Specific Process Knowledge/Thermal Process/Dope with Phosphorus: Difference between revisions

From LabAdviser
Pevo (talk | contribs)
No edit summary
Pevo (talk | contribs)
No edit summary
Line 292: Line 292:


<br clear="all" />
<br clear="all" />
====Phosphorous doping test after change of hardware====
Det er helt fint – Du må selvfølgelig gerne spørge til status på A4 ovnen.
Der blev åbnet POCl-flasken i går, og i dag laver vi en doteringstest med en testskive i ovnen. Jeg unloader testskiven i eftermiddag og måler sheet resistance på den for at kunne sammenligne med tidligere resultater i LabAdviser:
http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/Dope_with_Phosphorus
Det er denne proces, vi forsøger at gentage (se nederst på siden i LabAdviser):
-          RCA rens: Testskive (p-type, 1-20 Ωcm) + dummy-skiver
-          Forfor pre-dep i A4 ovnen: ”POCLNEW” (ny opskrift), 900°C, 15 min fosfor pre-dep + 20 min annealing
-          BHF dyp ved RCA-rensen: Fjern fosfor-glas fra testskiven
-          4 Point Probe: Mål sheet resistance på testskiven

Revision as of 08:54, 26 August 2022

Feedback to this page: click here


The Phosphorus Predep furnace (A4) can be used for phosphorus pre-deposition of silicon wafers, resulting in n-type doping. In the furnace, the silicon wafers are positioned vertically in a quarts boat.


Test of the Phosphorus Predep furnace

Purpose

To study the correlation between the temperature for the predeposition process and drive-in of the phosphorus doping in the Phosphorus Predep furnace (A4) at DTU Nanolab.

Experimental setup

20 boron doped device wafers (p-type) were used - Four wafers for each of the five different predeposition temperatures (see table below). In the furnace five dummy wafers were placed on each side of the device wafers. The dummy wafers nearest to the device wafers were changed in-between the runs to minimize doping from these dummy wafers.

Run # Temperature Process time with POCl3 Annealing time in O2 Wafer #
1 850 oC 15 minutes 20 minutes 1, 2, 3, 4
2 900 oC 15 minutes 20 minutes 5, 6, 7, 8
3 950 oC 15 minutes 20 minutes 9, 10, 11, 12
4 1000 oC 15 minutes 20 minuntes 13, 14, 15, 16
5 1050 oC 15 minutes 20 minutes 17, 18, 19, 20

After the predeposition two wafers from each run were taken out to be further processed. These wafers were: 1, 2, 5, 6, 9, 10, 13, 14, 17, 18. These wafers were dipped in BHF to remove the phosphorus glass layer before the drive-in process.

The drive-in process was made in the Phosphorus Drive-in furnace (A3) for all the mentioned wafers at same time. At the drive-in process a dummy wafer was placed in-between the wafers that have been predeposited at different temperatures, so doping from wafer to wafer was minimized. The phosphorus drive-in was done with the process "DRY1050" which is a dry oxidation at 1050 oC for 100 minutes and 20 minutes annealing. At the oxidation the O2 flow was 5 SLM, and the N2 flow for annealing was 3 SLM.

Results

Several measurements were done for the different device wafers. After the predeposition, the thickness of the grown phosphorus glass layer was measured, and the sheet resistance and slice resistivity were measured on the same wafer after a BHF etch.

Measurements after phosphorus predeposition
Ellipsometer (center point only) Four Point Probe
Wafer # Temperature [C] Oxide thickness [nm] Refractive index Sheet resistance [Ωsq] Slice Resistivity [Ωcm]
3 850 27,4 1,4623 311 6,22*10-3
7 900 45,27 1,4622 138,5 4,85*10-3
11 950 61,36 1,4625 16,1 1,29*10-3
15 1000 80,45 1,4624 7,4 0,89*10-3
19 1050 119,37 1,4623 6,6 1,49*10-3

The oxide layer was etch in BHF before the drive-in of the doping with a dry oxidation at 1050 oC for 100 minutes and 20 minutes annealing.

