Jump to content

Specific Process Knowledge/Lithography/UVExposure: Difference between revisions

From LabAdviser
Taran (talk | contribs)
Taran (talk | contribs)
 
(93 intermediate revisions by 3 users not shown)
Line 1: Line 1:
{{:Specific Process Knowledge/Lithography/authors_generic}}
{{cc-nanolab}}


'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/UVExposure click here]'''
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/UVExposure click here]'''
Line 14: Line 14:


!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#KS Aligner|KS Aligner]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: MA6-1|Aligner: MA6-1]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_MA6_-_2|Aligner: MA6 - 2]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_MA6-2|Aligner: MA6-2]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 01|Aligner: Maskless 01]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 01|Aligner: Maskless 01]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 02|Aligner: Maskless 02]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 02|Aligner: Maskless 02]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 03|Aligner: Maskless 03]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 03|Aligner: Maskless 03]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Inclined UV Lamp|Inclined UV Lamp]]</b>
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 04|Aligner: Maskless 04]]</b>
<!--|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Inclined UV Lamp|Inclined UV Lamp]]</b>-->


|-
|-
Line 28: Line 29:
*Back Side Alignment
*Back Side Alignment
*UV exposure
*UV exposure
OBS: this tool is in PolyFabLab
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Top Side Alignment
*Top Side Alignment
Line 46: Line 48:
*Maskless UV exposure
*Maskless UV exposure
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Top Side Alignment
*Maskless UV exposure
*Direct laser writing
OBS: this tool is in PolyFabLab
<!--|style="background:WhiteSmoke; color:black"|
*UV exposure
*UV exposure
*DUV exposure
*DUV exposure-->


|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Performance
!style="background:silver; color:black" align="center" valign="center" rowspan="4"|Performance
|style="background:LightGrey; color:black"|Minimum feature size
|style="background:LightGrey; color:black"|Minimum feature size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
Line 63: Line 70:
~1 µm
~1 µm
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
~1 µm
<!--|style="background:WhiteSmoke; color:black"|-->
|-
|-
|style="background:LightGrey; color:black"|Exposure light/filters/spectrum
|style="background:LightGrey; color:black"|Alignment accuracy
|style="background:WhiteSmoke; color:black"|
*TSA: ±2 µm
*BSA: ±5 µm
|style="background:WhiteSmoke; color:black"|
*TSA: ±1 µm
*BSA: ±2 µm
|style="background:WhiteSmoke; color:black"|
±2 µm<br>(±1 µm possible)
|style="background:WhiteSmoke; color:black"|
*TSA: ± 0.5 µm
*BSA: ± 1 µm
|style="background:WhiteSmoke; color:black"|
*TSA: ± 0.5 µm
*BSA: ± 1 µm
|style="background:WhiteSmoke; color:black"|
±1 µm
<!--|style="background:WhiteSmoke; color:black"|-->
 
|-
|style="background:LightGrey; color:black"|Exposure light
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*350W Hg lamp
*350W Hg lamp
*i-line filter (365nm notch filter)<br>Intensity in Constant Intensity mode:<br>8mW/cm<sup>2</sup> @ 365nm
*i-line filter (365nm bandpass filter)
*303nm filter optional
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*500W Hg-Xe lamp
*500W Hg-Xe lamp
*i-line filter (365nm notch filter)<br>Intensity in Constant Intensity mode:<br>11mW/cm<sup>2</sup> @ 365nm
*i-line filter (365nm bandpass filter)
*UV350 optics optional
*UV250 optics optional
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
365nm LED
365nm LED
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
375nm laserdiodes
375nm laser diode array
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
405nm laserdiodes
405nm laser diode array
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*365nm LED
*405nm laser diode
<!--|style="background:WhiteSmoke; color:black"|
*1000 W Hg-Xe lamp
*1000 W Hg-Xe lamp
*Dichroic mirror:
*Dichroic mirror:
**Near UV (350-450nm)
**Near UV (350-450nm)
**Mid UV (260-320nm)
**Mid UV (260-320nm)
**Deep UV (220-260nm)
**Deep UV (220-260nm)-->


|-
|-
Line 114: Line 144:
**Pneumatic auto-focus
**Pneumatic auto-focus
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Projection:
**Optical auto-focus
**Pneumatic auto-focus
*Direct laser writing:
**Optical auto-focus
**Pneumatic auto-focus
<!--|style="background:WhiteSmoke; color:black"|
*Flood exposure
*Flood exposure
*Proximity exposure with home-made chuck and maskholder
*Proximity exposure with home-made chuck and maskholder
*Inclined exposure
*Inclined exposure
*Rotating exposure
*Rotating exposure-->


|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
|style="background:LightGrey; color:black"|Batch size
|style="background:LightGrey; color:black"|Batch size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1 small sample, down to 10x10 mm<sup>2</sup>
*1 50 mm wafer
*1 100 mm wafer
*1 100 mm wafer
*1 150 mm wafer
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1 small sample, down to 10x10 mm<sup>2</sup>
*1 small sample, down to 10x10 mm<sup>2</sup>
Line 149: Line 183:
*1 150 mm wafer
*1 150 mm wafer
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*all sizes up to 8inch
*1 small sample, down to 3x3 mm<sup>2</sup>
*1 50 mm wafer
*1 100 mm wafer
*1 150 mm wafer
<!--|style="background:WhiteSmoke; color:black"|
*all sizes up to 8inch-->
 
