Specific Process Knowledge/Lithography/UVExposure: Difference between revisions
No edit summary |
|||
| (68 intermediate revisions by 2 users not shown) | |||
| Line 1: | Line 1: | ||
{{ | {{cc-nanolab}} | ||
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/ | '''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php?title=Specific_Process_Knowledge/Lithography/UVExposure click here]''' | ||
[[Category: Equipment|Lithography exposure]] | [[Category: Equipment|Lithography exposure]] | ||
| Line 11: | Line 11: | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment | !colspan="2" border="none" style="background:silver; color:black;" align="center"|Equipment | ||
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure# | |style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: MA6-1|Aligner: MA6-1]]</b> | ||
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: | |style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner:_MA6-2|Aligner: MA6-2]]</b> | ||
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 01|Aligner: Maskless 01]]</b> | |style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 01|Aligner: Maskless 01]]</b> | ||
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 02|Aligner: Maskless 02]]</b> | |style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 02|Aligner: Maskless 02]]</b> | ||
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 03|Aligner: Maskless 03]]</b> | |style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 03|Aligner: Maskless 03]]</b> | ||
|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Inclined UV Lamp|Inclined UV Lamp]]</b> | |style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Aligner: Maskless 04|Aligner: Maskless 04]]</b> | ||
<!--|style="background:WhiteSmoke; color:black"|<b>[[Specific_Process_Knowledge/Lithography/UVExposure#Inclined UV Lamp|Inclined UV Lamp]]</b>--> | |||
|- | |- | ||
| Line 46: | Line 47: | ||
*Maskless UV exposure | *Maskless UV exposure | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Top Side Alignment | |||
*Maskless UV exposure | |||
*Direct laser writing | |||
<!--|style="background:WhiteSmoke; color:black"| | |||
*UV exposure | *UV exposure | ||
*DUV exposure | *DUV exposure--> | ||
|- | |- | ||
| Line 53: | Line 58: | ||
|style="background:LightGrey; color:black"|Minimum feature size | |style="background:LightGrey; color:black"|Minimum feature size | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
~1 µm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
~1 µm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
~1 µm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
~1 µm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
~1 µm | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
~1 µm | |||
<!--|style="background:WhiteSmoke; color:black"|--> | |||
|- | |- | ||
|style="background:LightGrey; color:black"|Exposure light/filters/spectrum | |style="background:LightGrey; color:black"|Exposure light/filters/spectrum | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*350W Hg lamp | *350W Hg lamp | ||
*i-line filter (365nm notch filter) | *i-line filter (365nm notch filter)<br>Intensity in Constant Intensity mode:<br>8mW/cm<sup>2</sup> @ 365nm | ||
*303nm filter optional | *303nm filter optional | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*500W Hg-Xe lamp | *500W Hg-Xe lamp | ||
*i-line filter (365nm notch filter) | *i-line filter (365nm notch filter)<br>Intensity in Constant Intensity mode:<br>11mW/cm<sup>2</sup> @ 365nm | ||
*UV350 optics optional | *UV350 optics optional | ||
*UV250 optics optional | *UV250 optics optional | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
365nm LED | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
375nm laserdiodes | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
405nm laserdiodes | |||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*365nm LED | |||
*405nm diode laser | |||
<!--|style="background:WhiteSmoke; color:black"| | |||
*1000 W Hg-Xe lamp | *1000 W Hg-Xe lamp | ||
* | *Dichroic mirror: | ||
**Near UV (350-450nm) | |||
**Mid UV (260-320nm) | |||
**Deep UV (220-260nm)--> | |||
|- | |- | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Flood exposure | |||
*Proximity | *Proximity | ||
*Contact | *Contact: | ||
**Soft contact | **Soft contact | ||
**Hard contact | **Hard contact | ||
**Vacuum contact | **Vacuum contact | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Flood exposure | |||
*Proximity | *Proximity | ||
*Contact | *Contact: | ||
**Soft contact | **Soft contact | ||
**Hard contact | **Hard contact | ||
**Vacuum contact | **Vacuum contact | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Projection | *Projection: | ||
**Pneumatic auto-focus | **Pneumatic auto-focus | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Projection | *Projection: | ||
**Optical auto-focus | **Optical auto-focus | ||
**Pneumatic auto-focus | **Pneumatic auto-focus | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Projection | *Projection: | ||
**Pneumatic auto-focus | **Pneumatic auto-focus | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*Projection: | |||
**Optical auto-focus | |||
**Pneumatic auto-focus | |||
*Direct laser writing: | |||
**Optical auto-focus | |||
**Pneumatic auto-focus | |||
<!--|style="background:WhiteSmoke; color:black"| | |||
*Flood exposure | *Flood exposure | ||
*Proximity exposure with home-made chuck and maskholder | *Proximity exposure with home-made chuck and maskholder | ||
*Inclined exposure | *Inclined exposure | ||
*Rotating exposure | *Rotating exposure--> | ||
