Line 185: Line 185:


'''AZ 5214E and AZ 4562'''
'''AZ 5214E and AZ 4562'''
Compared to exposure on Aligner-6inch, the dose on KS-Aligner is doubled.
Compared to exposure on Aligner-6inch, the dose on KS-Aligner is doubled.
*1.5µm: 35-49 mJ/cm<sup>2</sup> corresponding to 5-7 seconds exposure at 7 mW/cm<sup>2</sup>.
*1.5µm: 35-49 mJ/cm<sup>2</sup> corresponding to 5-7 seconds exposure at 7 mW/cm<sup>2</sup>.
*2.2µm: 56-70 mJ/cm<sup>2</sup> corresponding to 8-10 seconds exposure at 7 mW/cm<sup>2</sup>.
*4.2µm: ~140 mJ/cm<sup>2</sup> corresponding to ~20 seconds exposure at 7 mW/cm<sup>2</sup>.
*10µm: ~280 mJ/cm<sup>2</sup> corresponding to ~40 seconds exposure at 7 mW/cm<sup>2</sup>.


'''AZ MiR 701'''
'''AZ MiR 701'''