Specific Process Knowledge/Characterization/AFM: Atomic Force Microscopy/Workspaces: Difference between revisions
Appearance
| Line 72: | Line 72: | ||
=Evaluation of used probes= | =Evaluation of used probes= | ||
This evaluation was done to know if the tip end of the ScanAsyst probes was typically dirty or broken when we got bad images. If it was dirty, maybe we could find a way to clean it. If it was broken off we needed to replace it. It turned out it was in most cases broken. | |||
{| border="1" cellspacing="1" cellpadding="2" align="left" | {| border="1" cellspacing="1" cellpadding="2" align="left" | ||
! Box number and probe number | ! Box number and probe number | ||