Line 8: Line 8:


==Choosing measurement parameters==
==Choosing measurement parameters==
A detailed manual on the Aeris XRD is found on the [https://wiki.iodp.tamu.edu/display/LMUG/Aeris+Advanced+User+Guide wiki of the International Ocean Discovery Project]. It has helpful hints on adjusting the hardware including choices on beam knife height and slit size. It also has some pictures from the XRDMP Creator software. Some info from there:


..."Scanning at angles less than 5° 2θ you will need to lower the knife (anti scatter)."
Some helpful tips can now be found in the [https://labmanager.dtu.dk/d4Show.php?id=6584&mach=463#dafs72253 manual on LabManager], section 4. They are mostly copied from the XRDMP Creater help document, which we for copyright reasons cannot make available here.


.... "at low angles you also need a narrower slit (which will reduce intensity but also the irradiated area)."
A detailed manual on the Aeris XRD is found online on the [https://wiki.iodp.tamu.edu/display/LMUG/Aeris+Advanced+User+Guide wiki of the International Ocean Discovery Project]. It also has helpful hints on adjusting the hardware including choices on beam knife height and slit size. It also has some pictures from the XRDMP Creator software. A useful nugget of info from there which you will not find in our manual concerns the beam knife position: ..."Scanning at angles less than 5° 2θ you will need to lower the knife (anti scatter)."
 
==Fluorescence suppression==
It is possible to adjust the Pulse Height Discrimination levels of the Pixcel 1D detector on our machine. This can help suppress signal from fluorescence, which is at a slightly different energy level than the pure copper K-alpha signal. The exact energy level of the fluorescence of course depends on the material you are measuring.
 
Here is a comparison of a measurement with and without fluorescence suppression:
[[File:fluoresc supp feoh.png|400px|caption:FeOH measured with and without fluorescence suppression.]]