Specific Process Knowledge/Characterization/XRD/XRD SmartLab: Difference between revisions

From LabAdviser
Reet (talk | contribs)
Created page with " == XRD SmartLab == The Rigaku SmartLab is an advanced XRD for measuring on thin films. All thin films can be measured without fixating the sample, as the system has a so cal..."
 
Eves (talk | contribs)
 
(46 intermediate revisions by 3 users not shown)
Line 1: Line 1:
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Characterization/XRD/XRD_SmartLab click here]'''


== XRD SmartLab ==
<i> Unless otherwise stated, this page is written by <b>DTU Nanolab internal</b><br>
All images and photos on this page belongs to <b>DTU Nanolab</b></i>
 
 
 
= XRD SmartLab =


The Rigaku SmartLab is an advanced XRD for measuring on thin films. All thin films can be measured without fixating the sample, as the system has a so called In-Plane arm.
The Rigaku SmartLab is an advanced XRD for measuring on thin films. All thin films can be measured without fixating the sample, as the system has a so called In-Plane arm.


[[image:XRD_SmartLab.jpg|200x200px|right|thumb|The XRD SmartLab located in cleanroom F-2]]
[[image:XRD_SmartLab.jpg|400x400px|right|thumb|The XRD SmartLab located in cleanroom F-2]]




'''The user manual(s), user APV(s), technical information, and contact information can be found in LabManager:'''  
'''The user manual(s), user APV(s), technical information, and contact information can be found in LabManager:'''  


[http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=428 XRD SmartLab in LabManager]
[http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=428 XRD SmartLab in LabManager] - requires login


Various measurement types including X-Ray reflectivity, Rocking curve, Theta-2theta, and Pole figures are described here:


[http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Characterization/XRD/Process_Info  Process information]
[http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Characterization/XRD/Process_Info  Process information]
The measurement settings and results of the pre-acceptance test are described in this document, also found on the Lab Manager:
*[https://labmanager.dtu.dk/view_binary.php?fileId=5413| (FAT) Factory Acceptance Test]  - requires login
The same measurement were completed as part of the acceptance, but no report was generated. In addition the inspection certificate is available here:
*[https://labmanager.dtu.dk/view_binary.php?fileId=5414| (SAT) Site Acceptance Test]  - requires login


==Software for analysis==
==Software for analysis==
The software packages used for data analysis are available on the equipment computer, but we recommend that you install it on your personal computer. To run the software you need a USB dongle with a license on, these can be borrowed from Rebecca and Kristian in room 347-077. We only have 9 dongles available, so when you are done please return the dongle to Nanolab.  
Please see the Data Analysis section at the main [[Specific Process Knowledge/Characterization/XRD|XRD page]].
 
==Characterization of thin films using X-ray reflectivity (XRR)==
 
Implementing the XRR method the user can calculate thickness, density, and roughness of the deposited material. The XRR is also widely employed for multilayer stack analysis.
 
Here are some results available:
 
* Characterization of [[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Aluminium/Al_Ebeam_evaporation_in_Temescal|e-beam deposited Al thin films]].
 
* Characterization of [[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Aluminium/Al_Sputtering_in_Cluster_Lesker_PC3#X-ray_reflectivity_analysis|DC sputtered Al thin films]].
 
* Characterization of [[Specific_Process_Knowledge/Characterization/XRD/XRD_SmartLab/ALD_deposited_alumina_and_titania_XRR#X-ray_reflectivity_results|ALD deposited Al<sub>2</sub>O<sub>3</sub> and TiO<sub>2</sub> ultrathin layers]].


The software can be found on "CleanroomDrive\_Equipment\XRD\Rigaku software\RILauncher", it should be possible to install the software without a dongle. To use the software you have to log in. The user is: Administrator. There is no password.
* Characterization of [[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_CrSi#X-ray_Reflectivity_Analysis|Sputtered deposited CrSi bilayer]].
 
* Characterization of [[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_MgO#X-Ray_Reflectivity|Sputtered deposited MgO thin films]].
 
* Characterization of [[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Scandium/Sc_Sputtering_in_Cluster_Lesker_PC3#X-ray_Reflectivity_(XRR)|Sputtered deposited Sc thin films]].
 
