Quality control of measurement accuracy: edited info on calibration
Reet (talk | contribs)
Quality control of measurement accuracy: edited info on calibration
Line 197: Line 197:
|}
|}
|}
|}
If the average value is within the calibration limit, but outside of the action limit, the control measurement should be redone after 1 week.
If the average value is outside the calibration limit, the instrument is calibrated straight away.  


If two consecutive measurements for a range are outside the action limit, the range should be calibrated.
If the average value is within the calibration limit, but outside of the action limit, the control measurement should be redone after 1 week. If two consecutive measurements for a range are outside the action limit, the range should be calibrated.  


The confidence interval on the step height is given as twice the standard deviation, or 95 % confidence intervals.
The accuracy of the step height calibration is 0.15 µm for the 24.865 µm standard and 30 Å for the 9160 Å standard. These values are equivalent to twice the standard deviation, or 95 % confidence intervals.