Specific Process Knowledge/Lithography/UVExposure: Difference between revisions

From LabAdviser
Taran (talk | contribs)
Taran (talk | contribs)
Line 396: Line 396:
|-
|-


!style="background:silver; color:black" align="left" valign="center" rowspan="5"|Performance
!style="background:silver; color:black" align="left" valign="center" rowspan="6"|Performance


|style="background:LightGrey; color:black"|Exposure mode
|style="background:LightGrey; color:black"|Exposure mode
Line 405: Line 405:
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
375nm or 405nm (laser diode arrays)
375nm or 405nm (laser diode arrays)
|-
|style="background:LightGrey; color:black"|Focusing method
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
*Optical
*Pneumatic
|-
|-
|style="background:LightGrey; color:black"|Minimum structure size
|style="background:LightGrey; color:black"|Minimum structure size
Line 420: Line 425:
|style="background:LightGrey; color:black"|Alignment modes
|style="background:LightGrey; color:black"|Alignment modes
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
|style="background:WhiteSmoke; color:black" align="center" colspan="2"|
Top side alignment
*Top side alignment
Backside alignment
*Backside alignment
Field alignment
*Field alignment


|-
|-

Revision as of 17:44, 27 March 2019

Feedback to this page: click here

UV Exposure Comparison Table

Equipment KS Aligner Aligner: MA6 - 2 Aligner: Maskless 01 Aligner: Maskless 02 Inclined UV Lamp
Purpose
  • Top Side Alignment
  • Back Side Alignment
  • UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • UV exposure
  • (DUV exposure)
  • Bond alignment
  • Top Side Alignment
  • Maskless UV exposure
  • Top Side Alignment
  • Back Side Alignment
  • Maskless UV exposure
  • UV exposure
  • DUV exposure
Performance Minimum feature size
  • 1.5µm down to 1µm
  • 1.5µm down to 0.5µm
  • 2µm down to 1µm
  • 1µm down to 0.6µm
Exposure light/filters/spectrum
  • 350W Hg lamp
  • i-line filter (365nm notch filter), intensity in Constant Intensity mode: 8mW/cm2 @ 365nm
  • 303nm filter optional
  • 500W Hg-Xe lamp
  • i-line filter (365nm notch filter), intensity in Constant Intensity mode: 11mW/cm2 @ 365nm
  • UV350 optics optional
  • UV250 optics optional
  • 365nm LED
  • 375nm laserdiodes

or

  • 405nm laserdiodes
  • 1000 W Hg-Xe lamp
  • near UV (350-450nm), mid UV (260-320nm), and deep UV (220-260nm)
Exposure mode
  • proximity, soft, hard, vacuum contact
  • proximity, soft, hard, vacuum contact
  • Projection
  • Projection
  • Flood exposure
  • Proximity exposure with home-made chuck and maskholder
Substrates Batch size
  • 1 small sample
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 5x5 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 small sample, down to 3x3 mm2
  • 1 50 mm wafer
  • 1 100 mm wafer
  • 1 150 mm wafer
  • 1 200 mm wafer
  • all sizes up to 8inch
Allowed materials
  • All cleanroom materials except copper and steel
  • Dedicated 2inch chuck for III-V materials
  • All cleanroom materials except copper and steel
  • Dedicated chuck for III-V materials
  • All cleanroom materials
  • All cleanroom materials
  • All cleanroom materials


KS Aligner

The KS Aligner is placed in E-4

Feedback to this section: click here

SUSS Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA microscope. It is also possible to make IR- light alignment.

Training videos: Operation Alignment (on Aligner: MA6-2, but KS Aligner is the same model)

The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager.

Process information

The KS Aligner has an i-line notch filter installed. This results in an exposure light peak around 365nm with a FWHM of 7nm. Dependent on the spectral sensitivity of the resist, the optimal dose may be increased compared to broadband exposure on the Aligner-6inch.

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

vacuum contact, hard contact, soft contact, proximity, flood exposure

Exposure light/filters
  • broadband (i-, g-, h-line)
  • 365 nm (i-line)
  • 303 nm
Minimum structure size

down to 1.25µm

Mask size
  • 5x5 inch
  • 7x7 inch
  • special holder for 4 x 2" designs on 5x5 inch
Alignment modes
  • Top side (TSA)
  • Backside (BSA)
Substrates Substrate size
  • small pieces 1x1cm
  • 50 mm wafers
  • 100 mm wafers
  • 150 mm wafers
Allowed materials

All cleanroom materials

Dedicated chuck for III-V materials

Batch

1


Aligner: MA6 - 2

The Aligner: MA6 - 2 is placed in E-4

Feedback to this section: click here

The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, low vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA (video) microscope.

