=== | === Image acquisition === | ||
1. Choose '''TEM imaging''' under '''Technique Manager''' for general image acquisition.<br /> | 1. Choose '''TEM imaging''' under '''Technique Manager''' for general image acquisition.<br /> | ||
| Line 57: | Line 57: | ||
* '''Correct Drift''' aligns this stack of individual frames.<br /> | * '''Correct Drift''' aligns this stack of individual frames.<br /> | ||
* This means the signal-to-noise in each individual frame should be high enough to get a decent summed final image.<br /> | * This means the signal-to-noise in each individual frame should be high enough to get a decent summed final image.<br /> | ||
==== Diffraction mode ==== | ==== Diffraction mode ==== | ||
| Line 64: | Line 63: | ||
* Increasing the exposure results in summing individual frames (as in imaging mode).<br /> | * Increasing the exposure results in summing individual frames (as in imaging mode).<br /> | ||
* As the minimum exposure time for the individual frame is 4x higher, the signal-to-noise in the individual frame is better compared to imaging mode. This can help for very low dose acquisition.<br /> | * As the minimum exposure time for the individual frame is 4x higher, the signal-to-noise in the individual frame is better compared to imaging mode. This can help for very low dose acquisition.<br /> | ||
<br /> | |||
=== ''in situ'' acquisition === | === ''in situ'' acquisition === | ||