==The ATV Furnace==
==The ATV Furnace==


[[Image:ATV_ovn.jpg|thumb|365x365px|Multipurpose annealing furnace. Positioned in cleanroom B-1]]
[[Image:ATV_ovn.jpg|thumb|365x365px|Multipurpose Annealing Furnace. Positioned in cleanroom B-1]]
[[Image:ATV_ovn_boat.jpg|thumb|335x335px|Boat with wafer for the multipurpose annealing furnace.]]
[[Image:ATV_ovn_boat.jpg|thumb|335x335px|Boat with wafers for the Multipurpose Annealing Furnace.]]


The Multipurpose annealing furnace it made by the company ATV Technologie, and it was installed in the cleanroom in 2015.  
The Multipurpose annealing furnace it made by ATV Technologie, and it was installed in the cleanroom in 2015.  


The purpose of the Multipurpose annealing furnace is annealing and dry oxidation of different samples and resist pyrolysis. Annealling and resist pyrolysis can be done in vaccum or at atmospheric pressure, in a N<sub>2</sub>, H<sub>2</sub> or a mixture of the two gasses.  
The purpose of the Multipurpose annealing furnace is annealing and dry oxidation of different samples and resist pyrolysis. Annealling and resist pyrolysis can be done in vaccum or at atmospheric pressure, in a N<sub>2</sub>, H<sub>2</sub> atmosphere or a mixture of the two gasses.  


All process gasses (except purge nitrogen) are heated heated, before they are introduced into the furnace at the door side.
All process gasses (except purge nitrogen) are heated, before they are introduced into the furnace at the door side.


Is is possible to change all quartz ware in the furnace (the furnace tube, the door sealing and the wafer boat). A the moment Danchip is has three different sets of quartz ware:
Is is possible to change all quartz ware in the furnace (the furnace tube, the door sealing and the wafer boat). At the moment Danchip is has two different sets of quartz ware:


*Clean: Dedicated for clean samples that have been RCA cleaned
*Metal: Dedicated for different samples that cannot be RCA cleaned. Also samples with metals are allowed in the furnace, when this quartz set is mounted
*Metal: Dedicated for different samples that cannot be RCA cleaned. Also samples with metals are allowed in the furnace, when this quartz set is mounted
*Resist pyrolysis: Dedicated for resist pyrolysis.  
*Resist pyrolysis: Dedicated for resist pyrolysis.  
Line 28: Line 27:
The furnace tube is heated by use on 12 long heaters situated along the furnace tube and combined in three groups (top, bottom left and bottom right) and two flat heaters situated in the ends of the furnace tube. The this way the temperature will be very uniform everywhere in the furnace tube. The heating can be done very fast, up to 30 <sup>o</sup>C/min. For atmospheric pressure processes the maximum temperature is 1100 <sup>o</sup>C, and for vacuum processes the maximum temperature is 1050 <sup>o</sup>C.  
The furnace tube is heated by use on 12 long heaters situated along the furnace tube and combined in three groups (top, bottom left and bottom right) and two flat heaters situated in the ends of the furnace tube. The this way the temperature will be very uniform everywhere in the furnace tube. The heating can be done very fast, up to 30 <sup>o</sup>C/min. For atmospheric pressure processes the maximum temperature is 1100 <sup>o</sup>C, and for vacuum processes the maximum temperature is 1050 <sup>o</sup>C.  


The furnace body surrounding the furnace tube consists of a top and bottom half-shell. To cool down the furnace, the top half shelf can be lifted up, and cooling fans will then flow air from the surroundings around the furnace tube to cool it down. However, for safety reasons, the cooling fans will not the activated when there is hydrogen in the furnace, and cooling will then be very slow. Furthermore t
The furnace body surrounding the furnace tube consists of a top and bottom half-shell. To cool down the furnace, the top half shelf can be lifted up, and cooling fans will then flow air from the surroundings around the furnace tube to cool it down. However, for safety reasons, the cooling fans will not the activated when there is hydrogen in the furnace, and cooling will then be very slow. Furthermore it is not possible to open the furnace when the temperature is above 800 <sup>o</sup>C.
 
It is not possible to open the furnace when the temperature is above 800 <sup>o</sup>C.