The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.

Training videos: Operation Alignment

The user manual, APV, and contact information can be found in LabManager.

Process information

The Aligner: MA6 - 2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.

Equipment performance and process related parameters

Purpose

Mask alignment and UV exposure, potentially DUV exposure 1)

Bond alignment

Performance Exposure mode

vacuum contact, hard contact, soft contact, proximity, flood exposure

Exposure light/filters
  • broadband (i-, g-, h-line)
  • 365 nm (i-line)
  • "UV300" (280-350 nm)
  • DUV (240 nm) 1)
Minimum structure size

Typically 1.25 µm, possibly down to 0.8 µm 1)

Mask size
  • 5x5 inch
  • 7x7 inch
  • special holder for 4 x 2" designs on 5x5 inch
Alignment modes
  • Top side (TSA)
  • Backside (BSA)
Substrates Substrate size
  • small pieces 1x1cm
  • 50 mm wafers
  • 100 mm wafers
  • 150 mm wafers
Allowed materials

All cleanroom materials except copper and steel

Dedicated chuck for III-V materials

Batch

1

1) Not available yet.


Aligner: Maskless 01

Feedback to this section: click here

Aligner: Maskless 01 positioned in E-5

The MLA 100 Maskless Aligner located in the E-5 cleanroom is a direct exposure lithography tool. It is a UV LED exposure system, that exposes the patterns directly on photosensitive resists on chips, 2, 4, and 6 inch substrates, without prior fabrication of the mask. The system offers top side alignment with high accuracy.

The user manual and contact information can be found in LabManager: Equipment info in LabManager

Process information

Exposure technology

Process parameters

Substrate positioning

Alignment

Optimal use of the maskless aligner

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light

365nm (LED), FWHM=8nm

Minimum structure size

down to 3µm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes

Top side only

Substrates Substrate size
  • maximum writing area: 125x125 mm2
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 5x5 mm2
Allowed materials

All cleanroom materials

Batch

1


Aligner: Maskless 02

Feedback to this section: click here

THIS SECTION IS UNDER CONSTRUCTION

Process information

Equipment performance and process related parameters

Purpose

Alignment and UV exposure

Performance Exposure mode

Projection

Exposure light

375nm or 405nm (laser diode arrays)

Focusing method
  • Optical
  • Pneumatic
Minimum structure size

down to 600nm

Design formats
  • GDS-II
  • CIF
  • DXF
  • Gerber
  • HIMT format
Alignment modes
  • Top side alignment
  • Backside alignment
  • Field alignment
Substrates Substrate size
  • maximum writing area: 200x200 mm2
  • 200 mm wafer
  • 150 mm wafer
  • 100 mm wafer
  • 50 mm wafer
  • pieces down to 3x3 mm2
Allowed materials

All cleanroom materials

Batch

1


Inclined UV Lamp

Feedback to this section: click here

Inclined UV lamp is placed in E-5

The Inclined UV lamp is 1000 W Hg(Xe)lamp source designed for near UV, 350-450nm, mid UV, 260-320nm, and deep UV, 220-260nm exposures of resists and polymers. The exposure source can be also used to make an inclined exposure in air or in the media tank.

The tool was purchased in February 2009 from Newport. The exposure lamp has a official name: Oriel Flood Exposure Source, unit 92540. All other parts of equipment: substrate and mask holder with media tank, exhaust box around the tool, timer controller, were designed and build at DTU Danchip workshop.

The substrate and mask holder with a media tank was designed as part of Master Thesis of DTU Nanotech, Andres Kristensen group. The exhaust box was made as part of safety and the timer controller was build to control exposure time.

The technical specification and the general outline of the equipment can be found in LabManager.

The user manual(s), quality control procedure(s) and results and contact information can be found in LabManager.


Equipment performance and process related parameters

Purpose

UV exposure

Performance Exposure mode
  • Flood exposure
  • Proximity exposure with mask possible for 4inch substrate
Exposure light/filters
  • Near UV(350-450nm)
  • Mid UV (260-320nm)
  • Deep UV(220-260nm)
Minimum structure size
Mask size

5x5inch optinal

Alignment modes

No alignment possible

Substrates Substrate size

Up to 8inch substrates, different shapes

Allowed materials

All cleanroom materials

Batch

1