Specific Process Knowledge/Lithography/Aligners/MAvsMLA: Difference between revisions
| Line 19: | Line 19: | ||
==AZ 5214E== | ==AZ 5214E== | ||
===Resolution=== | ===Resolution=== | ||
'''Resolution at optimal conditions for different exposure equipment:''' | '''Resolution at optimal processing conditions for different exposure equipment:''' | ||
{| class="wikitable" | {| class="wikitable" | ||
|+ | |+ | ||
| Line 113: | Line 113: | ||
===Yield=== | ===Yield=== | ||
The yield is measured by inspecting all 116 lithographic resolution test structures in an optical microscope and recording the smallest resolved dots and lines. | The yield is measured by inspecting all 116 lithographic resolution test structures in an optical microscope and recording the smallest resolved dots and lines. | ||
<br clear="all" /> | <br clear="all" /> | ||
| Line 188: | Line 189: | ||
===Resist profile=== | ===Resist profile=== | ||
====Linewidth and sidewall angle vs exposure dose==== | ====Linewidth and sidewall angle vs exposure dose==== | ||
'''SEM images of 2µm triplets as a function of exposure dose:''' | '''SEM images of 2µm triplets as a function of exposure dose:''' | ||
{|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" | {|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" | ||
| Line 214: | Line 217: | ||
|} | |} | ||
<br clear="all" /> | |||
{|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" | {|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" | ||
| Line 238: | Line 241: | ||
|} | |} | ||
<br clear="all" /> | |||
'''Result of image analysis of SEM images of 2µm triplets:''' | '''Result of image analysis of SEM images of 2µm triplets:''' | ||
| Line 245: | Line 248: | ||
! width="510" | | ! width="510" | | ||
|+ | |+ | ||
|- border="0" align="center" | |- border="0" align="center" | ||
|[[Image:5214E_Linewidth vs dose.jpg|500px]] | |[[Image:5214E_Linewidth vs dose.jpg|500px]] | ||
| Line 251: | Line 254: | ||
|- align="center" | |- align="center" | ||
| colspan="2" | Blue circles: Exposed on Aligner: MA6-2 in vacuum contact. <br>Orange Squares: Exposed on Aligner: Maskless 02 (MLA2) using defoc 2 (QC defoc=2). | | colspan="2" | Width of 2µm line and sidewall angle vs. exposure dose for 1.5µm AZ 5214E. <br>Blue circles: Exposed on Aligner: MA6-2 in vacuum contact. <br>Orange Squares: Exposed on Aligner: Maskless 02 (MLA2) using defoc 2 (QC defoc=2). | ||
|} | |} | ||
<br clear="all" /> | <br clear="all" /> | ||
====Linewidth and sidewall angle vs defocus parameter==== | |||
'''SEM images as a function of defocus parameter for Aligner: Maskless 02 (MLA2) using dose 95mJ/cm<sup>2</sup>:''' | '''SEM images as a function of defocus parameter for Aligner: Maskless 02 (MLA2) using dose 95mJ/cm<sup>2</sup>:''' | ||
{|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" | {|border="1" cellspacing="0" cellpadding="3" style="text-align:left;" | ||
| Line 290: | Line 293: | ||
|} | |} | ||
<br clear="all" /> | |||
'''Result of image analysis of SEM images of 2µm triplets:''' | '''Result of image analysis of SEM images of 2µm triplets:''' | ||
| Line 296: | Line 299: | ||
! width="510" | | ! width="510" | | ||
|+ | |+ | ||
|- border="0" align="center" | |- border="0" align="center" | ||
|[[Image:5214E_Profile vs defoc.jpg|500px]] | |[[Image:5214E_Profile vs defoc.jpg|500px]] | ||
|- align="center" | |- align="center" | ||
| Blue circles: Width of 2µm line. <br>Orange Squares: Sidewall angle. | | Linewidth and sidewall angle vs. defocus value for 1.5µm AZ 5214E exposed in Aligner: Maskless 02 (MLA2) using dose 95mJ/cm<sup>2</sup>. <br>Blue circles: Width of 2µm line. <br>Orange Squares: Sidewall angle. | ||
|} | |} | ||
Latest revision as of 16:24, 15 June 2026
The content on this page, including all images and pictures, was created by DTU Nanolab staff, unless otherwise stated.
Feedback to this page: click here
Mask Aligner vs Maskless Aligner
Slides presented at NNUM in Uppsala 2026: MA6 vs MLA 2026_v09
Mask design used in the investigation: LithoTestAlign_2025_v10.gds
Additional data
THIS PAGE IS UNDER CONSTRUCTION
AZ 5214E
Resolution
Resolution at optimal processing conditions for different exposure equipment:
| Exposure equipment | Dose | Contact/Defoc | Resolution | Comment |
|---|---|---|---|---|
| Aligner: MA6-2 | 100 | Vacuum | 1µm | |
| Aligner: MA6-1 | 100 | Vacuum | 1.25µm | Probably over-exposed |
| Aligner: Maskless 02 (MLA2) | 90 | 2 | 1.25µm |
SEM images from different exposure equipment:
| 2µm triplets | Lines | Trenches | |
|---|---|---|---|
| Aligner: MA6-2
100mJ/cm2, vacuum contact |
|||
| Aligner: MA6-1
100mJ/cm2, vacuum contact |
|||
| Aligner: Maskless 02 (MLA2)
90mJ/cm2, defoc 2 |
SEM images of different size structures:
| 1µm triplets | 2µm triplets | 4µm triplets | |
|---|---|---|---|
| Aligner: MA6-1
100mJ/cm2, vacuum contact |
Yield
The yield is measured by inspecting all 116 lithographic resolution test structures in an optical microscope and recording the smallest resolved dots and lines.
Effect of exposure mode for mask aligner
Effect of WEC pressure for mask aligner (after WEC head service)
Resist profile
Linewidth and sidewall angle vs exposure dose
SEM images of 2µm triplets as a function of exposure dose:
| 85mJ/cm2 | 90mJ/cm2 | 95mJ/cm2 | 100mJ/cm2 | 105mJ/cm2 | |
|---|---|---|---|---|---|
| Aligner: MA6-2
Vacuum contact |
| 80mJ/cm2 | 90mJ/cm2 | 100mJ/cm2 | 110mJ/cm2 | |
|---|---|---|---|---|
| Aligner: Maskless 02 (MLA2)
Defoc 2 |
Result of image analysis of SEM images of 2µm triplets:
Linewidth and sidewall angle vs defocus parameter
SEM images as a function of defocus parameter for Aligner: Maskless 02 (MLA2) using dose 95mJ/cm2:
| Defoc -8 | Defoc -1 | Defoc 2 | Defoc 5 | Defoc 12 | |
|---|---|---|---|---|---|
| 2µm triplets | |||||
| Lines |
Result of image analysis of SEM images of 2µm triplets:
| Linewidth and sidewall angle vs. defocus value for 1.5µm AZ 5214E exposed in Aligner: Maskless 02 (MLA2) using dose 95mJ/cm2. Blue circles: Width of 2µm line. Orange Squares: Sidewall angle. |
AZ MiR 701
Resolution
- Table with optimal parameters and resolution for different aligners
- SEM pictures
Resist profile (sidewall angle and linewidth)
- SEM pictures to support graphs?