|- | |- | ||
| Line 146: | Line 167: | ||
*1 150 mm wafer | *1 150 mm wafer | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*all sizes up to 8inch | *1 small sample, down to 3x3 mm<sup>2</sup> | ||
*1 50 mm wafer | |||
*1 100 mm wafer | |||
*1 150 mm wafer | |||
<!--|style="background:WhiteSmoke; color:black"| | |||
*all sizes up to 8inch--> | |||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
| Line 162: | Line 189: | ||
*All cleanroom materials | *All cleanroom materials | ||
|style="background:WhiteSmoke; color:black"| | |style="background:WhiteSmoke; color:black"| | ||
*All cleanroom materials | *All PolyFabLab materials | ||
<!--|style="background:WhiteSmoke; color:black"| | |||
*All cleanroom materials--> | |||
|- | |- | ||
|} | |} | ||
==Aligner: MA6-1== | |||
== | [[Image:KSAligner in E-4.jpg|300x300px|thumb|The Aligner: MA6-1 is located in PolyFabLab]] | ||
[[Image:KSAligner in E-4.jpg|300x300px|thumb|The | |||
SUSS Mask Aligner MA6 is designed for high resolution photolithography. | SUSS Mask Aligner MA6 is designed for high resolution photolithography. | ||
| Line 176: | Line 202: | ||
Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment. | Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment. | ||
'''The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=44 LabManager] | '''Training videos:''' | ||
[https://www.youtube.com/watch?v=o8IBtfQHNzU Operation] | |||
[https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment] | |||
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=44 LabManager] - '''requires login''' | |||
===Exposure dose=== | ===Exposure dose=== | ||
[[Specific Process Knowledge/Lithography/Resist# | [[Specific Process Knowledge/Lithography/Resist#Aligner: MA6-1|Information on UV exposure dose]] | ||
===Process information=== | ===Process information=== | ||
The | The Aligner: MA6-1 has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm. | ||
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design and locations]]. | |||
===Quality Control (QC)=== | ===Quality Control (QC)=== | ||
| Line 191: | Line 225: | ||
|- | |- | ||
| | | | ||
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for | *[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for Aligner: MA6-1] - '''requires login'''<br> | ||
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=44 The newest QC data for | *[https://labmanager.dtu.dk/view_binary.php?type=data&mach=44 The newest QC data for Aligner: MA6-1] - '''requires login''' | ||
{| {{table}} | {| {{table}} | ||
| align="center" | | | align="center" | | ||
| Line 201: | Line 235: | ||
|- | |- | ||
|Measurement area | |Measurement area | ||
| | |100 mm | ||
|- | |- | ||
|Number of measurements | |Number of measurements | ||
| Line 210: | Line 244: | ||
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px" | {| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px" | ||
!QC limits | !QC limits | ||
! | !Aligner: MA6-1 | ||
|- | |- | ||
|Nominal intensity | |Nominal intensity | ||
| Line 229: | Line 263: | ||
===Equipment performance and process related parameters=== | ===Equipment performance and process related parameters=== | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Alignment and UV exposure | Alignment and UV exposure | ||
|- | |- | ||
| Line 241: | Line 275: | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
vacuum contact, hard contact, soft contact, proximity, flood exposure | vacuum contact, hard contact, soft contact, proximity, flood exposure | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Exposure light/filters | | style="background:LightGrey; color:black"|Exposure light/filters | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*broadband (i-, g-, h-line) | *broadband (i-, g-, h-line) | ||
*365 nm (i-line) | *365 nm (i-line) | ||
| Line 251: | Line 285: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Minimum structure size | |style="background:LightGrey; color:black"|Minimum structure size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
down to 1.25µm | down to 1.25µm | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Mask size | |style="background:LightGrey; color:black"|Mask size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*5x5 inch | *5x5 inch | ||
*7x7 inch | *7x7 inch | ||
| Line 261: | Line 295: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Alignment modes | |style="background:LightGrey; color:black"|Alignment modes | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*Top side (TSA) | *Top side (TSA) | ||
*Backside (BSA) | *Backside (BSA) | ||
| Line 267: | Line 301: | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"|Substrate size | |style="background:LightGrey; color:black"|Substrate size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
* small pieces 1x1cm | * small pieces 1x1cm | ||
* 50 mm wafers | * 50 mm wafers | ||
| Line 274: | Line 308: | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
All cleanroom materials | All cleanroom materials | ||
| Line 280: | Line 314: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Batch | |style="background:LightGrey; color:black"|Batch | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
1 | 1 | ||
|- | |- | ||
| Line 287: | Line 321: | ||
<br clear="all" /> | <br clear="all" /> | ||
==Aligner: MA6 - 2== | ==Aligner: MA6-2== | ||
[[Image:AlignerMA6-2 in E-4.jpg|300x300px|thumb|The Aligner: MA6-2 is placed in E-4]] | |||
[[Image:AlignerMA6-2 in E-4.jpg|300x300px|thumb|The Aligner: MA6 - 2 is placed in E-4]] | |||
The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. | The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. | ||
