Created page with "==From February 2024: RESIST PYROLYSIS FURNACE== '''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/Furnace:_Multipurpose_annealing#Multipurpose_annealing_furnace click here]''' ''This page is written by DTU Nanolab internal'' Thermal A2 Furnace A2 ==The..."
 
Pevo (talk | contribs)
 
(14 intermediate revisions by the same user not shown)
Line 1: Line 1:
==From February 2024: RESIST PYROLYSIS FURNACE==


'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/Furnace:_Multipurpose_annealing#Multipurpose_annealing_furnace click here]'''
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/Specific_Process_Knowledge/Thermal_Process/Furnace:_Multipurpose_annealing#Multipurpose_annealing_furnace click here]'''
Line 11: Line 9:
[[Category: Furnaces|A2]]
[[Category: Furnaces|A2]]


==The Multipurpose annealing furnace==
==The Resist Pyrolysis furnace (research tool)==


[[Image:ATV_ovn.jpg|thumb|365x365px|Multipurpose Annealing Furnace. Positioned in cleanroom B-1. Photo: DTU Nanolab internal]]
[[Image:ATV_ovn.jpg|thumb|365x365px|Resist Pyrolysis furnace. Positioned in cleanroom B-1. Photo: DTU Nanolab internal]]
[[Image:ATV_ovn_boat.jpg|thumb|335x335px|Boat with wafers for the Multipurpose Annealing Furnace. Photo: DTU Nanolab internal]]
[[Image:ATV_ovn_boat.jpg|thumb|335x335px|Boat with wafers for the Resist Pyrolysis furnace (from an oxidation test). Photo: DTU Nanolab internal]]


The Multipurpose Annealing Furnace it made by ATV Technologie, and it was installed in the cleanroom in 2015.
From February 2024: The name of the furnace has changed from "Multipurpose Anneal furnace" to "Resist Pyrolysis furnace" (or “Furnace: Resist Pyrolysis” in LabManager). At the same time, the furnace has become a research tool. This means, that the furnace is mainly available for users working on certain research projects at DTU Nanolab, and it should mainly be used for resist pyrolysis


The purpose of the Multipurpose Annealing Furnace is annealing and dry oxidation of different samples and resist pyrolysis. Annealing and resist pyrolysis can be done in vacuum or at atmospheric pressure, in a N<sub>2</sub> or H<sub>2</sub> atmosphere or a mixture of the two gasses.  
The Resist Pyrolysis furnace it made by ATV Technologie, and it was installed in the cleanroom in 2015.  


All process gasses (except purge nitrogen) are heated, before they are introduced into the furnace at the door side.
The furnace is mainly being used for resist pyrolysis. It can also be used for annealing in nitrogen (N<sub>2</sub>) and oxidation with oxygen (O<sub>2</sub>), but O<sub>2</sub> is mostly used for cleaning of the furnace after resist pyrolysis.


Is is possible to change all quartz ware in the furnace (the furnace tube, the door sealing and the wafer boat). At the moment Nanolab has two different sets of quartz ware:
Process steps can run either in vacuum or at atmospheric pressure. A small N<sub>2</sub> or O<sub>2</sub> flow (up to 2 slm) can be applied in vacuum steps.  


*Metal: Dedicated for different samples that cannot be RCA cleaned. Also samples with metals are allowed in the furnace, when this quartz set is mounted
The quartz parts in the furnace can be changed, depending on what processes and samples the furnace is used for, but it has been decided that a quartz set for resist resist pyrolysis will be mounted in the furnace all the time. This limits the number of materials that are allowed in the furnace.
*Resist pyrolysis: Dedicated for resist pyrolysis.  


Please note that all new materials have to be approved by the Thin Film group (thinfilm@nanolab.dtu.dk) before they are allowed in the furnace.  
All process gases (except purge nitrogen) are heated, before they are introduced into the furnace at the door side.


