Specific Process Knowledge/Characterization/AFM: Atomic Force Microscopy/Workspaces: Difference between revisions

Bghe (talk | contribs)
Bghe (talk | contribs)
Line 19: Line 19:
!Recommended mode
!Recommended mode
|Roughness measurement can be done in both Contact, Tapping and ScanAsyst mode. We recommend '''ScanAsyst mode''' because of less chance of (non-uniform) deformation of the sample and less tip wear - and because of ease of use.
|Roughness measurement can be done in both Contact, Tapping and ScanAsyst mode. We recommend '''ScanAsyst mode''' because of less chance of (non-uniform) deformation of the sample and less tip wear - and because of ease of use.
|For Topographic measurements with no steep/abrupt steps we also recommend '''ScanAsyst mode''' due to less chance of (non-uniform) deformation of the sample and less tip wear - and because of ease of use.
|For Topographic measurements with no steep/abrupt steps we also recommend '''ScanAsyst mode''' due to less chance of (non-uniform) deformation of the sample and less tip wear - and because of ease of use.  
|For Steep/abrupt steps but no high aspect ratio structures we still recommend '''ScanAsyst mode''' due to ease of use but it will scan slower than Tapping mode.  
|For Steep/abrupt steps but no high aspect ratio structures we still recommend '''ScanAsyst mode''' due to ease of use but it will scan slower than Tapping mode.  
|For High aspect ratio measurements we recommend Tapping mode. There is not yet a ScanAsyst probe developed for high aspect ratio. Therefor we recommend '''Tapping mode'''.
|For High aspect ratio measurements we recommend Tapping mode. There is not yet a ScanAsyst probe developed for high aspect ratio. Therefore we recommend '''Tapping mode'''.
|-
|-


Line 49: Line 49:
|}
|}


*By choosing the work space QNM in air (Mechanical properties - ?? - QNM in air) ScanAsyst is included and so are the nanomechanical properties like modulus, adhesion, dissipation and deformation