Specific Process Knowledge/Thin film deposition/Deposition of Chromium
Feedback to this page: click here
Unless otherwise stated, this page is written by DTU Nanolab internal
Studies of Cr deposition processes
Uniformity of Cr layers - Uniformity of Cr layers deposited with different methods and settings
Sputtering of Cr in Wordentec - Settings and deposition rates
Sputtering of Cr in Sputter system (Lesker) - Settings and deposition rates
Sputtering of Cr in Sputter-system Metal-Oxide (PC1) - Settings and deposition rates
Sputtering of Cr in Sputter-system Metal-Nitride (PC3) - Settings and deposition rates
Thermal evaporation of Cr in Thermal evaporator - Settings and deposition results
Stress in sputtered Cr films - Extremely high tensile stress in Cr films deposited at high temperature
Chromium deposition
Chromium can be deposited by e-beam evaporation, thermal evaporation, and sputter deposition. It should be noted that in e-beam evaporation, chromium does not melt but evaporates directly from the solid phase. In the chart below you can compare the different deposition equipment.
E-beam evaporation E-beam evaporation (E-beam evaporator (Temescal) and E-beam evaporator (10-pockets)) | E-beam evaporation and sputter deposition (Wordentec) | Thermal evaporation (Thermal evaporator) | Sputter deposition (Lesker sputterer) | Sputter deposition ((Sputter-system Metal-Oxide (PC1) and Sputter-system Metal-Nitride (PC3)) ) | |
---|---|---|---|---|---|
General description | E-beam deposition of Chromium | E-beam and sputter deposition of Chromium | Thermal deposition of Chromium | Sputter deposition of Chromium | Sputter deposition of Chromium |
Pre-clean | Ar ion gun (only in E-beam evaporator Temescal) | RF Ar clean | RF Ar clean | RF Ar clean | |
Layer thickness | 10Å to 1µm* | 10Å to 1µm* | 80 nm | at least up to 200 nm | at least up to 200 nm |
Deposition rate | 1 Å/s to 10 Å/s | 1 Å/s to 10 Å/s (e-beam)
Sputtering: Depends on process parameters. See here and process log. |
1 Å/s | Depends on process parameters. At least up to 1.48 Å/s. See process log. | Depends on process parameters. |
Batch size |
|
|
|
|
|
Allowed materials |
Almost any that does not degas. See the cross-contamination sheet. |
Almost any that does not degas. See the cross-contamination sheet. |
Almost any that does not degas. See the cross-contamination sheet. |
|
|
Comment | Takes approx. 20 min to pump down | Takes approx. 1 hour to pump down | Takes approx. 20 min to pump down | Takes approx. 5 min to pump down load lock and 6 min for transfer to processing chamber |
* For thicknesses above 300 nm, please request permission from metal@nanolab.dtu.dk to ensure there is enough material.