The furnace tube is heated by use of 12 long heaters situated along the furnace tube and combined in three groups (top, bottom left and bottom right) and two flat heaters situated in the ends of the furnace tube. The this way the temperature will be very uniform everywhere in the furnace tube. The heating can be done very fast, up to 30 <sup>o</sup>C/min. For atmospheric pressure processes the maximum temperature is 1100 <sup>o</sup>C, and for vacuum processes the maximum temperature is 1050 <sup>o</sup>C.  
The furnace tube is heated by use of 12 long heaters situated along the furnace tube and combined in three groups (top, bottom left and bottom right) and two flat heaters situated in the ends of the furnace tube. The this way the temperature will be very uniform everywhere in the furnace tube. However, the front heater is unfortunately broken at the moment. The heating can be done very fast, up to 30 <sup>o</sup>C/min. The maximum allowed processing time and temperature is 3 hours at 1050 <sup>o</sup>C.


The furnace body surrounding the furnace tube consists of a top and bottom half-shell. To cool down the furnace (when the temperature is below 800 <sup>o</sup>C), the top half shelf can be lifted up, and cooling fans will then flow air from the surroundings around the furnace tube to cool it down. However, for safety reasons, the cooling fans will not the activated when there is hydrogen in the furnace, and cooling will then be very slow.  
The furnace body surrounding the furnace tube consists of a top and bottom half-shell. To cool down the furnace (when the temperature is below 800 <sup>o</sup>C), the top half shelf can be lifted up, and cooling fans will then flow air from the surroundings around the furnace tube to cool it down.  


It is not possible to open the furnace when the temperature is above 300 <sup>o</sup>C.
It is not possible to open the furnace, when the temperature is above 300 <sup>o</sup>C.


For resist pyrolysis, samples with different resist layers are heated up to maximum 1100 <sup>o</sup>C in a nitrogen atmosphere. At high temperatures carbon is formed by pyrolysis of the resist. In this way conductive structures can be made from a resist patterned sample. If oxygen from the air or from outgassing of the resist is present in the furnace, the resist layer will be removed, thus it is important to evacuate the furnace and flush it with nitrogen, before a high temperature for resist pyrolysis is obtained. Pyrolysis of a large amount of resist may also be a problem due to resist outgassing.  
For resist pyrolysis, samples with different resist layers are heated up to maximum 1050 <sup>o</sup>C in a nitrogen atmosphere. At high temperatures carbon is formed by pyrolysis of the resist. In this way conductive structures can be made from a resist patterned sample. If oxygen from the air or from outgassing of the resist is present in the furnace, the resist layer will be removed, thus it is important to evacuate the furnace and flush it with nitrogen, before a high temperature for resist pyrolysis is obtained. Pyrolysis of a large amount of resist may also be a problem due to resist outgassing.  




'''The user manual, technical information and contact information can be found in LabManager:'''
'''The user manual, technical information and contact information can be found in LabManager:'''


'''[http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=340 Furnace: Multipurpose annealing]'''
'''[http://labmanager.dtu.dk/function.php?module=Machine&view=view&mach=340 Furnace: Resist Pyrolysis]'''
 


==Process information==
==Process information==
*[[Specific_Process_Knowledge/Thermal_Process/Resist_Pyrolysis_furnace/Acceptance test|Results from the Resist Pyrolysis (former Multipurpose Anneal) furnace acceptance test]]
*[[Specific_Process_Knowledge/Thermal_Process/Resist_Pyrolysis_furnace/Pyrolysis_with_Resist_Pyrlysis_furnace|Resist pyrolysis]]
*Annealing: look at the [[Specific Process Knowledge/Thermal Process/Annealing|Annealing]] page
*Oxidation: look at the [[Specific Process Knowledge/Thermal Process/Oxidation|Oxidation]] page
*Oxidation: look at the [[Specific Process Knowledge/Thermal Process/Oxidation|Oxidation]] page
*Annealing: look at the [[Specific Process Knowledge/Thermal Process/Annealing|Annealing]] page
*[[Specific_Process_Knowledge/Thermal_Process/Furnace:_Multipurpose_annealing/Acceptance test|Results from the Multipurpose Anneal Furnace acceptance test]]
*[[Specific_Process_Knowledge/Thermal_Process/Pyrolysis/Pyrolysis_with_Multipurpose_Anneal_Furnace|Pyrolysis]]


