Specific Process Knowledge/Lithography/UVExposure/aligner MA6-2: Difference between revisions
| Line 1: | Line 1: | ||
=Aligner: MA6-2= | |||
[[Image:AlignerMA6-2 in E-4.jpg|400px|thumb|The Aligner: MA6-2 is located in E-4.]] | [[Image:AlignerMA6-2 in E-4.jpg|400px|thumb|The Aligner: MA6-2 is located in E-4.]] | ||
| Line 18: | Line 18: | ||
The user manual, quality control procedures and results, user APVs, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=339 LabManager] - '''requires login''' | The user manual, quality control procedures and results, user APVs, and contact information can be found in [http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=339 LabManager] - '''requires login''' | ||
==Exposure dose== | |||
[[Specific Process Knowledge/Lithography/Resist/UVresist/exposureDoseMaskAligners#Aligner:_MA6-2|Information on UV exposure dose]] | [[Specific Process Knowledge/Lithography/Resist/UVresist/exposureDoseMaskAligners#Aligner:_MA6-2|Information on UV exposure dose]] | ||
==Process information== | |||
The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns. | The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns. | ||
| Line 67: | Line 67: | ||
|} | |} | ||
==Alignment== | |||
'''Top Side Alignment:''' | '''Top Side Alignment:''' | ||
*TSA microscope standard objectives: 5X, and 10X (20X available) | *TSA microscope standard objectives: 5X, and 10X (20X available) | ||
| Line 101: | Line 101: | ||
**High: 450µm x 650µm (950µm x 650µm full field) | **High: 450µm x 650µm (950µm x 650µm full field) | ||
=== Equipment performance and process related parameters | ==Light intensity and uniformity after lamp ignition== | ||
[[image:MA6-2_light_intensity_and_uniformity_from_ignition.png|400px|right|thumb|Light intensity and uniformity after lamp ignition. CP is constant power mode, CI is constant intensity mode. The accept limits for the CI mode is ±5%, the accept limit for the non-uniformity is 2%. CP mode has no accept limit.]] | |||
The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing. | |||
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value. | |||
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time. | |||
==Equipment performance and process related parameters== | |||
{|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | {|border="1" cellspacing="1" cellpadding="10" style="text-align:left;" | ||
| Line 161: | Line 171: | ||
|} | |} | ||
<sup>1)</sup> Not available | <sup>1)</sup> Not available. Requires tool change. Would require safety investigation (and a dedicated project) to become available. | ||
<br clear="all" /> | <br clear="all" /> | ||
Revision as of 12:51, 13 January 2026
Aligner: MA6-2

The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, low vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA (video) microscope.
The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.
Training videos:
The user manual, quality control procedures and results, user APVs, and contact information can be found in LabManager - requires login
Exposure dose
Information on UV exposure dose
Process information
The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.
Link to information about alignment mark design and locations.
Quality Control (QC)
| Quality Control (QC) for Aligner: MA6-2 | ||||||||||||||||
Power supply and/or lamp will be adjusted if intensity is outside the limit. |
Alignment
Top Side Alignment:
- TSA microscope standard objectives: 5X, and 10X (20X available)
- TSA microscope special objectives: 11.25X offset (for smaller separation)
- Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
- Alignment of smaller separations is possible using the "scan microscope" function
- Maximum distance between TSA microscope objectives: 160 mm
- TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)
BackSide Alignment:
- Minimum distance between BSA microscope objectives: 15 mm
- Maximum distance between BSA microscope objectives: 100 mm
- BSA microscope travel range: X +50mm / -16mm; Y +50mm / -20mm (towards flat)
- BSA chuck view ranges:
- 2": X +/- 8-22mm; Y +/- 0-6mm
- 4": X +/- 14-46mm; Y +/- 0-10mm
- 6": X +/- 14-69mm; Y +/- 0-10mm (OBS: Max. separation of BSA microscopes is 100mm)
Microscope field of view (W x H, splitfield):
- TSA 5X
- Oculars: 1.3mm x 2.6mm (Ø2.6mm full field)
- Camera: 350µm x 500µm (700µm x 500µm full field)
- TSA 10X
- Oculars: 0.6mm x 1.3mm (Ø1.3mm full field)
- Camera: 150µm x 250µm (350µm x 250µm full field)
- TSA special
- Oculars: 0.55mm x 1.1mm (Ø1.1mm full field)
- Camera: 150µm x 200µm (300µm x 200µm full field)
- BSA camera
- Low: 1.5mm x 2mm (3mm x 2mm full field)
- High: 450µm x 650µm (950µm x 650µm full field)
Light intensity and uniformity after lamp ignition

The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.
| Purpose |
Mask alignment and UV exposure, potentially DUV exposure 1) Bond alignment | ||
|---|---|---|---|
| Performance | Exposure mode |
vacuum contact, hard contact, soft contact, proximity, flood exposure | |
| Exposure light/filters |
| ||
| Minimum structure size |
Typically 1.25 µm, possibly down to 0.8 µm 1) | ||
| Mask size |
| ||
| Alignment modes |
| ||
| Substrates | Substrate size |
| |
| Allowed materials |
All cleanroom materials except copper and steel Dedicated chuck for III-V materials | ||
| Batch |
1 | ||
1) Not available. Requires tool change. Would require safety investigation (and a dedicated project) to become available.