For resist pyrolysis, samples with different resist layers are heated up to maximum 1100 <sup>o</sup>C in a nitrogen atmosphere. At high temperatures carbon is formed by pyrolysis of the resist. In this way conductive structures can be made from a resist patterned sample. If oxygen from the air or from outgassing of the resist is present in the furnace, the resist layer will be removed, thus it is important to evacuate the furnace and flush it with nitrogen, before a high temperature for resist pyrolysis is obtained. Pyrolysis of a large amount of resist may also be a problem due to resist outgassing.  
For resist pyrolysis, samples with different resist layers are heated up to maximum 1100 <sup>o</sup>C in a nitrogen atmosphere. At high temperatures carbon is formed by pyrolysis of the resist. In this way conductive structures can be made from a resist patterned sample. If oxygen from the air or from outgassing of the resist is present in the furnace, the resist layer will be removed, thus it is important to evacuate the furnace and flush it with nitrogen, before a high temperature for resist pyrolysis is obtained. Pyrolysis of a large amount of resist may also be a problem due to resist outgassing.  

Revision as of 12:26, 26 June 2017

Feedback to this page: click here

Multipurpose annealing furnace

The ATV Furnace

Multipurpose Annealing Furnace. Positioned in cleanroom B-1
Boat with wafers for the Multipurpose Annealing Furnace.

The Multipurpose annealing furnace it made by ATV Technologie, and it was installed in the cleanroom in 2015.

The purpose of the Multipurpose annealing furnace is annealing and dry oxidation of different samples and resist pyrolysis. Annealling and resist pyrolysis can be done in vaccum or at atmospheric pressure, in a N2, H2 atmosphere or a mixture of the two gasses.

All process gasses (except purge nitrogen) are heated, before they are introduced into the furnace at the door side.

Is is possible to change all quartz ware in the furnace (the furnace tube, the door sealing and the wafer boat). At the moment Danchip is has two different sets of quartz ware:

  • Metal: Dedicated for different samples that cannot be RCA cleaned. Also samples with metals are allowed in the furnace, when this quartz set is mounted
  • Resist pyrolysis: Dedicated for resist pyrolysis.

Please note that all new materials have to be approved by the Thin Film group (thinfilm@danchip.dtu.dk) before they are allowed in the furnace.

The furnace tube is heated by use on 12 long heaters situated along the furnace tube and combined in three groups (top, bottom left and bottom right) and two flat heaters situated in the ends of the furnace tube. The this way the temperature will be very uniform everywhere in the furnace tube. The heating can be done very fast, up to 30 oC/min. For atmospheric pressure processes the maximum temperature is 1100 oC, and for vacuum processes the maximum temperature is 1050 oC.

The furnace body surrounding the furnace tube consists of a top and bottom half-shell. To cool down the furnace, the top half shelf can be lifted up, and cooling fans will then flow air from the surroundings around the furnace tube to cool it down. However, for safety reasons, the cooling fans will not the activated when there is hydrogen in the furnace, and cooling will then be very slow. Furthermore it is not possible to open the furnace when the temperature is above 800 oC.

For resist pyrolysis, samples with different resist layers are heated up to maximum 1100 oC in a nitrogen atmosphere. At high temperatures carbon is formed by pyrolysis of the resist. In this way conductive structures can be made from a resist patterned sample. If oxygen from the air or from outgassing of the resist is present in the furnace, the resist layer will be removed, thus it is important to evacuate the furnace and flush it with nitrogen, before a high temperature for resist pyrolysis is obtained. Pyrolysis of a large amount of resist may also be a problem due to resist outgassing.


The user manual, technical information and contact information can be found in LabManager:

Furnace: Multipurpose annealing

Process information

Overview of the performance of the ATV furnace and process related parameters

Purpose
  • Dry oxidation of silicon
  • Annealing in N2, H2 or a mixture of the two gasses
  • Pyrolysis of different resists
Performance Film thickness
  • Dry oxidation: 50 Å to ~200 nm SiO2 (it takes too long to grow a thicker layer)
Process parameter range Process Temperature
  • No vacuum: 25 oC - 1100 oC
  • Vacuum: 25 oC - 1050 oC
Process pressure
  • 1 atm
  • Vacuum down to ~0.1 mbar (depends on gas flow)
Gasses on the system
  • N2: 20 slm
  • O2: 10 slm
  • H2: 5 slm (max 2 slm for vacuum processes)
  • N2 mix : 10 slm (for H2-N2 gas mixture)
Substrates Batch size
  • 1-30 50 mm, 100 mm or 150 mm wafers per run
  • 1-50 200 mm wafers per run (not possible with all quartz sets)
  • Smaller samples (placed in a Si carrier wafer)
Substrate materials allowed
  • Depends on the quartz set