Specific Process Knowledge/Etch/ICP Metal Etcher/Chromium/End point: Difference between revisions
Appearance
| Line 46: | Line 46: | ||
=== End point signal=== | === End point signal=== | ||
====Barc etch==== | ====Barc etch==== | ||
<gallery> | The End point signal shows a drop after 5.5 s and I stopped the etch after 46.4 s giving ''' 41 s after the drop started'''. | ||
<gallery caption="S047676 47s barc etch, Feb-2025, wafer center, 100 nm Cr mask, with 500 nm resist" widths="600px" heights="550px" perrow="2"> | |||
File:data 03 screen shot 01.jpg | File:data 03 screen shot 01.jpg | ||
File:data 03 screen shot 02.jpg | File:data 03 screen shot 02.jpg | ||
| Line 54: | Line 56: | ||
The End point signal shows a drop after 295 s and I stopped the etch after 465 s giving '''170 s after the drop started'''. | The End point signal shows a drop after 295 s and I stopped the etch after 465 s giving '''170 s after the drop started'''. | ||
<gallery caption="S047676 | <gallery caption="S047676 7:43 min Cr etch, Feb-2025, wafer center, 100 nm Cr mask, with 500 nm resist" widths="600px" heights="550px" perrow="2"> | ||
File:data 05 screen shot 02.jpg | File:data 05 screen shot 02.jpg | ||
File:data 05 screen shot 04.jpg | File:data 05 screen shot 04.jpg | ||