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*All cleanroom materials except copper and steel
*All PolyFabLab materials
*Dedicated 2inch chuck for III-V materials
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*All cleanroom materials except copper and steel
*All cleanroom materials except copper and steel
Line 165: Line 204:
*All cleanroom materials
*All cleanroom materials
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*All cleanroom materials
*All PolyFabLab materials
<!--|style="background:WhiteSmoke; color:black"|
*All cleanroom materials-->
|-  
|-  
|}
|}


==KS Aligner==
==Aligner: MA6-1==
[[Image:KSAligner in E-4.jpg|300x300px|thumb|The KS Aligner is placed in E-4]]
[[Image:KSAligner in E-4.jpg|400px|thumb|The Aligner: MA6-1 is located in PolyFabLab.]]


SUSS Mask Aligner MA6 is designed for high resolution photolithography.  
SUSS Mask Aligner MA6 is designed for high resolution photolithography.  
Line 177: Line 218:




'''Training videos:'''
'''Training videos:'''  
<br/>The training videos were made for MA6-2, but operation of MA6-1 is very similar as they are exactly the same model. Due to a different power supply, however, procedures for ignition and intensity control of the lamp is different for MA6-1. Please refer to the manual for the correct procedures.


[https://www.youtube.com/watch?v=o8IBtfQHNzU Operation]
[https://www.youtube.com/watch?v=o8IBtfQHNzU Operation]
Line 183: Line 225:
[https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment]
[https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment]


 
The user manual(s), quality control procedure(s) and results, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=44 LabManager]  - '''requires login'''
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=44 LabManager]  - '''requires login'''


===Exposure dose===
===Exposure dose===
[[Specific Process Knowledge/Lithography/Resist#KS_Aligner|Information on UV exposure dose]]
[[Specific Process Knowledge/Lithography/Resist#Aligner: MA6-1|Information on UV exposure dose]]


===Process information===
===Process information===
The KS Aligner has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.
The Aligner: MA6-1 has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.
 
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design and locations]].


<!--
===Quality Control (QC)===
===Quality Control (QC)===
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
Line 197: Line 241:
|-
|-
|
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for KS Aligner] - '''requires login'''<br>
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for Aligner: MA6-1] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=44 The newest QC data for KS Aligner] - '''requires login'''
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=44 The newest QC data for Aligner: MA6-1] - '''requires login'''
{| {{table}}
{| {{table}}
| align="center" |  
| align="center" |  
Line 207: Line 251:
|-  
|-  
|Measurement area
|Measurement area
|4"
|100 mm
|-  
|-  
|Number of measurements
|Number of measurements
Line 216: Line 260:
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!QC limits
!KS Aligner
!Aligner: MA6-1
|-
|-
|Nominal intensity
|Nominal intensity
Line 232: Line 276:
Power supply and/or lamp will be adjusted if intensity is outside the limit.
Power supply and/or lamp will be adjusted if intensity is outside the limit.
|}
|}
-->


===Equipment performance and process related parameters===
===Equipment performance and process related parameters===
Line 252: Line 297:
| style="background:LightGrey; color:black"|Exposure light/filters
| style="background:LightGrey; color:black"|Exposure light/filters
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*broadband (i-, g-, h-line)
*365 nm (i-line)
*365 nm (i-line)
*303 nm
*broadband (i-, g-, h-line), requires tool change
*(303 nm, requires tool change)
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
Line 263: Line 308:
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*5x5 inch
*5x5 inch
*7x7 inch
*special holder for 4 x 2" designs on 5x5 inch
|-
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side (TSA)
*Top side (TSA), ±2µm
*Backside (BSA)
*Backside (BSA), ±5µm
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
* small pieces 1x1cm
Mask exposure and alignment:
* 50 mm wafers
* 100 mm wafers
* 100 mm wafers
* 150 mm wafers
Flood exposure:
* samples up to 150 mm wafers
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All PolyFabLab materials
 
Dedicated chuck for III-V materials
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
Line 293: Line 334:
<br clear="all" />
<br clear="all" />


==Aligner: MA6 - 2==
==Aligner: MA6-2==
[[Image:AlignerMA6-2 in E-4.jpg|300x300px|thumb|The Aligner: MA6 - 2 is placed in E-4]]
[[Image:AlignerMA6-2 in E-4.jpg|400px|thumb|The Aligner: MA6-2 is located in E-4.]]


