==Accessories following the systems ==
==Accessories following the systems ==
{| border="2" cellspacing="0" cellpadding="2" valign="top"
{| border="2" cellspacing="0" cellpadding="2" valign="top"
Line 191: Line 195:
Phase: Phase imaging maps the phase lag between the periodic signal driving the cantilever and the oscillations of the cantilever. Changes in phase lag often indicate changes in the properties of the sample surface. Here the structuring in the graphene is very clear
Phase: Phase imaging maps the phase lag between the periodic signal driving the cantilever and the oscillations of the cantilever. Changes in phase lag often indicate changes in the properties of the sample surface. Here the structuring in the graphene is very clear


==AFM Icon-Pt 2: Acceptance test [[Image:section under construction.jpg|70px]]==
==AFM Icon-Pt 2: Acceptance test==
===Noise tests===
===Noise tests===
====Sensor noise====
====Sensor noise====
Line 271: Line 275:
====Tapping mode with TAP150A====
====Tapping mode with TAP150A====
*CL on
*CL on
*256*32
*256x32
*0dg
*0dg
*Step height: 1545/1532 nm
*Step height: 1545/1532 nm
[[File:SiO2-resist step Tapping mode with TAP150A.JPG|400px]]
====ScanAsyst with TAP150A====
====ScanAsyst with TAP150A====
*CL on
*CL on
Line 282: Line 287:
*Peak force frequency: 2 Hz
*Peak force frequency: 2 Hz
*Step height: 1537 nm
*Step height: 1537 nm
[[File:SiO2-resist step ScanAsyst mode with TAP150A.JPG|400px]]
====ScanAsyst with ScanAsyst-air====
====ScanAsyst with ScanAsyst-air====
*CL on
*CL on
Line 290: Line 296:
*Peak force frequency: 2 Hz
*Peak force frequency: 2 Hz
*Step height: 1545 nm
*Step height: 1545 nm
[[File:SiO2-resist step ScanAsyst mode with ScanAsyst-air.JPG|400px]]
====Tapping mode with RFESP-75====
====Tapping mode with RFESP-75====
*CL on
*CL on
Line 295: Line 302:
*0dg
*0dg
*Step height: 1525 nm
*Step height: 1525 nm
[[File:SiO2-resist step Tapping mode with RFESP-75.JPG|400px]]
===High Aspect ratio sample===
High aspect ratio samples (e.g. 100 nm wide trenches 300 nm deep and 2 µm wide trenches 6µm deep).  The probes for this: FIB6-400A.
This was done and work spaces were created but no images was saved.