| Line 299: | Line 330: | ||
The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required. | The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required. | ||
'''The user manual, quality control procedures and results, user APVs, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=339 LabManager] | '''Training videos:''' | ||
[https://www.youtube.com/watch?v=o8IBtfQHNzU Operation] | |||
[https://www.youtube.com/watch?v=rvUuXYgw-xU Alignment] | |||
The user manual, quality control procedures and results, user APVs, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=339 LabManager] - '''requires login''' | |||
===Exposure dose=== | ===Exposure dose=== | ||
[[Specific Process Knowledge/Lithography/Resist#Aligner: | [[Specific Process Knowledge/Lithography/Resist#Aligner:_MA6-2|Information on UV exposure dose]] | ||
===Process information=== | ===Process information=== | ||
The Aligner: MA6 - 2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns. | The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns. | ||
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design and locations]]. | |||
===Quality Control (QC)=== | ===Quality Control (QC)=== | ||
| Line 314: | Line 353: | ||
|- | |- | ||
| | | | ||
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=339 The QC procedure for Aligner: MA6-2]<br> | *[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=339 The QC procedure for Aligner: MA6-2] - '''requires login'''<br> | ||
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=339 The newest QC data for Aligner: MA6-2] | *[https://labmanager.dtu.dk/view_binary.php?type=data&mach=339 The newest QC data for Aligner: MA6-2] - '''requires login''' | ||
{| {{table}} | {| {{table}} | ||
| align="center" | | | align="center" | | ||
| Line 324: | Line 363: | ||
|- | |- | ||
|Measurement area | |Measurement area | ||
| | |100 mm | ||
|- | |- | ||
|Number of measurements | |Number of measurements | ||
| Line 351: | Line 390: | ||
===Alignment=== | ===Alignment=== | ||
'''Top Side Alignment:''' | '''Top Side Alignment:''' | ||
*TSA microscope standard objectives: 5X, and 10X (20X available) | *TSA microscope standard objectives: 5X, and 10X (20X available) | ||
| Line 386: | Line 424: | ||
=== Equipment performance and process related parameters === | === Equipment performance and process related parameters === | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Mask alignment and UV exposure, potentially DUV exposure <sup>1)</sup> | Mask alignment and UV exposure, potentially DUV exposure <sup>1)</sup> | ||
| Line 399: | Line 437: | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
vacuum contact, hard contact, soft contact, proximity, flood exposure | vacuum contact, hard contact, soft contact, proximity, flood exposure | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Exposure light/filters | | style="background:LightGrey; color:black"|Exposure light/filters | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*broadband (i-, g-, h-line) | *broadband (i-, g-, h-line) | ||
*365 nm (i-line) | *365 nm (i-line) | ||
| Line 410: | Line 448: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Minimum structure size | |style="background:LightGrey; color:black"|Minimum structure size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Typically 1.25 µm, possibly down to 0.8 µm <sup>1)</sup> | Typically 1.25 µm, possibly down to 0.8 µm <sup>1)</sup> | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Mask size | |style="background:LightGrey; color:black"|Mask size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*5x5 inch | *5x5 inch | ||
*7x7 inch | *7x7 inch | ||
| Line 420: | Line 458: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Alignment modes | |style="background:LightGrey; color:black"|Alignment modes | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*Top side (TSA) | *Top side (TSA) | ||
*Backside (BSA) | *Backside (BSA) | ||
| Line 426: | Line 464: | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"|Substrate size | |style="background:LightGrey; color:black"|Substrate size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
* small pieces 1x1cm | * small pieces 1x1cm | ||
* 50 mm wafers | * 50 mm wafers | ||
| Line 433: | Line 471: | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
All cleanroom materials except copper and steel | All cleanroom materials except copper and steel | ||
| Line 439: | Line 477: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Batch | |style="background:LightGrey; color:black"|Batch | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
1 | 1 | ||
|- | |- | ||
| Line 445: | Line 483: | ||
<sup>1)</sup> Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available. | <sup>1)</sup> Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available. | ||
===Light intensity and uniformity after lamp ignition=== | |||
[[image:MA6-2_light_intensity_and_uniformity_from_ignition.png|400px|right|thumb|Light intensity and uniformity after lamp ignition. CP is constant power mode, CI is constant intensity mode. The accept limits for the CI mode is ±5%, the accept limit for the non-uniformity is 2%. CP mode has no accept limit.]] | |||
The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing. | |||
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value. | |||
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time. | |||
<br clear="all" /> | <br clear="all" /> | ||
== Aligner: Maskless 01 == | == Aligner: Maskless 01 == | ||
