Specific Process Knowledge/Thermal Process/Dope with Phosphorus

From LabAdviser

Feedback to this page: click here


The Phosphorus Predep furnace (A4) can be used for predeposition of silicon wafers with phosphorus, resulting in N-type doping. In the furnace, the silicon wafers are positioned in a quarts boat.

Test of the Phosphorus Predep furnace

Purpose

To study the coherence between the temperature for the predeposition process and drive-in of the doping with phosphorus in the Phosphorus Predep furnace (A4) at DTU Danchip.

Experimental setup

20 boron doped wafers (p-type) were used - Four wafers for each of the five different predeposition temperatures. In the furnace five dummy wafers were placed on each side of the processed wafers. The dummy wafer nearest to the test wafers were changed in-between the runs to miniminze doping from this dummy wafer.

Run # Temperature Process time with POCl3 Anneal time in N2 Wafer #
1 850 oC 15 minutes 20 miuntes 1, 2, 3, 4
2 900 oC 15 minutes 20 miuntes 5, 6, 7, 8
3 950 oC 15 minutes 20 miuntes 9, 10, 11, 12
4 1000 oC 15 minutes 20 miuntes 13, 14, 15, 16
5 1050 oC 15 minutes 20 miuntes 17, 18, 19, 20

After the predeposition two wafers from each run were taken out to be further processed. These wafers were: 1, 2, 5, 6, 9, 10, 13, 14, 17, 18.

These wafers were dipped in BHF to remove the phosphorus glass layer before the drive-in process. The drive-in process was made in the Phosphorus Drive-in furnace (A3). At the drive-in process a dummy wafer was placed in-between the wafers from different temperatures so doping from wafer to wafer was minimized. The drive-in was done with the process "DRY1050" with is a dry oxidation at 1050 oC for 100 minutes and 20 minutes annealing. At the oxidation was the O2 flow was 5 SLM, and the N2 flow for annealing was 3 SLM.

Result

Several measurements were done for the different wafers in the process. After the pre-deposition thickness of the Phosphorus glass layer was measured and the sheet resistance and slice resistivity measured on the same wafer after a BHF etch.

Measurement after Predeposition
Ellipsometer (center point only) Four Point Probe
Wafer # Temperature [C] Thinkness [nm] Refrative index Sheet resistance [Ωsq] Slice Resistivity [Ωcm]
3 850 27,4 1,4623 311 17,32
7 900 45,27 1,4622 138,5 7,61
11 950 61,36 1,4625 16,12 0,859
15 1000 80,45 1,4624 7,4 0,392
19 1050 119,37 1,4623 6,6 0,246


Measurement after Drive-in
Avg. five point on Filmtek Four Point Probe
Wafer # Temperature [C] Thinkness [nm] Refrative index Sheet resistance [Ωsq] Slice Resistivity [Ωcm]
2 850 110,44 1,4654 189,7 13,06
6 900 116,28 1,4629 101,6 5,32
9 950 137,06 1,4604 10,05 0,527
14 1000 141,46 1,4651 4,72 0,216
18 1050 139,87 1,4659 3,23 0,165
Test 110,71 1,46286 - -


Resistivity
Sheet resistance
SIMS Measurement After Pre-dep
SIMS Measurement After Drive-in Process at 1050 C

Looking at the 'SIMS Measurement after Drive-in Process at 1050 C' it can be seen that 'Pre-dep at 950 C' and 'Pre-dep at 1000 C' are crossing each other but they should not do that. There have only been meassured on one wafer so there is not that mush statistical data to verify it with.