Specific Process Knowledge/Lithography/Descum: Difference between revisions
Line 195: | Line 195: | ||
{| {{table}} | {| {{table}} | ||
| align="center" | | | align="center" | | ||
{| border="1" cellspacing="1" cellpadding=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
|- style="background:LightGrey" | |- style="background:LightGrey" | ||
| ||FW/REV|| C2/C1 || Power | | ||FW/REV|| C2/C1 || Power | ||
Line 223: | Line 223: | ||
{| {{table}} | {| {{table}} | ||
| align="center" | | | align="center" | | ||
{| border="1" cellspacing="1" cellpadding=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
|- style="background:LightGrey" | |- style="background:LightGrey" | ||
| ||FW/REV|| C2/C1 || Oxygen || Pressure | | ||FW/REV|| C2/C1 || Oxygen || Pressure | ||
Line 249: | Line 249: | ||
{| {{table}} | {| {{table}} | ||
| align="center" | | | align="center" | | ||
{| border="1" cellspacing="1" cellpadding=" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
|- style="background:LightGrey" | |- style="background:LightGrey" | ||
| ||FW/REV|| C2/C1 || Oxygen || Pressure | | ||FW/REV|| C2/C1 || Oxygen || Pressure |
Revision as of 11:23, 3 February 2023
This section, including all images and pictures, is created by DTU Nanolab staff unless otherwise stated.
Feedback to this page: click here
Plasma Asher 1
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login
Descum of AZ5214E resist on 50mm silicon wafer. Wafer was placed horisontally in chamber on a 100 mm carier wafer.
Recipe 1:
Note: Plasma asher was cold before use
- O2 flow: 70 ml/min
- N2 flow: 70 ml/min
- Power: 150 W
|
Recipe 2:
Note: Plasma asher was cold before use
- O2 flow: 500 ml/min
- N2 flow: 0 ml/min
- Power: 500 W
|
Descum tests on UV resists
Conny Hjort & Jesper Hanberg, September 2021
Descum of different resists, AZ5214E, AZ701 MiR and AZ2020 nLOF, on a single 100mm wafer was tested. Wafer was placed vertically in the middle of glass carrier.
Recipe settings:
Note: plasma Asher was cold before use.
- O2 flow: 70 ml/min
- N2 flow: 70 ml/min
Power: 150 W
Minor temperature rise during processing was observed, but not more than 5 degrees. Starting chamber pressure was around 0.5 mbar.
1,5 um AZ5214E resist:
|
1,5 um AZ5214E resist placed horizontally in the carrier:
|
1,5 um AZ701MiR resist:
|
1,5 um AZ 2020nLOF resist:
|
Plasma Asher 2
Jitka Urbánková & Jesper Hanberg, December 2019
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login
Descum of AZ Mir 701 resist on 100mm silicon wafer. Five wafers were placed vertically in chamber.
recipe 1:
- O2 flow: 100 ml/min
- N2 flow: 100 ml/min
- Power: 150 W
|
recipe 2:
- O2 flow: 500 ml/min
- N2 flow: 0 ml/min
- Power: 200 W
|
A linear time dependence was observed after etching 7 minutes or more (recipe 2).
Plasma Asher 3: Descum
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager - requires login
Plasma Asher 3 is specifically used for controlled descum process after lithography. Please note that you only can process a single 100 mm wafer, or one small sample, at a time. The plasma asher is equipped with 2 gaslines: oxygen and nitrogen, but all standard processes use only oxygen (as recommended by Diener).
Ashing of AZ MiR701 resist:
You can use two different descum process developments: you can either change power settings or processing chamber pressure.
Testing different power settings:
Recipe settings:
- O2 flow: 5 sccm
- N2 flow: 0
- Pressure: 0.2 mbar
- Power: Varied
Experiment parameters:
|
Testing different pressure settings:
Recipe settings:
- O2 flow: varied
- N2 flow: 0
- Pressure: varied
- Power: V100% (100 W)
Experiment parameters:
|
Ashing of AZ5214E resist:
Recipe settings:
- O2 flow: varied
- N2 flow: 0
- Pressure: varied
- Power: V100% (100 W)
Experiment parameters:
|