==Overview of the performance of the ATV furnace and process related parameters==
==Overview of the performance of the Resist Pyrolysis furnace and process related parameters==


{| border="2" cellspacing="3" cellpadding="5"  
{| border="2" cellspacing="3" cellpadding="5"  
|-
|-
!style="background:silver; color:black;" align="center"|Purpose  
!style="background:silver; color:black;" align="center"|Purpose  
|style="background:LightGrey; color:black"|
|style="background:LightGrey; color:black"|Processes
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Pyrolysis of different resists
*Annealing in N<sub>2</sub>
*Dry oxidation of silicon  
*Dry oxidation of silicon  
*Annealing in N<sub>2</sub>, H<sub>2</sub> or a mixture of the two gasses
*Pyrolysis of different resists
|-
|-
!style="background:silver; color:black" align="center"|Performance
!style="background:silver; color:black" align="center"|Performance
Line 66: Line 62:
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Process parameter range
!style="background:silver; color:black" align="center" valign="center" rowspan="3"|Process parameter range
|style="background:LightGrey; color:black"|Process Temperature
|style="background:LightGrey; color:black"|Process temperature
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*No vacuum: 25 <sup>o</sup>C - 1100 <sup>o</sup>C
*No vacuum: 25 <sup>o</sup>C - 1050 <sup>o</sup>C, max 3 hours at 1050 <sup>o</sup>C
*Vacuum: 25 <sup>o</sup>C - 1050 <sup>o</sup>C  
*Vacuum: 25 <sup>o</sup>C - 1050 <sup>o</sup>C, max 3 hours at 1050 <sup>o</sup>C  
|-
|-
|style="background:LightGrey; color:black"|Process pressure
|style="background:LightGrey; color:black"|Process pressure
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1 atm
*1 atm
*Vacuum down to ~0.1 mbar (depends on gas flow)
*Vacuum down to ~0.1 mbar (depends on the gas flow)
|-
|-
|style="background:LightGrey; color:black"|Gasses on the system
|style="background:LightGrey; color:black"|Gasses on the system
Line 80: Line 76:
*N<sub>2</sub>: 20 slm
*N<sub>2</sub>: 20 slm
*O<sub>2</sub>: 10 slm
*O<sub>2</sub>: 10 slm
*H<sub>2</sub>: 5 slm (max 2 slm for vacuum processes)
*H<sub>2</sub>: 5 slm (max 2 slm for vacuum processes) - Not available
*N<sub>2</sub> mix : 10 slm (for H<sub>2</sub>-N<sub>2</sub> gas mixture)
*N<sub>2</sub> mix : 10 slm (for H<sub>2</sub>-N<sub>2</sub> gas mixture) - Do not use
|-
|-
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
!style="background:silver; color:black" align="center" valign="center" rowspan="2"|Substrates
Line 87: Line 83:
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*1-30 50 mm, 100 mm or 150 mm wafers per run
*1-30 50 mm, 100 mm or 150 mm wafers per run
*1-50 200 mm wafers per run (not possible with all quartz sets)
*Smaller samples (placed on a Si carrier wafer)
*Smaller samples (placed on Si carrier wafers)
|-
|-
|style="background:LightGrey; color:black"|Substrate materials allowed
|style="background:LightGrey; color:black"|Substrate materials allowed
|style="background:WhiteSmoke; color:black"|
|style="background:WhiteSmoke; color:black"|
*Depends on the quartz set
*Only samples for resist pyrolysis, and all sample materials have to be approved by DTU Nanolab. Samples with metals and III-V materials are NOT allowed
|-  
|-  
|}
|}

