Specific Process Knowledge/Lithography/LiftOff

From LabAdviser

Feedback to this page: click here

Comparing HMDS priming

Equipment Lift-off Wet Bench Lift-off (4", 6")
Purpose
  • HMDS priming
  • HMDS priming only
  • HMDS priming and spin coating
Priming chemical

hexamethyldisilizane (HMDS)

Performance Contact angle

standard recipe 82° (on SiO2)

60° - 90°; standard recipe 82° (on SiO2)

Process parameters Process temperature

150°C

50°C

Process time

32.5 minutes

3 min / wafer

Substrates Substrate size
  • 50 mm wafers
  • 100 mm wafers
  • 150 mm wafers

100 mm wafers

Allowed materials

All cleanroom materials

Silicon (with oxide, nitride, or metal films or patterning)

Glass (borosilicate and quartz)

Batch

1 - 25, multiple batches possible

1 - 25



Lift-off Wet Bench

Acetone lift-off: positioned in cleanroom 3

This bench is only for wafers with metal!

Here are the main rules for lift-off bench use:

  • Place the wafers in a dedicated wafer holder.
  • Put the holder in the acetone and start the ultrasound. The strip off time is depending of resist thickness.
  • Rinse your wafers for 4-5 min. in running water after stripping.

Find more info about the lift-off process here: Specific Process Knowledge/Photolithography/AZ5214E standard resist - reverse process


Lift-off (4", 6")

bla bla bla