[[Image:Heidelberg_MLA100.jpg|300x300px|thumb|Aligner: Maskless 01 positioned in E-5 (2017)]] | [[Image:Heidelberg_MLA100.jpg|300x300px|thumb|Aligner: Maskless 01 positioned in E-5 (2017)]] | ||
| Line 460: | Line 506: | ||
The system offers top side alignment with high accuracy. | The system offers top side alignment with high accuracy. | ||
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]]. | |||
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=422 LabManager] | |||
The user manual and contact information can be found in LabManager: | |||
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=422 LabManager] - '''requires login''' | |||
===Exposure dose and defocus=== | ===Exposure dose and defocus=== | ||
| Line 475: | Line 524: | ||
=== Equipment performance and process related parameters === | === Equipment performance and process related parameters === | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Alignment and UV exposure | Alignment and UV exposure | ||
|- | |- | ||
| Line 487: | Line 535: | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Projection | Projection | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Exposure light | | style="background:LightGrey; color:black"|Exposure light | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
365nm (LED), FWHM=8nm | 365nm (LED), FWHM=8nm | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Focusing method | |style="background:LightGrey; color:black"|Focusing method | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Pneumatic | Pneumatic | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Minimum structure size | |style="background:LightGrey; color:black"|Minimum structure size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
down to 1µm | down to 1µm | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Design formats | |style="background:LightGrey; color:black"|Design formats | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*'''GDS-II''' | *'''GDS-II''' | ||
*CIF | *CIF | ||
| Line 511: | Line 559: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Alignment modes | |style="background:LightGrey; color:black"|Alignment modes | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Top side only | Top side only | ||
| Line 517: | Line 565: | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"|Substrate size | |style="background:LightGrey; color:black"|Substrate size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
* maximum writing area: 125x125 mm<sup>2</sup> | * maximum writing area: 125x125 mm<sup>2</sup> | ||
* 150 mm wafer | * 150 mm wafer | ||
| Line 525: | Line 573: | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
All cleanroom materials | All cleanroom materials | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Batch | |style="background:LightGrey; color:black"|Batch | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
1 | 1 | ||
|- | |- | ||
| Line 537: | Line 585: | ||
==Aligner: Maskless 02== | ==Aligner: Maskless 02== | ||
MLA150 | MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023). | ||
'''Special features''' | '''Special features''' | ||
* | *Optical Autofocus | ||
*Backside Alignment | *Backside Alignment | ||
*Basic Gray Scale Exposure | *Basic Gray Scale Exposure | ||
| Line 551: | Line 595: | ||
*High Aspect Ratio Mode for exposure of thick resists | *High Aspect Ratio Mode for exposure of thick resists | ||
*200 x 200 mm exposure field | *200 x 200 mm exposure field | ||
*Separate conversion PC | |||
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]]. | |||
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]''' | '''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]''' | ||
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=440 LabManager] | Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=440 LabManager] - '''requires login''' | ||
===Exposure dose and defocus=== | ===Exposure dose and defocus=== | ||
| Line 578: | Line 614: | ||
=== Equipment performance and process related parameters === | === Equipment performance and process related parameters === | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Alignment and UV exposure | Alignment and UV exposure | ||
|- | |- | ||
| Line 590: | Line 626: | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Projection | Projection | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Exposure light | | style="background:LightGrey; color:black"|Exposure light | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*375 | *375 nm<br> | ||
(laser diode arrays) | (laser diode arrays) | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Focusing method | |style="background:LightGrey; color:black"|Focusing method | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*Optical | *Optical | ||
* | *Pneumatic | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Minimum structure size | |style="background:LightGrey; color:black"|Minimum structure size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
down to | down to 1 µm | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Design formats | |style="background:LightGrey; color:black"|Design formats | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*'''GDS-II''' | *'''GDS-II''' | ||
*CIF | *CIF | ||
| Line 617: | Line 652: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Alignment modes | |style="background:LightGrey; color:black"|Alignment modes | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*Top side alignment | *Top side alignment | ||
*Backside alignment | *Backside alignment | ||
| Line 625: | Line 660: | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"|Substrate size | |style="background:LightGrey; color:black"|Substrate size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
