The Aligner-6inch has a long pass filter designed for SU-8 exposure installed. The SU-8 filter has no transmission in the 250-330nm range, and close to full transmission in the 400-500nm range. In the 350-370nm range, the transmission is approximately 0.5. This results in a broadband exposure light consisting of the i-line (365nm), the h-line (405nm), and the g-line (435nm) from the Hg spectrum. Dependent on the spectral sensitivity of the resist, the optimal dose may be decreased compared to i-line exposure on the KS-Aligner.
The Aligner-6inch has a long pass filter designed for SU-8 exposure installed. The SU-8 filter has no transmission in the 250-330nm range, and close to full transmission in the 400-500nm range. In the 350-370nm range, the transmission is approximately 0.5. This results in a broadband exposure light consisting of the i-line (365nm), the h-line (405nm), and the g-line (435nm) from the Hg spectrum. Dependent on the spectral sensitivity of the resist, the optimal dose may be decreased compared to i-line exposure on the KS-Aligner.


*[[Specific Process Knowledge/Lithography/UVExposure/UVExposure dose|Information on UV exposure dose]]
*[[Specific Process Knowledge/Lithography/UVExposure_Dose|Information on UV exposure dose]]