==Pegasus 3 - 150mm silicon etching==
==Pegasus 3 - 150mm silicon etching==
 
{{Contentbydryetch}}
 
[[Image:05x.jpg |frame|left|x300px|The Pegasus 3 cassette station (left), operator station (center) and support computer with Picoscope process monitoring.{{photo1}} ]]
 
<span style="background:#FF2800">THIS PAGE IS UNDER CONSTRUCTION</span>[[image:Under_construction.png|300px]]
 
'''The tool is ready for user training. '''
 
[[Image:05x.jpg |frame|left|x300px|The Pegasus 3 cassette station (left), operator station (center) and support computer with Picoscope process monitoring. ]]




Line 24: Line 18:




'''[[Specific Process Knowledge/Etch/DRIE-Pegasus 3/StandardRecipes|Standard recipes]]'''
'''Standard recipes'''
*[[Specific Process Knowledge/Etch/DRIE-Pegasus/Pegasus-3/DREM | The DREM recipes 0.5 kW and 0.2 kW ]]
 
 


*[[Specific Process Knowledge/Etch/DRIE-Pegasus 3/Barc etch |Barc etch]]


'''Hardware changes'''
'''Hardware changes'''
Line 32: Line 28:
Compared to a standard SPTS DRIE Pegasus chamber, the Pegasus 3 has had the same modifications as the Pegasus 1 with the exception of the Claritas EPD system. The changes are listed below.
Compared to a standard SPTS DRIE Pegasus chamber, the Pegasus 3 has had the same modifications as the Pegasus 1 with the exception of the Claritas EPD system. The changes are listed below.


*[[Specific Process Knowledge/Etch/DRIE-Pegasus/showerheadchange|Change of Pegasus 1 showerhead in 2019]]
*[[Specific Process Knowledge/Etch/DRIE-Pegasus/showerheadchange|High flow plenum upgrade]]
*[[Specific Process Knowledge/Etch/DRIE-Pegasus/picoscope|Addition of Picoscope oscilloscope system for process monitoring]]


*[[Specific Process Knowledge/Etch/DRIE-Pegasus/picoscope|Addition of Picoscope oscilloscope system for process monitoring on Pegasus 1 in February 2019]]
The original (fast switching and with a maximum flow of 1200 sccm) SF<sub>6</sub> MFC was causing problems for a long a time so it was replaced by two new MFC's:
* SF6-1: Fast switching MFC for Bosch processes with a maximum flow rate of 600 sccm. It sits in the position of the old 1200 sccm MFC on top of the plasma source.
* SF6-2: Normal MFC for continuous processes with a maximum flow rate of 100 sccm. It is installed in the gas cabinet next to the Ar MFC.




=== Other etch processes ===
=== Other etch processes ===


*[[Specific Process Knowledge/Etch/DRIE-Pegasus/nanoetch|Continuous nanostructure etches including nano1.42]]
*[[Specific Process Knowledge/Etch/DRIE-Pegasus/DUVetch|Etch processes with DUV masks]]
*[[Specific Process Knowledge/Etch/DRIE-Pegasus/Barc|BARC etches]]
*[[Specific Process Knowledge/Etch/DRIE-Pegasus/Isotropic|Isotropic etches]]


More processes, such as for DUV resist, are currently being developed, but they are not quite 'ready for publication' at LabAdviser so please contact Jonas (mailto:jmli@dtu.dk) for more information.
More processes, such as for DUV resist, are currently being developed, but they are not quite 'ready for publication' at LabAdviser so please contact Jonas (mailto:jmli@dtu.dk) for more information.
Line 54: Line 46:


  <!-- *[[Specific Process Knowledge/Etch/DRIE-Pegasus/VeeryDeeep| Very deep etching]] -->
  <!-- *[[Specific Process Knowledge/Etch/DRIE-Pegasus/VeeryDeeep| Very deep etching]] -->
*[[/SiO2 etch|Etch of very thin layers of SiO2]]


=== Wafer bonding ===
=== Wafer bonding ===

Latest revision as of 12:50, 25 January 2024

Feedback to this page: click here


Pegasus 3 - 150mm silicon etching

Unless otherwise stated, the content of this page was created by the dry etch group at DTU Nanolab

The Pegasus 3 cassette station (left), operator station (center) and support computer with Picoscope process monitoring.Photo: DTU Nanolab internal


The user manual(s) is available, technical information and contact information can be found in LabManager:

Equipment info in LabManager

Process information

Standard recipes



Hardware changes

Compared to a standard SPTS DRIE Pegasus chamber, the Pegasus 3 has had the same modifications as the Pegasus 1 with the exception of the Claritas EPD system. The changes are listed below.

The original (fast switching and with a maximum flow of 1200 sccm) SF6 MFC was causing problems for a long a time so it was replaced by two new MFC's:

  • SF6-1: Fast switching MFC for Bosch processes with a maximum flow rate of 600 sccm. It sits in the position of the old 1200 sccm MFC on top of the plasma source.
  • SF6-2: Normal MFC for continuous processes with a maximum flow rate of 100 sccm. It is installed in the gas cabinet next to the Ar MFC.


Other etch processes

More processes, such as for DUV resist, are currently being developed, but they are not quite 'ready for publication' at LabAdviser so please contact Jonas (mailto:jmli@dtu.dk) for more information.


Wafer bonding

To find information on how to bond wafers or chips to a carrier wafer, click here.

Internal Nanolab Process log for Pegasus 3

Process log at Nanolab [1]