* maximum writing area: 200x200 mm<sup>2</sup> | * maximum writing area: 200x200 mm<sup>2</sup> | ||
* 200 mm wafer | * 200 mm wafer | ||
| Line 634: | Line 669: | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
All cleanroom materials | All cleanroom materials | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Batch | |style="background:LightGrey; color:black"|Batch | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
1 | 1 | ||
|- | |- | ||
| Line 648: | Line 683: | ||
==Aligner: Maskless 03== | ==Aligner: Maskless 03== | ||
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH | MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020. | ||
| Line 660: | Line 691: | ||
*Separate conversion PC (Power PC) | *Separate conversion PC (Power PC) | ||
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]]. | |||
'''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]''' | '''[https://www.youtube.com/playlist?list=PLjWVU97LayHAiCabstMfAUeeWyQoQI_cV Training videos]''' | ||
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=464 LabManager] | Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=464 LabManager] - '''requires login''' | ||
===Exposure dose and defocus=== | ===Exposure dose and defocus=== | ||
| Line 676: | Line 708: | ||
=== Equipment performance and process related parameters === | === Equipment performance and process related parameters === | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Alignment and UV exposure | Alignment and UV exposure | ||
|- | |- | ||
| Line 688: | Line 720: | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Projection | Projection | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Exposure light | | style="background:LightGrey; color:black"|Exposure light | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
405nm (laser diode array) | 405nm<br>(laser diode array) | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Focusing method | |style="background:LightGrey; color:black"|Focusing method | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Pneumatic | Pneumatic | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Minimum structure size | |style="background:LightGrey; color:black"|Minimum structure size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
down to | down to 1 µm | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Design formats | |style="background:LightGrey; color:black"|Design formats | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*'''GDS-II''' | *'''GDS-II''' | ||
*CIF | *CIF | ||
| Line 712: | Line 744: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Alignment modes | |style="background:LightGrey; color:black"|Alignment modes | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*Top side alignment | *Top side alignment | ||
*Backside alignment | *Backside alignment | ||
| Line 720: | Line 752: | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"|Substrate size | |style="background:LightGrey; color:black"|Substrate size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
* maximum writing area: 150x150 mm<sup>2</sup> | * maximum writing area: 150x150 mm<sup>2</sup> | ||
* 150 mm wafer | * 150 mm wafer | ||
| Line 728: | Line 760: | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
All cleanroom materials | All cleanroom materials | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Batch | |style="background:LightGrey; color:black"|Batch | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
1 | 1 | ||
|- | |- | ||
| Line 739: | Line 771: | ||
<br clear="all" /> | <br clear="all" /> | ||
== Aligner: Maskless 04 == | |||
<!-- | |||
<span style="background:#FF2800">THIS SECTION IS UNDER CONSTRUCTION</span>[[image:Under_construction.png|200px]] | |||
--> | |||
[[Image:Heidelberg_uMLA.JPG|300x300px|thumb|Aligner: Maskless 04 is installed in PolyFabLab in building 347.]] | |||
The logon password for the PC is "mla" (without quotation marks). | |||
The µMLA Tabletop Maskless Aligner from Heidelberg, located in PolyFabLab in building 347, is a direct exposure lithography tool installed in 2024. | |||
It is a UV exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of a mask. | |||
The system offers top side alignment with good accuracy. | |||
Link to information about [[Specific_Process_Knowledge/Pattern_Design#Alignment_marks|alignment mark design]]. | |||
The user manual and contact information can be found in LabManager: | |||
Equipment info in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=544 LabManager] - '''requires login''' | |||
<!-- | |||
===Exposure dose and defocus=== | |||
[[Specific Process Knowledge/Lithography/Resist#Aligner:_Maskless_04|Information on UV exposure dose]] | |||
--> | |||
===[[Specific Process Knowledge/Lithography/Aligners/Aligner: Maskless 04 processing|Process information]]=== | |||
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Exposure_technology|Exposure technology]] | |||
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Process_Parameters|Process parameters]] | |||
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Substrate_positioning|Substrate positioning]] | |||
*[[Specific_Process_Knowledge/Lithography/Aligners/Aligner:_Maskless_04_processing#Alignment|Alignment]] | |||
=== Equipment performance and process related parameters === | |||
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | |||
!style="background:silver; color:black;" align="center" width="60"|Purpose | |||
|style="background:LightGrey; color:black"| | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
