using a wornAFM probe]]
| [[File:AFMWornTip.jpg|thumb|left|alt=Image taken using a worn probe| AFM image of the tip checker sample taken using a wornAFM probe]]
| [[File:AFMNewTip.jpg|thumb|left|alt=Image taken using a new probe| AFM image of the tip checker sample taken using a new AFM probe]]
| [[File:AFMNewTip.jpg|thumb|left|alt=Image taken using a new probe| AFM image of the tip checker sample taken using a new AFM probe]]
Line 215: Line 214:
! Image Parameters and
! Image Parameters and
Roughness information
Roughness information
|
* Acceleration voltage: 3kV
* Detector: SE
|  
|  
* Size: 1µm  
* Size: 1µm  
Line 232: Line 228:
|-
|-
! Comments  
! Comments  
| Example
| The shape of the probe causes all features to look like an oval pointing towards the top left of the image. Low measured roughness as the worn tip skates across the top of the triangles.
| The shape of the probe causes all features to look like an oval pointing towards the top left of the image. Low measured roughness as the worn tip skates across the top of the triangles.
| Same sample taken with a new probe. The triangular structures are pointing in different directions, as expected based on SEM image. Higher measured roughness as the sharper tip can map the trenches between the triangles.
| Same sample taken with a new probe. The triangular structures are pointing in different directions, as expected based on SEM image. Higher measured roughness as the sharper tip can map the trenches between the triangles.