The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography.  
The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography.  
Line 313: Line 354:


===Exposure dose===
===Exposure dose===
[[Specific Process Knowledge/Lithography/Resist#Aligner:_MA6_-_2|Information on UV exposure dose]]
[[Specific Process Knowledge/Lithography/Resist#Aligner:_MA6-2|Information on UV exposure dose]]


===Process information===
===Process information===
The Aligner: MA6 - 2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.
The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.
 
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design and locations]].


===Quality Control (QC)===
===Quality Control (QC)===
Line 333: Line 376:
|-  
|-  
|Measurement area
|Measurement area
|4"
|100 mm
|-  
|-  
|Number of measurements
|Number of measurements
Line 365: Line 408:


*Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
*Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
**Alignment of smaller separations is possible using the "scan microscope" function
*Maximum distance between TSA microscope objectives: 160 mm
*Maximum distance between TSA microscope objectives: 160 mm
*TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)
*TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)
Line 375: Line 419:
**2": X +/-  8-22mm; Y +/- 0-6mm
**2": X +/-  8-22mm; Y +/- 0-6mm
**4": X +/- 14-46mm; Y +/- 0-10mm
**4": X +/- 14-46mm; Y +/- 0-10mm
**6": X +/- 14-69mm; Y +/- 0-10mm
**6": X +/- 14-69mm; Y +/- 0-10mm (OBS: Max. separation of BSA microscopes is 100mm)




Line 429: Line 473:
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side (TSA)
*Top side (TSA), ±1µm (machine spec: ±2µm)
*Backside (BSA)
*Backside (BSA), ±2µm (machine spec: ±5µm)
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
Line 453: Line 497:


<sup>1)</sup> Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.
<sup>1)</sup> Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.
===Light intensity and uniformity after lamp ignition===
[[image:MA6-2_light_intensity_and_uniformity_from_ignition.png|400px|right|thumb|Light intensity and uniformity after lamp ignition. CP is constant power mode, CI is constant intensity mode. The accept limits for the CI mode is ±5%, the accept limit for the non-uniformity is 2%. CP mode has no accept limit.]]
The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.


<br clear="all" />
<br clear="all" />
Line 458: Line 512:
== Aligner: Maskless 01 ==
== Aligner: Maskless 01 ==


[[Image:Heidelberg_MLA100.jpg|300x300px|thumb|Aligner: Maskless 01 positioned in E-5 (2017)]]
[[Image:Heidelberg_MLA100.jpg|400px|thumb|Aligner: Maskless 01 is located in E-4.]]


The logon password for the PC is "mla" (without quotation marks).
The logon password for the PC is "mla" (without quotation marks).
Line 466: Line 520:
The system offers top side alignment with high accuracy.
The system offers top side alignment with high accuracy.


Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]].


The user manual and contact information can be found in LabManager:
The user manual and contact information can be found in LabManager:
Line 480: Line 535:
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_01_processing#Alignment|Alignment]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_01_processing#Alignment|Alignment]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_01_processing#Optimal_use_of_the_maskless_aligner|Optimal use of the maskless aligner]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_01_processing#Optimal_use_of_the_maskless_aligner|Optimal use of the maskless aligner]]
===Quality Control (QC)===
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 01 (MLA1) - Dose and Defoc'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=422 The QC procedure for Aligner: Maskless 01 (MLA1)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=422 The newest QC data for Aligner: Maskless 01 (MLA1)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center"
! QC:
! Dose and defocus test on 1.5µm AZ5214E
! QC limits
|-
|Dose test
|
Step size is 5 mJ/cm<sup>2</sup><br>
Test range is last QC value ± 10 mJ/cm<sup>2</sup> (5 steps total)
|none
|-
|Defoc test
|
Step size is 1 defoc<br>
Test range is last QC value ± 4 (5 steps total)
|none
|-
|}
|}
|}
<br/>
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 01 (MLA1) - Alignment'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=422 The QC procedure for Aligner: Maskless 01 (MLA1)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=422 The newest QC data for Aligner: Maskless 01 (MLA1)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center"
! QC Recipe:
! Alignment accuracy test
! QC limits
|-
|Topside alignment
|
Expose an overlay design after automatic alignment to 4 alignment marks using the High Res camera and applying scaling and shearing.<br>
Alignment accuracy in 9 points across a 100mm wafer is measured, the average alignment error is reported.
| Must be better than 1µm
|-
|}
| align="center" valign="top"|
|-
|}
Camera offsets will be adjusted if alignment error is outside the limit.
|}
<br/>


=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===
Line 519: Line 635:
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Top side only
Top side only, ±2µm (±1µm can be achieved)


|-
|-
Line 534: Line 650:
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
<br>Total height variation across the substrate must be less than ±40 µm - including wafer bow
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
Line 544: Line 661:


==Aligner: Maskless 02==
==Aligner: Maskless 02==
[[File:Aligner MLA 2.jpg|400px|thumb|Aligner: Maskless 02 is located in E-5.]]
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023).
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023).


