LabAdviser/314/Microscopy 314-307/TEM/Tips/GIF-apertures: Difference between revisions
No edit summary |
|||
(3 intermediate revisions by the same user not shown) | |||
Line 1: | Line 1: | ||
'''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/Microscopy_314-307/ | '''Feedback to this page''': '''[mailto:labadviser@nanolab.dtu.dk?Subject=Feed%20back%20from%20page%20http://labadviser.nanolab.dtu.dk/index.php/LabAdviser/314/Microscopy_314-307/TEM/Tips/GIF-apertures click here]''' | ||
(''content by Jens Kling, March 2020'') | (''content by Jens Kling @DTU Nanolab, March 2020'') | ||
[[Category:314]] | [[Category:314]] | ||
[[Category:314-Microscopy]] | [[Category:314-Microscopy]] | ||
Line 21: | Line 21: | ||
[[File:GIF-apertures.png|300px| | [[File:GIF-apertures.png|300px|left|thumb|Rack behind the microscope PC]] | ||
<br clear="all" /> | <br clear="all" /> |
Latest revision as of 14:26, 27 June 2023
Feedback to this page: click here
(content by Jens Kling @DTU Nanolab, March 2020)
GIF apertures are not responding
From time to time, there might be some issues with the GIF apertures not responding and moving. If this is the case, try the following:
- Close the software "Filter Control" and "DM" on the microscope PC.
- Go to the cabinet back of the microscope, where the controlers for the CCD cameras and GIF are (see image).
- Switch off the main power supply at the bottom (GIF electronics; right above the OneView controler).
- Wait 10s
- Switch the power back on.
- Make sure the two CCD controlers don't show an error (orange LED above RESET); if they do, press "Reset").
- Start "Filter Control" and "DM" on the microscope PC.