Alignment and UV exposure | |||
|- | |||
!style="background:silver; color:black" align="left" valign="center" rowspan="6"|Performance | |||
|style="background:LightGrey; color:black"|Exposure mode | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
*Projection (Raster mode) | |||
*Direct laser writing (Vector mode) | |||
|- | |||
| style="background:LightGrey; color:black"|Exposure light | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
*365nm (LED) for projection | |||
*405nm (diode laser) for direct laser writing | |||
|- | |||
|style="background:LightGrey; color:black"|Focusing method | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
Pneumatic or Optical | |||
|- | |||
|style="background:LightGrey; color:black"|Minimum structure size | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
Down to 1µm | |||
|- | |||
|style="background:LightGrey; color:black"|Design formats | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
*GDS-II | |||
*CIF | |||
*DXF | |||
*Gerber | |||
*HIMT format | |||
|- | |||
|style="background:LightGrey; color:black"|Alignment modes | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
Top side only | |||
|- | |||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | |||
|style="background:LightGrey; color:black"|Substrate size | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
* maximum writing area: 150x150 mm<sup>2</sup> | |||
* 150 mm wafer | |||
* 100 mm wafer | |||
* 50 mm wafer | |||
* pieces down to 3x3 mm<sup>2</sup> with optical autofocus | |||
|- | |||
| style="background:LightGrey; color:black"|Allowed materials | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
All PolyFabLab materials with sufficient stiffness and flatness. | |||
|- | |||
|style="background:LightGrey; color:black"|Batch | |||
|style="background:WhiteSmoke; color:black" colspan="2"| | |||
1 | |||
|- | |||
|} | |||
<br clear="all" /> | |||
<!-- | |||
==Inclined UV Lamp== | ==Inclined UV Lamp== | ||
[[Image:Inclined UV lamp_1.jpg|300x300px|right|thumb|Inclined UV lamp is placed in CX-1]] | [[Image:Inclined UV lamp_1.jpg|300x300px|right|thumb|Inclined UV lamp is placed in CX-1]] | ||
| Line 748: | Line 874: | ||
The technical specification and the general outline of the equipment can be found in LabManager. | The technical specification and the general outline of the equipment can be found in LabManager. | ||
The user manual(s), quality control procedure(s) and results and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=259 LabManager] - '''requires login''' | |||
<br clear="all" /> | <br clear="all" /> | ||
=== Equipment performance and process related parameters === | === Equipment performance and process related parameters === | ||
{| border=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
!style="background:silver; color:black;" align="center" width="60"|Purpose | !style="background:silver; color:black;" align="center" width="60"|Purpose | ||
|style="background:LightGrey; color:black"| | |style="background:LightGrey; color:black"| | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
UV exposure | UV exposure | ||
|- | |- | ||
| Line 764: | Line 892: | ||
|style="background:LightGrey; color:black"|Exposure mode | |style="background:LightGrey; color:black"|Exposure mode | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
*Flood exposure | *Flood exposure | ||
*Proximity exposure with mask possible for 4inch substrate | *Proximity exposure with mask possible for 4inch substrate | ||
| Line 771: | Line 899: | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Exposure light/filters | | style="background:LightGrey; color:black"|Exposure light/filters | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
* Near UV(350-450nm) | * Near UV(350-450nm) | ||
* Mid UV (260-320nm) | * Mid UV (260-320nm) | ||
| Line 778: | Line 906: | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Minimum structure size | |style="background:LightGrey; color:black"|Minimum structure size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Mask size | |style="background:LightGrey; color:black"|Mask size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
5x5inch optinal | 5x5inch optinal | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Alignment modes | |style="background:LightGrey; color:black"|Alignment modes | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
No alignment possible | No alignment possible | ||
| Line 792: | Line 920: | ||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | !style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | ||
|style="background:LightGrey; color:black"|Substrate size | |style="background:LightGrey; color:black"|Substrate size | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
Up to 8inch substrates, different shapes | Up to 8inch substrates, different shapes | ||
|- | |- | ||
| style="background:LightGrey; color:black"|Allowed materials | | style="background:LightGrey; color:black"|Allowed materials | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
All cleanroom materials | All cleanroom materials | ||
|- | |- | ||
|style="background:LightGrey; color:black"|Batch | |style="background:LightGrey; color:black"|Batch | ||
|style="background:WhiteSmoke; color:black | |style="background:WhiteSmoke; color:black" colspan="2"| | ||
1 | 1 | ||
|- | |- | ||
| Line 806: | Line 934: | ||
<br clear="all" /> | <br clear="all" /> | ||
--> | |||
Revision as of 16:01, 10 March 2025
The content on this page, including all images and pictures, was created by DTU Nanolab staff, unless otherwise stated.
