Jump to content

LabAdviser/314/Microscopy 314-307/FIB/Hydra: Difference between revisions

Fryma (talk | contribs)
Fryma (talk | contribs)
Line 58: Line 58:
* Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
* Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
* Make sure you delete all saved positions before finishing your microscope section.  
* Make sure you delete all saved positions before finishing your microscope section.  
* When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece.
* When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece. Enable safe stage movement.
* Use the logbook
* Use the logbook
* NEVER work without a valid link between the Z and the FWD
* NEVER work without a valid link between the Z and the FWD
Line 64: Line 64:
* DO NOT install ANY software on either the microscope or support PC,.
* DO NOT install ANY software on either the microscope or support PC,.
* Transfer your files at the end of each session.  We will be deleting files from the support computer without notice.
* Transfer your files at the end of each session.  We will be deleting files from the support computer without notice.
* ALWAYS make sure the EBSD detectors are RETRACTED when opening and closing the chamber. '''relevant?'''
* ONLY INSERT EBSD CAMERA if the sample is at the correct position (in case of a special EBSD holder the stage must be tilted to 16 °). '''Relevant?'''


=Who may operate Helios 5 Hydra UX DualBeam=
=Who may operate Helios 5 Hydra UX DualBeam=