== Picoscope examples ==
== Picoscope examples ==


Using the Picoscope is straightforward. The Start/Stop recording buttons are located in the lower left corner of the GUI; the recording has to be done manually. The settings in the top of the GUI
Using the Picoscope is straightforward. The Start/Stop recording buttons are located in the lower left corner of the GUI; the recording has to be done manually. The settings in the top of the GUI may look like this
[[image:Picodump.png|800px]]


[[image:PicoSetup.PNG|700px]]


are usually:
 
{| {{table}}
{| border="1" cellspacing="0" cellpadding="1"  align="center"
! colspan="3" |[[Specific Process Knowledge/Etch/DRIE-Pegasus/processD#The_new_standard_process_Process_D4 |Process D4 ]]
|-
! Channel
! Label
! Measurement
! Measurement range
|-
! A
| SF<sub>6</sub>
| DC
| ±10V
|-
! B
| SF<sub>6</sub>2.2
| 1
|-
! Pressure (mtorr)
| 26
| 20
|-
! Coil power (W)
| 2500
| 2000
|-
! Platen power (W)
| 35
| 0
|-
! Common
| colspan="2" | Temperature 0 degs
|-
|}


= New recipes=
= New recipes=