Line 556: Line 674:
*Separate conversion PC
*Separate conversion PC


Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]].


'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''
Line 569: Line 688:
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_02_processing#Features|Features]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_02_processing#Features|Features]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_02_processing#Alignment|Alignment]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_02_processing#Alignment|Alignment]]
===Quality Control (QC)===
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 02 (MLA2) - Dose and Defoc'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=440 The QC procedure for Aligner: Maskless 02 (MLA2)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=440 The newest QC data for Aligner: Maskless 02 (MLA2)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Dose and defocus test on 1.5µm AZ5214E
|-
|Dose test
|Last QC value ± 20 mJ/cm<sup>2</sup> (9 steps total)
|-
|Defoc test
|Last QC value ± 8 (9 steps total)
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!Aligner: Maskless 02 (MLA2)
|-
|Dose
|none
|-
|Defoc
|none
|-
|}
|-
|}
Dose and defoc values are reported in the QC data sheet.
|}
<br/>
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 02 (MLA2) - Alignment'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=440 The QC procedure for Aligner: Maskless 02 (MLA2)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=440 The newest QC data for Aligner: Maskless 02 (MLA2)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Alignment accuracy test
|-
|Topside alignment
|Expose an overlay design after automatic alignment to 4 alignment marks using the High Res camera and applying scaling and shearing.
Alignment accuracy in 9 points across a 100mm wafer is measured, the average alignment error is reported.
|-
|Backside alignment
|Expose an overlay design after automatic alignment to 4 backside alignment marks and applying scaling and shearing. Rotate the wafer 180° and repeat.
Alignment accuracy in 9 points across a 100mm wafer is measured, half the average alignment error is reported.
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!Aligner: Maskless 02 (MLA2)
|-
|Topside alignment error
|>0.5µm
|-
|Backside alignment error
|>1µm
|-
|}
|-
|}
Camera offsets will be adjusted if alignment error is outside the limit.
|}
<br/>


=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===
Line 611: Line 810:
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side alignment
*Top side alignment, ±0.5µm
*Backside alignment
*Backside alignment, ±1.0µm
*Field alignment (chip-by-chip TSA)
*Field alignment (chip-by-chip TSA), ±0.25µm (within 5x5mm<sup>2</sup> area)


|-
|-
Line 629: Line 828:
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
<br>Total height variation across the substrate must be less than ±90 µm - including wafer bow
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
Line 641: Line 841:


==Aligner: Maskless 03==
==Aligner: Maskless 03==
[[File:Aligner MLA 3.jpg|400px|thumb|Aligner: Maskless 03 is located in E-5.]]
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020.
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020.


Line 649: Line 850:
*Separate conversion PC (Power PC)
*Separate conversion PC (Power PC)


Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]].


'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]'''
Line 662: Line 864:
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_03_processing#Features|Features]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_03_processing#Features|Features]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_03_processing#Alignment|Alignment]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_03_processing#Alignment|Alignment]]
===Quality Control (QC)===
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 03 (MLA3) - Dose and Defoc'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=464 The QC procedure for Aligner: Maskless 03 (MLA3)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=464 The newest QC data for Aligner: Maskless 03 (MLA3)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Dose and defocus test on 1.5µm AZ5214E
|-
|Dose test
|Last QC value ± 20 mJ/cm<sup>2</sup> (9 steps total)
|-
|Defoc test
|Last QC value ± 8 (9 steps total)
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!Aligner: Maskless 03 (MLA3)
|-
|Dose
|none
|-
|Defoc
|none
|-
|}
|-
|}
Dose and defoc values are reported in the QC data sheet.
|}
<br/>
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 03 (MLA3) - Alignment'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=464 The QC procedure for Aligner: Maskless 03 (MLA3)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=464 The newest QC data for Aligner: Maskless 03 (MLA3)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Alignment accuracy test
|-
|Topside alignment
|Expose an overlay design after automatic alignment to 4 alignment marks using the High Res camera and applying scaling and shearing.
Alignment accuracy in 9 points across a 100mm wafer is measured, the average alignment error is reported.
|-
|Backside alignment
|Expose an overlay design after automatic alignment to 4 backside alignment marks and applying scaling and shearing. Rotate the wafer 180° and repeat.
Alignment accuracy in 9 points across a 100mm wafer is measured, half the average alignment error is reported.
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!Aligner: Maskless 03 (MLA3)
|-
|Topside alignment error
|>0.5µm
|-
|Backside alignment error
|>1µm
|-
|}
|-
|}
Camera offsets will be adjusted if alignment error is outside the limit.
|}
<br/>


