=Important Notes=
=Important Notes=
 
* ALWAYS make sure the EDS (and EBSD '''relevant?''') detectors are RETRACTED when opening and closing the chamber.
* Do NOT adjust the Nitrogen regulator!!! ''' Relevant?'''
* Always open and close the chamber GENTLY whilst viewing the live CCD image.
* Only home the stage with open door or if in vacuum only with holders that you are 100% sure it is safe to do so.
* Make sure you delete all saved positions before finishing your microscope section.
* When moving between saved positions, BE CAREFUL! If you are unsure move the Z height to the lowest point and lock it so as not to collide with the pole piece.
* Use the logbook
* NEVER work without a valid link between the Z and the FWD
* ONLY INSERT EBSD CAMERA if the sample is at the correct position (in case of a special EBSD holder the stage must be tilted to 16 °). '''Relevant?'''
* Always use GLOVES. Only vacuum-ready samples and holders are allowed inside the chamber.
* DO NOT install ANY software on either the microscope or support PC,.
* Transfer your files at the end of each session.  We will be deleting files from the support computer without notice.