==Quality Control (QC)==
==Quality Control (QC)==
The purpose of the QC process is to check the intensity and uniformity of the exposure light in the Aligner: MA6-2, using a UV optometer. The optometer is placed at 5 fixed positions, in the exposure area on a dedicated QC chuck, and the intensity of the light is measured.
The 5 intensity measurements are then used to calculate the average intensity as well as the peak uniformity error: <math>PUE[%] = \frac{min-max}{min+max} \sdot 100</math>.
'''Intensity accept limit:'''<br>
The intensity should be corrected if the average intensity (from the 5 measurement points) deviates more than ±5% from nominal value. The nominal value for the Aligner: MA6-2 is 11 mW/cm2, which means that the accept range is 10.45 - 11.55 mW/cm2
'''Uniformity accept limit:'''<br>
The uniformity should be adjusted if the uniformity error is greater than 2%.
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=339 The QC procedure for Aligner: MA6-2] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=339 The newest QC data for Aligner: MA6-2] - '''requires login'''
<!--
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: MA6-2'''
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: MA6-2'''
Line 173: Line 187:
Power supply and/or lamp will be adjusted if intensity is outside the limit.
Power supply and/or lamp will be adjusted if intensity is outside the limit.
|}
|}
-->