=== Equipment performance and process related parameters ===
=== Equipment performance and process related parameters ===
Line 702: Line 984:
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
*Top side alignment
*Top side alignment, ±0.5µm
*Backside alignment
*Backside alignment, ±1.0µm
*Field alignment (chip-by-chip TSA)
*Field alignment (chip-by-chip TSA), ±0.25µm (within 5x5mm<sup>2</sup> area)


|-
|-
Line 719: Line 1,001:
|style="background:WhiteSmoke; color:black" colspan="2"|
|style="background:WhiteSmoke; color:black" colspan="2"|
All cleanroom materials
All cleanroom materials
<br>Total height variation across the substrate must be less than ±90 µm - including wafer bow
|-
|style="background:LightGrey; color:black"|Batch
|style="background:WhiteSmoke; color:black" colspan="2"|
1
|-
|}
<br clear="all" />
== Aligner: Maskless 04 ==
[[Image:Heidelberg_uMLA.JPG|400px|thumb|Aligner: Maskless 04 is located in PolyFabLab in building 347.]]
The logon password for the PC is "mla" (without quotation marks).
The µMLA Tabletop Maskless Aligner from Heidelberg, located in PolyFabLab in building 347, is a direct exposure lithography tool installed in 2024.
It is a UV exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of a mask.
The system offers top side alignment with good accuracy.
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]].
The user manual and contact information can be found in LabManager:
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=544 LabManager] - '''requires login'''
<!--
===Exposure dose and defocus===
[[Specific Process Knowledge/Lithography/Resist#Aligner:_Maskless_04|Information on UV exposure dose]]
-->
===[[Specific Process Knowledge/Lithography/Aligners/Aligner: Maskless 04 processing|Process information]]===
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Exposure_technology|Exposure technology]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Process_Parameters|Process parameters]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Substrate_positioning|Substrate positioning]]
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Alignment|Alignment]]
=== Equipment performance and process related parameters ===
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;"
!style="background:silver; color:black;" align="center" width="60"|Purpose
|style="background:LightGrey; color:black"|
|style="background:WhiteSmoke; color:black" colspan="2"|
Alignment and UV exposure
|-
!style="background:silver; color:black" align="left" valign="center" rowspan="6"|Performance
|style="background:LightGrey; color:black"|Exposure mode
|style="background:WhiteSmoke; color:black" colspan="2"|
*Projection (Raster mode)
*Direct laser writing (Vector mode)
|-
| style="background:LightGrey; color:black"|Exposure light
|style="background:WhiteSmoke; color:black" colspan="2"|
*365nm (LED) for projection
*405nm (diode laser) for direct laser writing
|-
|style="background:LightGrey; color:black"|Focusing method
|style="background:WhiteSmoke; color:black" colspan="2"|
Pneumatic or Optical
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:WhiteSmoke; color:black" colspan="2"|
Down to 1µm
|-
|style="background:LightGrey; color:black"|Design formats
|style="background:WhiteSmoke; color:black" colspan="2"|
*GDS-II
*CIF
*DXF
*Gerber
*HIMT format
|-
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" colspan="2"|
Top side only, ±1µm
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black" colspan="2"|
* maximum writing area: 150x150 mm<sup>2</sup>
* 150 mm wafer
* 100 mm wafer
* 50 mm wafer
* pieces down to 3x3 mm<sup>2</sup> with optical autofocus
|-
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black" colspan="2"|
All PolyFabLab materials with sufficient stiffness and flatness.
<br>Total height variation across the substrate must be less than ±80 µm - including wafer bow
|-
|-
|style="background:LightGrey; color:black"|Batch
|style="background:LightGrey; color:black"|Batch
Line 728: Line 1,101:
<br clear="all" />
<br clear="all" />


<!--
==Inclined UV Lamp==
==Inclined UV Lamp==
[[Image:Inclined UV lamp_1.jpg|300x300px|right|thumb|Inclined UV lamp is placed in CX-1]]
[[Image:Inclined UV lamp_1.jpg|300x300px|right|thumb|Inclined UV lamp is placed in CX-1]]
Line 797: Line 1,171:


<br clear="all" />
<br clear="all" />
-->

Latest revision as of 14:17, 19 December 2025

The content on this page, including all images and pictures, was created by DTU Nanolab staff, unless otherwise stated.