Latest revision as of 13:57, 21 March 2024

Feedback to this page: click here

This page is written by DTU Nanolab internal

The Resist Pyrolysis furnace (research tool)

Resist Pyrolysis furnace. Positioned in cleanroom B-1. Photo: DTU Nanolab internal
Boat with wafers for the Resist Pyrolysis furnace (from an oxidation test). Photo: DTU Nanolab internal

From February 2024: The name of the furnace has changed from "Multipurpose Anneal furnace" to "Resist Pyrolysis furnace" (or “Furnace: Resist Pyrolysis” in LabManager). At the same time, the furnace has become a research tool. This means, that the furnace is mainly available for users working on certain research projects at DTU Nanolab, and it should mainly be used for resist pyrolysis

The Resist Pyrolysis furnace it made by ATV Technologie, and it was installed in the cleanroom in 2015.

The furnace is mainly being used for resist pyrolysis. It can also be used for annealing in nitrogen (N2) and oxidation with oxygen (O2), but O2 is mostly used for cleaning of the furnace after resist pyrolysis.

Process steps can run either in vacuum or at atmospheric pressure. A small N2 or O2 flow (up to 2 slm) can be applied in vacuum steps.

The quartz parts in the furnace can be changed, depending on what processes and samples the furnace is used for, but it has been decided that a quartz set for resist resist pyrolysis will be mounted in the furnace all the time. This limits the number of materials that are allowed in the furnace.

All process gases (except purge nitrogen) are heated, before they are introduced into the furnace at the door side.

The furnace tube is heated by use of 12 long heaters situated along the furnace tube and combined in three groups (top, bottom left and bottom right) and two flat heaters situated in the ends of the furnace tube. The this way the temperature will be very uniform everywhere in the furnace tube. However, the front heater is unfortunately broken at the moment. The heating can be done very fast, up to 30 oC/min. The maximum allowed processing time and temperature is 3 hours at 1050 oC.

The furnace body surrounding the furnace tube consists of a top and bottom half-shell. To cool down the furnace (when the temperature is below 800 oC), the top half shelf can be lifted up, and cooling fans will then flow air from the surroundings around the furnace tube to cool it down.

It is not possible to open the furnace, when the temperature is above 300 oC.

For resist pyrolysis, samples with different resist layers are heated up to maximum 1050 oC in a nitrogen atmosphere. At high temperatures carbon is formed by pyrolysis of the resist. In this way conductive structures can be made from a resist patterned sample. If oxygen from the air or from outgassing of the resist is present in the furnace, the resist layer will be removed, thus it is important to evacuate the furnace and flush it with nitrogen, before a high temperature for resist pyrolysis is obtained. Pyrolysis of a large amount of resist may also be a problem due to resist outgassing.


The user manual, technical information and contact information can be found in LabManager:

Furnace: Resist Pyrolysis

Process information

Overview of the performance of the Resist Pyrolysis furnace and process related parameters

Purpose Processes
  • Pyrolysis of different resists
  • Annealing in N2
  • Dry oxidation of silicon
Performance Film thickness
  • Dry oxidation: 50 Å to ~200 nm SiO2 (it takes too long to grow a thicker oxide layer)
Process parameter range Process temperature
  • No vacuum: 25 oC - 1050 oC, max 3 hours at 1050 oC
  • Vacuum: 25 oC - 1050 oC, max 3 hours at 1050 oC
Process pressure
  • 1 atm
  • Vacuum down to ~0.1 mbar (depends on the gas flow)
Gasses on the system
  • N2: 20 slm
  • O2: 10 slm
  • H2: 5 slm (max 2 slm for vacuum processes) - Not available
  • N2 mix : 10 slm (for H2-N2 gas mixture) - Do not use
Substrates Batch size
  • 1-30 50 mm, 100 mm or 150 mm wafers per run
  • Smaller samples (placed on a Si carrier wafer)
Substrate materials allowed
  • Only samples for resist pyrolysis, and all sample materials have to be approved by DTU Nanolab. Samples with metals and III-V materials are NOT allowed