===Quality Control (QC)===
[[Image:section under construction.jpg|70px]]
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 01 (MLA1) - Dose and Defoc'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=422 The QC procedure for Aligner: Maskless 01 (MLA1)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=422 The newest QC data for Aligner: Maskless 01 (MLA1)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Dose and defocus test on 1.5µm AZ5214E
|-
|Dose test
|Last QC value ± 20 mJ/cm2 (9 steps total)
|-
|Defoc test
|Last QC value ± 8 mJ/cm2 (9 steps total)
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!Aligner: Maskless 01 (MLA1)
|-
|Dose
|none
|-
|Defoc
|none
|-
|}
|-
|}
Dose and defoc values are reported in the QC data sheet.
|}
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for Aligner: Maskless 01 (MLA1) - Dose and Defoc'''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=422 The QC procedure for Aligner: Maskless 01 (MLA1)] - '''requires login'''<br>
*[https://labmanager.dtu.dk/view_binary.php?type=data&mach=422 The newest QC data for Aligner: Maskless 01 (MLA1)] - '''requires login'''
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Dose and defocus test on 1.5µm AZ5214E
|-
|Dose test
|Last QC value ± 20 mJ/cm2 (9 steps total)
|-
|Defoc test
|Last QC value ± 8 mJ/cm2 (9 steps total)
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:400px"
!QC limits
!Aligner: Maskless 01 (MLA1)
|-
|Nominal intensity
|11 mW/cm<sup>2</sup> @ 365 nm
|-
|Intensity deviation from nominal
|≤5%
|-
|Intensity non-uniformity
|≤2%
|-
|}
|-
|}
Power supply and/or lamp will be adjusted if intensity is outside the limit.
|}