Feedback to this page: click here

UV Exposure Comparison Table

Equipment Aligner: MA6-1 Aligner: MA6-2 Aligner: Maskless 01 Aligner: Maskless 02 Aligner: Maskless 03 Aligner: Maskless 04
Purpose
  • Top Side Alignment
  • Back Side Alignment
  • UV exposure

OBS: this tool is in PolyFabLab

  • Top Side Alignment
  • Back Side Alignment
  • UV exposure
  • (DUV exposure)
  • Bond alignment
  • Top Side Alignment
  • Maskless UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • Maskless UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • Maskless UV exposure
  • Top Side Alignment
  • Maskless UV exposure
  • Direct laser writing

OBS: this tool is in PolyFabLab

Performance Minimum feature size

~1 µm

~1 µm

~1 µm

~1 µm

~1 µm

~1 µm

Alignment accuracy
  • TSA: ±2 µm
  • BSA: ±5 µm
  • TSA: ±1 µm
  • BSA: ±2 µm

±2 µm
(±1 µm possible)

  • TSA: ± 0.5 µm
  • BSA: ± 1 µm
  • TSA: ± 0.5 µm
  • BSA: ± 1 µm

±1 µm

Exposure light
  • 350W Hg lamp
  • i-line filter (365nm bandpass filter)
  • 500W Hg-Xe lamp
  • i-line filter (365nm bandpass filter)

365nm LED

375nm laser diode array

405nm laser diode array

  • 365nm LED
  • 405nm laser diode
Exposure mode
  • Flood exposure
  • Proximity
  • Contact:
    • Soft contact
    • Hard contact
    • Vacuum contact
  • Flood exposure
  • Proximity
  • Contact:
    • Soft contact
    • Hard contact
    • Vacuum contact
  • Projection:
    • Pneumatic auto-focus
  • Projection:
    • Optical auto-focus
    • Pneumatic auto-focus
  • Projection:
    • Pneumatic auto-focus
  • Projection:
    • Optical auto-focus
    • Pneumatic auto-focus
  • Direct laser writing:
    • Optical auto-focus
    • Pneumatic auto-focus
Substrates Batch size
  • 1 100 mm wafer
  • 1 small sample, down to 10x10 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 5x5 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 3x3 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 200 mm wafer
  • 1 small sample, down to 5x5 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 3x3 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
Allowed materials
  • All PolyFabLab materials
  • All cleanroom materials except copper and steel
  • Dedicated chuck for III-V materials
  • All cleanroom materials
  • All cleanroom materials
  • All cleanroom materials
  • All PolyFabLab materials

Aligner: MA6-1

The Aligner: MA6-1 is located in PolyFabLab.

SUSS Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.


Training videos:
The training videos were made for MA6-2, but operation of MA6-1 is very similar as they are exactly the same model. Due to a different power supply, however, procedures for ignition and intensity control of the lamp is different for MA6-1. Please refer to the manual for the correct procedures.

Operation

Alignment

The user manual(s), quality control procedure(s) and results, and contact information can be found in LabManager - requires login

Exposure dose

Information on UV exposure dose

Process information

The Aligner: MA6-1 has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.

Link to information about alignment mark design and locations.


Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

vacuum contact, hard contact, soft contact, proximity, flood exposure

Exposure light/filters
  • 365 nm (i-line)
  • broadband (i-, g-, h-line), requires tool change
  • (303 nm, requires tool change)
Minimum structure size

down to 1.25µm

Mask size
  • 5x5 inch
Alignment modes
  • Top side (TSA), ±2µm
  • Backside (BSA), ±5µm
Substrates Substrate size

Mask exposure and alignment:

  • 100 mm wafers

Flood exposure:

  • samples up to 150 mm wafers
Allowed materials

All PolyFabLab materials

Batch

1


Aligner: MA6-2

The Aligner: MA6-2 is located in E-4.

The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, low vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA (video) microscope.

The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.


Training videos:

Operation

Alignment


The user manual, quality control procedures and results, user APVs, and contact information can be found in LabManager - requires login

Exposure dose

Information on UV exposure dose

Process information

The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.

Link to information about alignment mark design and locations.

Quality Control (QC)

Quality Control (QC) for Aligner: MA6-2
QC Recipe: Manual intensity measurement
Measurement area 100 mm
Number of measurements 5
QC limits Aligner: MA6-2
Nominal intensity 11 mW/cm2 @ 365 nm
Intensity deviation from nominal ≤5%
Intensity non-uniformity ≤2%

Power supply and/or lamp will be adjusted if intensity is outside the limit.

Alignment

Top Side Alignment:

  • TSA microscope standard objectives: 5X, and 10X (20X available)
  • TSA microscope special objectives: 11.25X offset (for smaller separation)
  • Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
    • Alignment of smaller separations is possible using the "scan microscope" function
  • Maximum distance between TSA microscope objectives: 160 mm
  • TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)

BackSide Alignment:

  • Minimum distance between BSA microscope objectives: 15 mm
  • Maximum distance between BSA microscope objectives: 100 mm
  • BSA microscope travel range: X +50mm / -16mm; Y +50mm / -20mm (towards flat)
  • BSA chuck view ranges:
    • 2": X +/- 8-22mm; Y +/- 0-6mm
    • 4": X +/- 14-46mm; Y +/- 0-10mm
    • 6": X +/- 14-69mm; Y +/- 0-10mm (OBS: Max. separation of BSA microscopes is 100mm)