* Characterization of [[Specific_Process_Knowledge/Thin_film_deposition/Deposition_of_Scandium_Nitride/ScN_Reactive_Sputtering_in_Cluster_Lesker_PC3#X-ray_Reflectivity_(XRR)|Sputtered deposited ScN thin films]].
 
<!-- * Characterization of [[Specific_Process_Knowledge/Characterization/XRD/XRD_SmartLab/ALD_deposited_alumina_and_titania_XRR_and_SE_comparison|ALD deposited Al<sub>2</sub>O<sub>3</sub> and TiO<sub>2</sub> ultrathin layers]].-->
 
==Grazing Incidence X-ray Diffraction XRD (GiXRD)==
 
* Evaluation of [[Specific Process Knowledge/Characterization/XRD/XRD SmartLab/Instrumental broading in GiXRD|Instrumental broading in GiXRD]].
 
 
==θ/2θ X-ray Diffraction==
 
* Evaluation of [[Specific Process Knowledge/Characterization/XRD/XRD SmartLab/Instrumental broading in T2T|Instrumental broadening in symmetrical θ/2θ measurements]].


==Equipment performance and process related parameters==
==Equipment performance and process related parameters==
Line 25: Line 72:
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment  
|style="background:WhiteSmoke; color:black"|<b>XRD SmartLab</b>
|style="background:WhiteSmoke; color:black"|<b>XRD SmartLab</b>
|style="background:WhiteSmoke; color:black"|<b>XRD Powder</b>
|-
|-
!style="background:silver; color:black;" align="center" width="60"|Purpose  
!style="background:silver; color:black;" align="center" width="60"|Purpose  
Line 42: Line 88:
*Roughness
*Roughness
*Density
*Density
|style="background:WhiteSmoke; color:black"|
 
*Phase ID
*Crystal Size
*Crystallinity
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="6"|X-ray generator
!style="background:silver; color:black" align="center" valign="center" rowspan="6"|X-ray generator
Line 52: Line 95:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
3 kW
3 kW
|style="background:WhiteSmoke; color:black"|
600 W
|-
|-
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Line 59: Line 100:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
20 to 45 kV
20 to 45 kV
|style="background:WhiteSmoke; color:black"|
40 kV
|-
|-
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Line 66: Line 105:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
2 to 60 mA
2 to 60 mA
|style="background:WhiteSmoke; color:black"|
15 mA
|-
|-
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Type
Type
|style="background:WhiteSmoke; color:black"|
Sealed tube
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
Sealed tube
Sealed tube
Line 78: Line 113:
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Target
Target
|style="background:WhiteSmoke; color:black"|
Cu
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
Cu
Cu
Line 87: Line 120:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
0.4 mm x 8 mm (Line/Point)
0.4 mm x 8 mm (Line/Point)
|style="background:WhiteSmoke; color:black"|
0.4 mm x 12 mm (Line)
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="5"|Goniometer
!style="background:silver; color:black" align="center" valign="center" rowspan="4"|Goniometer
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Scanning mode
Scanning mode
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
incident / receiver coupled or independent
incident / receiver coupled or independent
|style="background:WhiteSmoke; color:black"|
incident / receiver coupled
|-
|-
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Line 102: Line 131:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
300 mm
300 mm
|style="background:WhiteSmoke; color:black"|
145 mm
|-
|-
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Line 109: Line 136:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
0.0001° (0.36")
0.0001° (0.36")
|style="background:WhiteSmoke; color:black"|
0.001° (3.6")
|-
|-
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|
Line 120: Line 145:
*X,Y:&plusmn;50 mm for a 100 mm wafer
*X,Y:&plusmn;50 mm for a 100 mm wafer
*Rx,Ry:-5~+5°
*Rx,Ry:-5~+5°
|style="background:WhiteSmoke; color:black"|
Fixed with rotation
|-
|style="background:LightGrey; color:black"|
Sample size
|style="background:WhiteSmoke; color:black"|
Diameter: 150 mm
Thickness: 0~21 mm
|style="background:WhiteSmoke; color:black"|
Powders
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Optics
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Optics
|style="background:LightGrey; color:black"|Incident side
|style="background:LightGrey; color:black"|Incident side
Line 139: Line 155:
*Soller slit
*Soller slit
*Variable divergence slit
*Variable divergence slit
|style="background:WhiteSmoke; color:black"|
*0.04° soller slit
*Ni and Cu filter
*Divergence slits
*Beam mask
|-
|-
|style="background:LightGrey; color:black"|Receiver side
|style="background:LightGrey; color:black"|Receiver side
Line 151: Line 162:
*Parallel slit analysers (PSA)
*Parallel slit analysers (PSA)
*Ge(220)x2 analyser
*Ge(220)x2 analyser
|style="background:WhiteSmoke; color:black"|
*0.04° soller slit
*Ni filter
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates
|style="background:LightGrey; color:black"|Substrate size
|style="background:LightGrey; color:black"|Substrate size
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
up to 150 mm wafers
up to 150 mm wafers; thickness up to 21 mm
|style="background:WhiteSmoke; color:black"|
Only for powders
|-
|-
| style="background:LightGrey; color:black"|Allowed materials
| style="background:LightGrey; color:black"|Allowed materials
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
All materials
All materials allowed in the cleanroom
|style="background:WhiteSmoke; color:black"|
All materials have to be approved
|-  
|-  
|}
|}


