====Quality Control (QC)====
====Quality Control (QC)====


'''THIS SECTION IS UNDER CONSTRUCTION''' [[Image:section under construction.jpg|70px]]
{| border="1" cellspacing="2" cellpadding="2" colspan="3"
|bgcolor="#98FB98" |'''Quality Control (QC) for KS Aligner [[Image:section under construction.jpg|70px]] '''
|-
|
*[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for KS Aligner]<br>
*[http://labmanager.dtu.dk/view_binary.php?fileId=3838 The newest QC data for KS Aligner]
{| {{table}}
| align="center" |
{| border="1" cellspacing="1" cellpadding="2"  align="center" style="width:300px"
! QC Recipe:
! Manual intensity measurement
|-
|Measurement area
|4"
|-
|Number of measurements
|5
|-
|}
| align="center" valign="top"|
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width:300px"
!QC limits
!KS Aligner
|-
|Nominal intensity
|8 mW/cm<sup>2</sup>
|-
|Intensity deviation from nominal
|≤5%
|-
|Intensity non-uniformity
|≤2%
|-
|}
|-
|}
Power supply and/or lamp will be adjusted if intensity is outside the limit.
|}