Specific Process Knowledge/Etch/DRIE-Pegasus/picoscope: Difference between revisions
Line 51: | Line 51: | ||
</gallery> | </gallery> | ||
One can see the multiplexing of the Bosch process by alternation of the gas flows of | One can see the multiplexing of the Bosch process by alternation of the gas flows of | ||
<span style="color:#FFFFFF; background:blue">SF<sub>6</sub></span> and <span style="background:red">C<sub>4</sub>F<sub>8</sub></span> | <span style="color:#FFFFFF; background:blue">SF<sub>6</sub></span> (that has flow initially at 350 sccm for the first 1.5 second then 550 sccm for 5.5 seconds in the etch phase) and <span style="background:red">C<sub>4</sub>F<sub>8</sub></span> (at 200 sccm in the dep phase) | ||
<gallery caption="4 cycles of Process D4 " widths="400" heights="500" perrow="2"> | <gallery caption="4 cycles of Process D4 " widths="400" heights="500" perrow="2"> | ||
Line 64: | Line 64: | ||
[[Image:PrD4 SPTS.png |frame|left|x500px|4 cycles of the recipe PrD4 recorded with the SPTS software]] | [[Image:PrD4 SPTS.png |frame|left|x500px|4 cycles of the recipe PrD4 recorded with the SPTS software]] | ||
[[Image:Process A pico.png |frame|right|x500px|4 cycles of the recipe | [[Image:Process A pico.png |frame|right|x500px|4 cycles of the recipe Process A recorded with the Picoscope]] | ||
[[Image:Process A SPTS.png |frame|left|x500px|4 cycles of the recipe PrD4 recorded with the SPTS software]] | [[Image:Process A SPTS.png |frame|left|x500px|4 cycles of the recipe PrD4 recorded with the SPTS software]] |
Revision as of 21:27, 17 March 2020
Feedback to this page: click here
Process optimization using the Picoscope
Before going into details on why it makes sense to optimize the processes using the picoscope process monitoring, we need to have a look at how this is usually done.
Standard process parameter monitoring
The original standard recipes on Pegasus 1 differ in many ways. The second step of one of them is listed below:
Process A Step 2 parameters | ||
---|---|---|
Parameter | Etch | Dep |
Gas flow (sccm) | SF6 350 (1.5 s), 550 | C4F8 200 |
Cycle time (secs) | 7.0 | 4.0 |
Pressure (mtorr) | 25 (1.5 s), 150 | 25 |
Coil power (W) | 2800 | 2000 |
Platen power (W) | 140 (1.5) 45 | 0 |
Common | Temperature 20 degs |
When running any recipe on the SPTS Pro software, a set of process parameters are recorded as the process runs. This data can be accessed later by looking up the datalog. Which parameters are recorded is dictated by the socalled logging recipe that is selected for every process recipe. The list of possible parameters to include in a logging recipe is long and is comprised of both input parameters such as 'Coil forward power demand' and 'SF6 flow setpoint', and measured values such as 'Platen temperature' and 'Platen DC Bias'. During the process itself, one can also activate the 'Trace' to monitor the process parameters in real time.
To the left in the figure below, some of the most important process parameters are shown for a process run of the recipe Process A.
-
Left: Recorded with the SPTS software
-
Right: Recorded with the Picoscope
One can see the multiplexing of the Bosch process by alternation of the gas flows of SF6 (that has flow initially at 350 sccm for the first 1.5 second then 550 sccm for 5.5 seconds in the etch phase) and C4F8 (at 200 sccm in the dep phase)
-
Left: Recorded with the SPTS software
-
Right: Recorded with the Picoscope