LabAdviser/314/Microscopy 314-307/FIB/Hydra

From LabAdviser

THIS PAGE IS UNDER CONSTRUCTION

Feedback to this page: click here

Helios 5 Hydra UX DualBeam

Reading Materials

Technical Notes

Resolution

  • Electron beam:
    • At optimum WD: 0.7 nm at 1 kV • 1.0 nm at 500 V (ICD)
    • At coincident point: 0.6 nm at 15 kV • 1.2 nm at 1 kV
  • Xe Ion beam resolution at coincident point:
    • <20 nm at 30 kV using preferred statistical method
    • <10 nm at 30 kV using selective edge method

Attachments

- PFIB column with unique inductively coupled plasma (ICP) source supporting four ion species with fast switching capability. Ion species available: Xe, Ar, O, N

- MultiChem Gas Injection System (GIS) (for W and Pt both E and I beams. C deposition E beam only. XeF2 and H2O etching Ion beam only. Check this)

- Thermo Scientific EasyLift™ NanoManipulator for precise in situ sample manipulation

- Everhart-Thornley SE detector (ETD)

- In-chamber electron and ion detector (ICE) for secondary ions (SI) and electrons (SE)

- EDS detector

- In-chamber Nav-Cam Sample Navigation Camera

- Integrated plasma cleaner

May have: (need to check)

- Thermo Scientific CryoCleaner Decontamination Device

- WDS

- EBSD

- Retractable low-voltage, high-contrast directional solid-state backscatter electron detector (DBS)

- Elstar Column in-lens SE/BSE detector (TLD-SE, TLD-BSE)

- Elstar Column in-column SE/BSE detector (ICD)*


To find the basic instructions for operating the instrument, the reader is referred to the labmanager manual.

Important Notes

  • ALWAYS make sure the EDS (and EBSD relevant?) detectors are RETRACTED when opening and closing the chamber.
  • Do NOT adjust the Nitrogen regulator!!! Relevant?
  • Always open and close the chamber GENTLY whilst viewing the live CCD image.
  • Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
  • Make sure you delete all saved positions before finishing your microscope section.
  • When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece.
  • Use the logbook
  • NEVER work without a valid link between the Z and the FWD
  • ONLY INSERT EBSD CAMERA if the sample is at the correct position (in case of a special EBSD holder the stage must be tilted to 16 °). Relevant?
  • Always use GLOVES. Only vacuum-ready samples and holders are allowed inside the chamber.
  • DO NOT install ANY software on either the microscope or support PC,.
  • Transfer your files at the end of each session. We will be deleting files from the support computer without notice.

Who may operate Helios 5 Hydra UX DualBeam

Equipment performance and process related parameters

Process information