Blue circles: Exposed on Aligner: MA6-2 in vacuum contact.
Orange Squares: Exposed on Aligner: Maskless 02 (MLA2) using defoc 2 (QC defoc=2).

Blue circles: Exposed on Aligner: MA6-2 in vacuum contact.
Orange Squares: Exposed on Aligner: Maskless 02 (MLA2) using defoc 2 (QC defoc=2).

Blue circles: Width of 2µm line. Orange Squares: Sidewall angle.
AZ nLOF 2020
Resolution
- Table with optimal parameters and resolution for different aligners
- SEM pictures
Resist profile (sidewall angle and linewidth)
- SEM pictures to support graphs?

Blue circles: Exposed on Aligner: MA6-2 in vacuum contact.
Orange Squares: Exposed on Aligner: Maskless 01 (MLA1) using defoc -4 (QC defoc=-2).
Grey triangles: Exposed on Aligner: Maskless 02 (MLA2) using defoc 0 (QC defoc=2).

Blue circles: Exposed on Aligner: MA6-2 in vacuum contact.
Orange Squares: Exposed on Aligner: Maskless 01 (MLA1) using defoc -4 (QC defoc=-2).
Grey triangles: Exposed on Aligner: Maskless 02 (MLA2) using defoc 0 (QC defoc=2).

Blue circles: Width of 2µm line. Orange Squares: Sidewall angle.

Blue circles: Width of 2µm line. Orange Squares: Sidewall angle.