Microscope field of view (W x H, splitfield):

  • TSA 5X
    • Oculars: 1.3mm x 2.6mm (Ø2.6mm full field)
    • Camera: 350µm x 500µm (700µm x 500µm full field)
  • TSA 10X
    • Oculars: 0.6mm x 1.3mm (Ø1.3mm full field)
    • Camera: 150µm x 250µm (350µm x 250µm full field)
  • TSA special
    • Oculars: 0.55mm x 1.1mm (Ø1.1mm full field)
    • Camera: 150µm x 200µm (300µm x 200µm full field)
  • BSA camera
    • Low: 1.5mm x 2mm (3mm x 2mm full field)
    • High: 450µm x 650µm (950µm x 650µm full field)

Equipment performance and process related parameters

Purpose

Mask alignment and UV exposure, potentially DUV exposure 1)

Bond alignment

Performance Exposure mode

vacuum contact, hard contact, soft contact, proximity, flood exposure

Exposure light/filters
  • broadband (i-, g-, h-line)
  • 365 nm (i-line)
  • "UV300" (280-350 nm)
  • DUV (240 nm) 1)
Minimum structure size

Typically 1.25 µm, possibly down to 0.8 µm 1)

Mask size
  • 5x5 inch
  • 7x7 inch
  • special holder for 4 x 2" designs on 5x5 inch
Alignment modes
  • Top side (TSA), ±1µm (machine spec: ±2µm)
  • Backside (BSA), ±2µm (machine spec: ±5µm)
Substrates Substrate size
  • small pieces 1x1cm
  • 50 mm wafers
  • 100 mm wafers
  • 150 mm wafers
Allowed materials

All cleanroom materials except copper and steel

Dedicated chuck for III-V materials

Batch

1

1) Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.

Light intensity and uniformity after lamp ignition

Light intensity and uniformity after lamp ignition. CP is constant power mode, CI is constant intensity mode. The accept limits for the CI mode is ±5%, the accept limit for the non-uniformity is 2%. CP mode has no accept limit.

The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.

If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.

The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.


Aligner: Maskless 01

Aligner: Maskless 01 is located in E-4.

The logon password for the PC is "mla" (without quotation marks).

The MLA 100 Maskless Aligner located in the E-4 cleanroom is a direct exposure lithography tool installed in 2017. It is a UV LED exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of the mask. The system offers top side alignment with high accuracy.

Link to information about alignment mark design.

The user manual and contact information can be found in LabManager:

Equipment info in LabManager - requires login

Exposure dose and defocus

Information on UV exposure dose

Process information

Quality Control (QC)

Quality Control (QC) for Aligner: Maskless 01 (MLA1) - Dose and Defoc
QC: Dose and defocus test on 1.5µm AZ5214E QC limits
Dose test

Step size is 5 mJ/cm2
Test range is last QC value ± 10 mJ/cm2 (5 steps total)

none
Defoc test

Step size is 1 defoc
Test range is last QC value ± 4 (5 steps total)

none


Quality Control (QC) for Aligner: Maskless 01 (MLA1) - Alignment
QC Recipe: Alignment accuracy test QC limits
Topside alignment

Expose an overlay design after automatic alignment to 4 alignment marks using the High Res camera and applying scaling and shearing.
Alignment accuracy in 9 points across a 100mm wafer is measured, the average alignment error is reported.

Must be better than 1µm

Camera offsets will be adjusted if alignment error is outside the limit.


Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light

365nm (LED), FWHM=8nm

Focusing method

Pneumatic

Minimum structure size

down to 1µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes

Top side only, ±2µm (±1µm can be achieved)

Substrates Substrate size
  • maximum writing area: 125x125 mm2
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 5x5 mm2
Allowed materials

All cleanroom materials
Total height variation across the substrate must be less than ±40 µm - including wafer bow

Batch

1


Aligner: Maskless 02

Aligner: Maskless 02 is located in E-5.

MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023).


Special features

  • Optical Autofocus
  • Backside Alignment
  • Basic Gray Scale Exposure
  • Advanced Field Alignment Mode for alignment to individual chips/devices on the substrate
  • High Aspect Ratio Mode for exposure of thick resists
  • 200 x 200 mm exposure field
  • Separate conversion PC

Link to information about alignment mark design.

Training videos

Equipment info in LabManager - requires login

Exposure dose and defocus

Information on UV exposure dose

Process information

Quality Control (QC)

Quality Control (QC) for Aligner: Maskless 02 (MLA2) - Dose and Defoc
QC Recipe: Dose and defocus test on 1.5µm AZ5214E
Dose test Last QC value ± 20 mJ/cm2 (9 steps total)
Defoc test Last QC value ± 8 (9 steps total)
QC limits Aligner: Maskless 02 (MLA2)
Dose none
Defoc none

Dose and defoc values are reported in the QC data sheet.


