Specific Process Knowledge/Lithography/UVExposure: Difference between revisions
Line 219: | Line 219: | ||
'''The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager:''' | '''The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager:''' | ||
Equipment info in [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=201 LabManager] | Equipment info in [http://labmanager.danchip.dtu.dk/function.php?module=Machine&view=view&mach=201 LabManager] | ||
<br clear="all" /> | |||
=== Equipment performance and process related parameters === | |||
{| border="2" cellspacing="0" cellpadding="2" | |||
!style="background:silver; color:black;" align="center" width="60"|Purpose | |||
|style="background:LightGrey; color:black"| | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
Alignment and UV exposure | |||
|- | |||
!style="background:silver; color:black" align="left" valign="top" rowspan="5"|Performance | |||
|style="background:LightGrey; color:black"|Exposure mode | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
*soft contact, hard contact, proximity, flood exposure | |||
*automatic loading optional for 6inch substrates | |||
*automatic pattern recognition optional for special alignment marks | |||
|- | |||
| style="background:LightGrey; color:black"|Exposure light/filters | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
*365 nm, light intensity 7,0 mW&cm2 in Constant Intansity (CI2)mode | |||
*303 nm filters optinal | |||
|- | |||
|style="background:LightGrey; color:black"|Minimum structure size | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
* down to 1um exposure with 4inch chuck | |||
* down to 5um exposure with 6inch chuck | |||
|- | |||
|style="background:LightGrey; color:black"|Mask size | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
*5x5inch | |||
*7x7inch | |||
|- | |||
|style="background:LightGrey; color:black"|Alignment modes | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
*Top Side Alignment(TSA) | |||
*Bottom Side Alignment(BSA) | |||
|- | |||
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Substrates | |||
|style="background:LightGrey; color:black"|Substrate size | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
* 50 mm wafers | |||
* 110 mm wafers | |||
* 150 mm wafers | |||
|- | |||
| style="background:LightGrey; color:black"|Allowed materials | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
All cleanroom materials except III-V materials | |||
|- | |||
|style="background:LightGrey; color:black"|Batch | |||
|style="background:WhiteSmoke; color:black" align="center" colspan="2"| | |||
*1 for 4inch substrates | |||
*25 for 6inch substrates in automatic loading mode | |||
|- | |||
|} | |||
<br clear="all" /> | |||
Revision as of 09:18, 4 October 2013
UV Exposure Comparison Table
Equipment | KS Aligner | Aligner-6inch | III-V Aligner | Inclined UV Lamp | |
---|---|---|---|---|---|
Purpose |
|
|
|
| |
Performance | Minimum feature size |
|
|
|
|
Exposure light/filters/spectrum |
|
|
|
| |
Exposure mode |
|
|
|
| |
Process parameter range | Positive Process |
|
|
|
|
Negative Process |
then 210mW/cm2 flood exposure after PEB |
then 210mW/cm2 flood exposure after PEB |
|
| |
Substrates | Batch size |
|
|
|
|
Allowed materials |
|
|
|
|
KS Aligner
Feedback to this section: click here
SUSS Mask Aligner MA6 is designed for high resolution photolithography.
The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, hard, soft, proximity) are supplied.
Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager: Equipment info in LabManager
Purpose |
Alignment and UV exposure | ||
---|---|---|---|
Performance | Exposure mode |
soft contact, hard contact, proximity, flood exposure | |
Exposure light/filters |
| ||
Minimum structure size |
down to 1µm | ||
Mask size |
| ||
Alignment modes |
| ||
Substrates | Substrate size |
| |
Allowed materials |
| ||
Batch |
1 |
Aligner-6inch
Feedback to this section: click here
Aligner-6inch, EVG620 aligner, is designed for high resolution photolithography. The machine can be used for 2, 4 and 6 inch substrates. Cassette-to-cassette handling option is available only for 6inch substrates. The automatic pattern recognition software is available for the special alignment marks design recommended of EVGroup. Please contact Danchip staff for further information. Available exposure mode: proximity, soft, hard and vacuum contact. Two alignment options are available: top side alignment (TSA) and back side alignment (BSA). IR-light alignment also an option.
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager: Equipment info in LabManager
Purpose |
Alignment and UV exposure | ||
---|---|---|---|
Performance | Exposure mode |
| |
Exposure light/filters |
| ||
Minimum structure size |
| ||
Mask size |
| ||
Alignment modes |
| ||
Substrates | Substrate size |
| |
Allowed materials |
All cleanroom materials except III-V materials | ||
Batch |
|
III-V Aligner
Feedback to this section: click here
The SÜSS MicroTec MA1006 mask aligner located in the III-V cleanroom is dedicated for processing of III-V compound semiconductors.
Specific use of the mask aligner can be found in the standard resist recipes.
Purpose |
Alignment and UV exposure | ||
---|---|---|---|
Performance | Exposure mode |
soft contact, hard contact, proximity, flood exposure | |
Exposure light/filters |
365 nm, 405 nm | ||
Minimum structure size |
down to 2µm | ||
Mask size |
5x5inch | ||
Alignment modes |
Top side only | ||
Substrates | Substrate size |
| |
Allowed materials |
III-V materials | ||
Batch |
1 |
Inclined UV Lamp
Feedback to this section: click here
The Inclined UV lamp is 1000 W Hg(Xe)lamp source designed for near UV, 350-450nm, mid UV, 260-320nm, and deep UV, 220-260nm exposures of resists and polymers. The exposure source can be also used to make an inclined exposure in air or in the media tank.
The tool was purchased in February 2009 from Newport. The exposure lamp has a official name: Oriel Flood Exposure Source, unit 92540. All other parts of equipment: substrate and mask holder with media tank, exhaust box around the tool, timer controller, were designed and build at DTU Danchip workshop.
The substrate and mask holder with a media tank was designed as part of Master Thesis of DTU Nanotech, Andres Kristensen group. The exhaust box was made as part of safety and the timer controller was build to control exposure time.
The technical specification and the general outline of the equipment can be found in LabManager.
The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager: Equipment info in LabManager