<br clear="all" />
<br clear="all" />

Latest revision as of 08:39, 31 July 2024

Feedback to this page: click here

Unless otherwise stated, this page is written by DTU Nanolab internal
All images and photos on this page belongs to DTU Nanolab


XRD SmartLab

The Rigaku SmartLab is an advanced XRD for measuring on thin films. All thin films can be measured without fixating the sample, as the system has a so called In-Plane arm.

The XRD SmartLab located in cleanroom F-2


The user manual(s), user APV(s), technical information, and contact information can be found in LabManager:

XRD SmartLab in LabManager - requires login


Various measurement types including X-Ray reflectivity, Rocking curve, Theta-2theta, and Pole figures are described here:

Process information


The measurement settings and results of the pre-acceptance test are described in this document, also found on the Lab Manager:

The same measurement were completed as part of the acceptance, but no report was generated. In addition the inspection certificate is available here:


Software for analysis

Please see the Data Analysis section at the main XRD page.

Characterization of thin films using X-ray reflectivity (XRR)

Implementing the XRR method the user can calculate thickness, density, and roughness of the deposited material. The XRR is also widely employed for multilayer stack analysis.

Here are some results available:


Grazing Incidence X-ray Diffraction XRD (GiXRD)


θ/2θ X-ray Diffraction

Equipment performance and process related parameters

Equipment XRD SmartLab
Purpose Crystal structure analysis and thin film thickness measurement
  • Phase ID
  • Crystal Size
  • Crystallinity
  • Quality and degree of orientation
  • 3D orientation
  • Latice strain
  • Composition
  • Twist
  • 3D lattice constant
  • Thickness
  • Roughness
  • Density
X-ray generator

Maximum rated output

3 kW

Rated tube voltage

20 to 45 kV

Rated tube current

2 to 60 mA

Type

Sealed tube

Target

Cu

Focus size

0.4 mm x 8 mm (Line/Point)

Goniometer

Scanning mode

incident / receiver coupled or independent

Goniomenter radius

300 mm

Minimum step size

0.0001° (0.36")

Sample stage

  • χ:-5~+95°
  • φ:0~360°
  • Z:-4~+1 mm
  • X,Y:±50 mm for a 100 mm wafer
  • Rx,Ry:-5~+5°
Optics Incident side
  • Cross Beam Optics(CBO)
  • Ge(220)x2 monochromator
  • In-Plane Parallel Slit Collimator (PSC)
  • Soller slit
  • Variable divergence slit
Receiver side
  • Automatic variable scattering slit
  • Automatic variable receiver slit
  • Parallel slit analysers (PSA)
  • Ge(220)x2 analyser
Substrates Substrate size

up to 150 mm wafers; thickness up to 21 mm

Allowed materials

All materials allowed in the cleanroom