Quality Control (QC) for Aligner: Maskless 02 (MLA2) - Alignment
QC Recipe: Alignment accuracy test
Topside alignment Expose an overlay design after automatic alignment to 4 alignment marks using the High Res camera and applying scaling and shearing.

Alignment accuracy in 9 points across a 100mm wafer is measured, the average alignment error is reported.

Backside alignment Expose an overlay design after automatic alignment to 4 backside alignment marks and applying scaling and shearing. Rotate the wafer 180° and repeat.

Alignment accuracy in 9 points across a 100mm wafer is measured, half the average alignment error is reported.

QC limits Aligner: Maskless 02 (MLA2)
Topside alignment error >0.5µm
Backside alignment error >1µm

Camera offsets will be adjusted if alignment error is outside the limit.


Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light
  • 375 nm

(laser diode arrays)

Focusing method
  • Optical
  • Pneumatic
Minimum structure size

down to 1 µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes
  • Top side alignment, ±0.5µm
  • Backside alignment, ±1.0µm
  • Field alignment (chip-by-chip TSA), ±0.25µm (within 5x5mm2 area)
Substrates Substrate size
  • maximum writing area: 200x200 mm2
  • 200 mm wafer
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 3x3 mm2 1)
Allowed materials

All cleanroom materials
Total height variation across the substrate must be less than ±90 µm - including wafer bow

Batch

1

1) with optical autofocus


Aligner: Maskless 03

Aligner: Maskless 03 is located in E-5.

MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020.


Special features:

  • Backside Alignment
  • Advanced Field Alignment Mode for alignment to individual chips/devices on the substrate
  • Separate conversion PC (Power PC)

Link to information about alignment mark design.

Training videos

Equipment info in LabManager - requires login

Exposure dose and defocus

Information on UV exposure dose

Process information

Quality Control (QC)

Quality Control (QC) for Aligner: Maskless 03 (MLA3) - Dose and Defoc
QC Recipe: Dose and defocus test on 1.5µm AZ5214E
Dose test Last QC value ± 20 mJ/cm2 (9 steps total)
Defoc test Last QC value ± 8 (9 steps total)
QC limits Aligner: Maskless 03 (MLA3)
Dose none
Defoc none

Dose and defoc values are reported in the QC data sheet.


Quality Control (QC) for Aligner: Maskless 03 (MLA3) - Alignment
QC Recipe: Alignment accuracy test
Topside alignment Expose an overlay design after automatic alignment to 4 alignment marks using the High Res camera and applying scaling and shearing.

Alignment accuracy in 9 points across a 100mm wafer is measured, the average alignment error is reported.

Backside alignment Expose an overlay design after automatic alignment to 4 backside alignment marks and applying scaling and shearing. Rotate the wafer 180° and repeat.

Alignment accuracy in 9 points across a 100mm wafer is measured, half the average alignment error is reported.

QC limits Aligner: Maskless 03 (MLA3)
Topside alignment error >0.5µm
Backside alignment error >1µm

Camera offsets will be adjusted if alignment error is outside the limit.


Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light

405nm
(laser diode array)

Focusing method

Pneumatic

Minimum structure size

down to 1 µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes
  • Top side alignment, ±0.5µm
  • Backside alignment, ±1.0µm
  • Field alignment (chip-by-chip TSA), ±0.25µm (within 5x5mm2 area)
Substrates Substrate size
  • maximum writing area: 150x150 mm2
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 5x5 mm2
Allowed materials

All cleanroom materials
Total height variation across the substrate must be less than ±90 µm - including wafer bow

Batch

1


Aligner: Maskless 04

Aligner: Maskless 04 is located in PolyFabLab in building 347.

The logon password for the PC is "mla" (without quotation marks).

The µMLA Tabletop Maskless Aligner from Heidelberg, located in PolyFabLab in building 347, is a direct exposure lithography tool installed in 2024. It is a UV exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of a mask. The system offers top side alignment with good accuracy.

Link to information about alignment mark design.

The user manual and contact information can be found in LabManager:

Equipment info in LabManager - requires login

Process information

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode
  • Projection (Raster mode)
  • Direct laser writing (Vector mode)
Exposure light
  • 365nm (LED) for projection
  • 405nm (diode laser) for direct laser writing
Focusing method

Pneumatic or Optical

Minimum structure size

Down to 1µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes

Top side only, ±1µm

Substrates Substrate size
  • maximum writing area: 150x150 mm2
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 3x3 mm2 with optical autofocus
Allowed materials

All PolyFabLab materials with sufficient stiffness and flatness.
Total height variation across the substrate must be less than ±80 µm - including wafer bow

Batch

1