Specific Process Knowledge/Lithography/UVExposure: Difference between revisions
Appearance
| Line 166: | Line 166: | ||
{| border="1" cellspacing="2" cellpadding="2" colspan="3" | {| border="1" cellspacing="2" cellpadding="2" colspan="3" | ||
|bgcolor="#98FB98" |'''Quality Control (QC) for | |bgcolor="#98FB98" |'''Quality Control (QC) for KS Aligner [[Image:section under construction.jpg|70px]] ''' | ||
|- | |- | ||
| | | | ||
*[http://labmanager.dtu.dk/d4Show.php?id= | *[http://labmanager.dtu.dk/d4Show.php?id=5644&mach=44 The QC procedure for KS Aligner]<br> | ||
*[http://labmanager.dtu.dk/view_binary.php?fileId= | *[http://labmanager.dtu.dk/view_binary.php?fileId=3838 The newest QC data for KS Aligner] | ||
{| {{table}} | {| {{table}} | ||
| align="center" | | | align="center" | | ||
{| border="1" cellspacing="1" cellpadding="2" align="center" style="width: | {| border="1" cellspacing="1" cellpadding="2" align="center" style="width:300px" | ||
! QC Recipe: | ! QC Recipe: | ||
! | ! Manual intensity measurement | ||
|- | |- | ||
| | |Measurement area | ||
| | |4" | ||
|- | |- | ||
| | |Number of measurements | ||
| | |5 | ||
|- | |- | ||
|} | |} | ||
| align="center" valign="top"| | | align="center" valign="top"| | ||
{| border="2" cellspacing="1" cellpadding="2" align="center" style="width: | {| border="2" cellspacing="1" cellpadding="2" align="center" style="width:300px" | ||
!QC limits | !QC limits | ||
! | !KS Aligner | ||
|- | |- | ||
|Nominal | |Nominal intensity | ||
| | |8 mW/cm2 | ||
|- | |- | ||
| | |Intensity deviation from nominal | ||
|< | |<5% | ||
|- | |- | ||
| | |Intensity non-uniformity | ||
|< | |<2% | ||
|- | |- | ||
|} | |} | ||
|- | |- | ||
|} | |} | ||
Power supply and/or lamp will be adjusted if intensity is outside the limit. | |||
|} | |} | ||
=== Equipment performance and process related parameters === | === Equipment performance and process related parameters === | ||