Jump to content

Specific Process Knowledge/Lithography/UVExposure: Difference between revisions

Taran (talk | contribs)
Taran (talk | contribs)
Line 290: Line 290:


===Process information===
===Process information===
The Aligner-6inch has a long pass filter designed for SU-8 exposure installed. The SU-8 filter has no transmission in the 250-330nm range, and close to full transmission in the 400-500nm range. In the 350-370nm range, the transmission is approximately 0.5. This results in a broadband exposure light consisting of the i-line (365nm), the h-line (405nm), and the g-line (435nm) from the Hg spectrum. Dependent on the spectral sensitivity of the resist, the optimal dose may be decreased compared to i-line exposure on the KS-Aligner.
'''Positive tone resists:'''
'''Positive tone resists:'''