Specific Process Knowledge/Lithography/Coaters/Spin Coater: RCD8 processing: Difference between revisions
Appearance
| Line 27: | Line 27: | ||
=Automatic dispense= | =Automatic dispense= | ||
Automatic dispense on the Spin Coater: RCD8 is done using a Nordson EFD Performus VII pressurised cartridge dispense system. The syringe containing the resist is mounted on the media arm, and the resist is dispensed using a nitrogen pressure supplied by the control box. The dispense rate is controlled by the nitrogen pressure, while the dispense time, and thus the volume, is set in the recipe. The dispensed volume may be estimated by measuring the distance the resist level is displaced during a dispense, and calculating the corresponding volume using the diameter of the syringe (23 mm). | Automatic dispense on the Spin Coater: RCD8 is done using a Nordson EFD Performus VII pressurised cartridge dispense system. The syringe containing the resist is mounted on the media arm, and the resist is dispensed using a nitrogen pressure supplied by the control box. The dispense rate is controlled by the nitrogen pressure, while the dispense time, and thus the volume, is set in the recipe. The dispensed volume may be estimated by measuring the distance the resist level is displaced during a dispense, and calculating the corresponding volume using the diameter of the syringe (23 mm). A displacement of 10mm corresponds to a dispensed volume of 4 mL. | ||
Observed dispense rates (uncertainty from measurements ~0.1 ml/s): | Observed dispense rates (uncertainty from measurements ~0.1 ml/s): | ||