Feedback to this page: click here
UV Exposure Comparison Table
| Equipment | Aligner: MA6-1 | Aligner: MA6-2 | Aligner: Maskless 01 | Aligner: Maskless 02 | Aligner: Maskless 03 | Aligner: Maskless 04 | |
|---|---|---|---|---|---|---|---|
| Purpose |
|
|
|
|
|
| |
| Performance | Minimum feature size |
~1 µm |
~1 µm |
~1 µm |
~1 µm |
~1 µm |
~1 µm |
| Exposure light/filters/spectrum |
|
|
365nm LED |
375nm laserdiodes |
405nm laserdiodes |
| |
| Exposure mode |
|
|
|
|
|
| |
| Substrates | Batch size |
|
|
|
|
|
|
| Allowed materials |
|
|
|
|
|
| |
Aligner: MA6-1

SUSS Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.
Training videos:
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login
Exposure dose
Information on UV exposure dose
Process information
The Aligner: MA6-1 has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm.
Link to information about alignment mark design and locations.
Quality Control (QC)
| Quality Control (QC) for KS Aligner | ||||||||||||||||
Power supply and/or lamp will be adjusted if intensity is outside the limit. |
| Purpose |
Alignment and UV exposure | ||
|---|---|---|---|
| Performance | Exposure mode |
vacuum contact, hard contact, soft contact, proximity, flood exposure | |
| Exposure light/filters |
| ||
| Minimum structure size |
down to 1.25µm | ||
| Mask size |
| ||
| Alignment modes |
| ||
| Substrates | Substrate size |
| |
| Allowed materials |
All cleanroom materials Dedicated chuck for III-V materials | ||
| Batch |
1 | ||
Aligner: MA6-2

The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, low vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA (video) microscope.
The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.
Training videos:
The user manual, quality control procedures and results, user APVs, and contact information can be found in LabManager - requires login
Exposure dose
Information on UV exposure dose
Process information
The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.
Link to information about alignment mark design and locations.
Quality Control (QC)
| Quality Control (QC) for Aligner: MA6-2 | ||||||||||||||||
Power supply and/or lamp will be adjusted if intensity is outside the limit. |
Alignment
Top Side Alignment:
- TSA microscope standard objectives: 5X, and 10X (20X available)
- TSA microscope special objectives: 11.25X offset (for smaller separation)
- Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
- Maximum distance between TSA microscope objectives: 160 mm
- TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)
BackSide Alignment:
- Minimum distance between BSA microscope objectives: 15 mm
- Maximum distance between BSA microscope objectives: 100 mm
- BSA microscope travel range: X +50mm / -16mm; Y +50mm / -20mm (towards flat)
- BSA chuck view ranges:
- 2": X +/- 8-22mm; Y +/- 0-6mm
- 4": X +/- 14-46mm; Y +/- 0-10mm
- 6": X +/- 14-69mm; Y +/- 0-10mm
Microscope field of view (W x H, splitfield):
- TSA 5X
- Oculars: 1.3mm x 2.6mm (Ø2.6mm full field)
- Camera: 350µm x 500µm (700µm x 500µm full field)
- TSA 10X
- Oculars: 0.6mm x 1.3mm (Ø1.3mm full field)
- Camera: 150µm x 250µm (350µm x 250µm full field)
- TSA special
- Oculars: 0.55mm x 1.1mm (Ø1.1mm full field)
- Camera: 150µm x 200µm (300µm x 200µm full field)
- BSA camera
- Low: 1.5mm x 2mm (3mm x 2mm full field)
- High: 450µm x 650µm (950µm x 650µm full field)
| Purpose |
Mask alignment and UV exposure, potentially DUV exposure 1) Bond alignment | ||
|---|---|---|---|
| Performance | Exposure mode |
vacuum contact, hard contact, soft contact, proximity, flood exposure | |
| Exposure light/filters |
| ||
| Minimum structure size |
Typically 1.25 µm, possibly down to 0.8 µm 1) | ||
| Mask size |
| ||
| Alignment modes |
| ||
| Substrates | Substrate size |
| |
| Allowed materials |
All cleanroom materials except copper and steel Dedicated chuck for III-V materials | ||
| Batch |
1 | ||
1) Not available yet. Requires tool change. Would require safety investigation (and a dedicated project) to become available.
