=== image acquisition ===
=== Image acquisition ===


1. Choose '''TEM imaging''' under '''Technique Manager''' for general image acquisition.<br />
1. Choose '''TEM imaging''' under '''Technique Manager''' for general image acquisition.<br />
Line 57: Line 57:
* '''Correct Drift''' aligns this stack of individual frames.<br />
* '''Correct Drift''' aligns this stack of individual frames.<br />
* This means the signal-to-noise in each individual frame should be high enough to get a decent summed final image.<br />
* This means the signal-to-noise in each individual frame should be high enough to get a decent summed final image.<br />


==== Diffraction mode ====
==== Diffraction mode ====
Line 64: Line 63:
* Increasing the exposure results in summing individual frames (as in imaging mode).<br />
* Increasing the exposure results in summing individual frames (as in imaging mode).<br />
* As the minimum exposure time for the individual frame is 4x higher, the signal-to-noise in the individual frame is better compared to imaging mode. This can help for very low dose acquisition.<br />
* As the minimum exposure time for the individual frame is 4x higher, the signal-to-noise in the individual frame is better compared to imaging mode. This can help for very low dose acquisition.<br />
<br />


=== ''in situ'' acquisition ===
=== ''in situ'' acquisition ===