Measurements after phosphorus drive-in
Avg. five point on Filmtek Four Point Probe
Wafer # Temperature [oC] Oxide thickness [nm] Refractive index Sheet resistance [Ωsq] Slice Resistivity [Ωcm]
2 850 110,44 1,4654 189,7 3,79*10-3
6 900 116,28 1,4629 101,6 3,56*10-3
9 950 137,06 1,4604 10,0 0,8*10-3
14 1000 141,46 1,4651 4,72 0,57*10-3
18 1050 139,87 1,4659 3,23 0,73*10-3
Blank Si wafer 110,71 1,46286 - -


Resistivity
Sheet resistance
SIMS Measurement After Pre-dep
SIMS Measurement After Drive-in Process at 1050 oC

Looking at the "SIMS Measurement After Drive-in Process at 1050 oC" graph, it can be seen that the two curves "Pre-dep at 950 oC" and "Pre-dep at 1000 oC" are crossing each other, but according to the theory they should not do that. Only one wafer has been measured so there is not that much statistical data to verify it with.

Test of the Phosphorus Predep furnace

Purpose

To find the result of a specific process using the Phosphorus Predep furnace (A4) followed by a drive in and oxidating the substrate in the Phosphorus drive in furnace (A3).

Experimental setup

Substrate: p-type Si(100,)

Resistivity: 1-20 Ωcm

Recipe: POCL900

Temperature: 900 °C

Time: 15 min

Anneal: 20 min in N2

After removing the oxide in BHF the wafers were oxidized in Phosphor Drive in Furnace (A3) using DRY1050 at 1050°C for 70 min without further annealing in N2. Results are listed in the table below:

Results

Measurements after phosphorus predeposition
Afer predeposition After Oxidation
Test date Sheet resistance [Ω/sq] Oxide thickness [nm] Sheet resistance [Ω/sq
Aug2012 47.0 93.1 41.8
Jan2013 50.5 91.3 55.6
Aug2013 58.0 91.8 41.8
Jan2014 54.1 93.3 33.8
Jan2015 50.8 94.2 39.4


One wafer from Jan2015 was analyzed with SIMS and the profile is shown on the graph below:

SIMS Measurement After Predeposition, before and after oxidation respectively


Result from GreenBelt, @nanolab,Jan 2015


Test of the Phosphorus Predep furnace

Purpose

To find the doping profile after a specific process flow. The test was made by only performing the high-temperature process step after the pre-deposition step done in the Phosphorus Predep furnace (A4). The subsequently high temperature processes was done by using the Phosphorous drive in furnace (A3) and the Oxide Anneal Furnace (C3).

Experimental setup

Substrate: p-type Si(100,)

Resistivity: 1-20 Ωcm

The total thermal budget for the wafers:

Predeposition in Furnace (A4): POCL1000, 15 min, no further anneal

Wet Oxidation in Furnace (A3): WET1050, 30 min, no further anneal

Anneal in Furnace (C1): ANN1000, 30 min in N2

Wet oxidation in Furnace (C1): WET1050, 30 min, no further anneal

Results

The wafer was analyzed with SIMS and the profile is shown on the graph below:


SIMS Measurement after Phosphorous Predeposition and several oxidations and annealings


Result from Trine Holm Christensen, @Space, Feb. 2015



Phosphorous doping test after change of hardware

Det er helt fint – Du må selvfølgelig gerne spørge til status på A4 ovnen.

Der blev åbnet POCl-flasken i går, og i dag laver vi en doteringstest med en testskive i ovnen. Jeg unloader testskiven i eftermiddag og måler sheet resistance på den for at kunne sammenligne med tidligere resultater i LabAdviser: http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/Dope_with_Phosphorus

Det er denne proces, vi forsøger at gentage (se nederst på siden i LabAdviser):

- RCA rens: Testskive (p-type, 1-20 Ωcm) + dummy-skiver - Forfor pre-dep i A4 ovnen: ”POCLNEW” (ny opskrift), 900°C, 15 min fosfor pre-dep + 20 min annealing - BHF dyp ved RCA-rensen: Fjern fosfor-glas fra testskiven - 4 Point Probe: Mål sheet resistance på testskiven