Light intensity and uniformity after lamp ignition

The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.
Aligner: Maskless 01

The logon password for the PC is "mla" (without quotation marks).
The MLA 100 Maskless Aligner located in the E-4 cleanroom is a direct exposure lithography tool installed in 2017. It is a UV LED exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of the mask. The system offers top side alignment with high accuracy.
Link to information about alignment mark design.
The user manual and contact information can be found in LabManager:
Equipment info in LabManager - requires login
Exposure dose and defocus
Information on UV exposure dose
Process information
- Exposure technology
- Process parameters
- Substrate positioning
- Alignment
- Optimal use of the maskless aligner
| Purpose |
Alignment and UV exposure | ||
|---|---|---|---|
| Performance | Exposure mode |
Projection | |
| Exposure light |
365nm (LED), FWHM=8nm | ||
| Focusing method |
Pneumatic | ||
| Minimum structure size |
down to 1µm | ||
| Design formats |
| ||
| Alignment modes |
Top side only | ||
| Substrates | Substrate size |
| |
| Allowed materials |
All cleanroom materials | ||
| Batch |
1 | ||
Aligner: Maskless 02
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH (installed 2019 as WMI (0.6µm resolution), rebuilt to WMII 2023).
Special features
- Optical Autofocus
- Backside Alignment
- Basic Gray Scale Exposure
- Advanced Field Alignment Mode for alignment to individual chips/devices on the substrate
- High Aspect Ratio Mode for exposure of thick resists
- 200 x 200 mm exposure field
- Separate conversion PC
Link to information about alignment mark design.
Equipment info in LabManager - requires login
Exposure dose and defocus
Information on UV exposure dose
Process information
| Purpose |
Alignment and UV exposure | ||
|---|---|---|---|
| Performance | Exposure mode |
Projection | |
| Exposure light |
(laser diode arrays) | ||
| Focusing method |
| ||
| Minimum structure size |
down to 1 µm | ||
| Design formats |
| ||
| Alignment modes |
| ||
| Substrates | Substrate size |
| |
| Allowed materials |
All cleanroom materials | ||
| Batch |
1 | ||
1) with optical autofocus
Aligner: Maskless 03
MLA150 WMII maskless aligner from Heidelberg Instruments GmbH, installed 2020.
Special features:
- Backside Alignment
- Advanced Field Alignment Mode for alignment to individual chips/devices on the substrate
- Separate conversion PC (Power PC)
Link to information about alignment mark design.
Equipment info in LabManager - requires login
Exposure dose and defocus
Information on UV exposure dose
Process information
| Purpose |
Alignment and UV exposure | ||
|---|---|---|---|
| Performance | Exposure mode |
Projection | |
| Exposure light |
405nm | ||
| Focusing method |
Pneumatic | ||
| Minimum structure size |
down to 1 µm | ||
| Design formats |
| ||
| Alignment modes |
| ||
| Substrates | Substrate size |
| |
| Allowed materials |
All cleanroom materials | ||
| Batch |
1 | ||
Aligner: Maskless 04
The logon password for the PC is "mla" (without quotation marks).
The µMLA Tabletop Maskless Aligner from Heidelberg, located in PolyFabLab in building 347, is a direct exposure lithography tool installed in 2024. It is a UV exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of a mask. The system offers top side alignment with good accuracy.
Link to information about alignment mark design.
The user manual and contact information can be found in LabManager:
Equipment info in LabManager - requires login
Process information
| Purpose |
Alignment and UV exposure | ||
|---|---|---|---|
| Performance | Exposure mode |
| |
| Exposure light |
| ||
| Focusing method |
Pneumatic or Optical | ||
| Minimum structure size |
Down to 1µm | ||
| Design formats |
| ||
| Alignment modes |
Top side only | ||
| Substrates | Substrate size |
| |
| Allowed materials |
All PolyFabLab materials with sufficient stiffness and flatness. | ||
| Batch |
1 | ||