Specific Process Knowledge/Lithography/UVExposure/aligner MA6-2
Aligner: MA6-2
Tool description
The Aligner: MA6-2 is a manual UV exposure system, which can be used for mask alignment and exposure of UV resists on chips, 50 mm, 100 mm, and 150 mm substrates.
The Süss MicroTek Mask Aligner MA6 is designed for high resolution photolithography. The 365nm exposure wavelength version is capable of 1.25 (1.0) um resolution in vacuum contact. All contact exposure programs (vacuum, low vacuum, hard, soft, proximity) are supplied. Two alignment options are available: top side alignment (TSA) with a split field or a video microscope and back side alignment (BSA) with BSA (video) microscope.
The aligner can also be used for bond alignment (for Wafer Bonder 02). Special training is required.
| Product: | Süss mask aligner MA6 |
|---|---|
| Year of purchase: | 2014 |
| Location: | Cleanroom E-4 |
User manual
The user manual and contact information can be found in LabManager - requires login
Tool training
Training on the tool requires users to complete the lithography TPT followed by the online tool training and a hands-on authorization training.
The tool training videos are part of the online tool training, but can also be viewed here (Operation) and here (Alignment).
| Purpose |
|
|---|---|
| Exposure modes |
|
| Exposure light/filters |
|
| Minimum resolution | ~1.25 µm |
| Mask size |
|
| Alignment modes |
|
| Substrate size |
|
| Allowed materials |
|
| Substrate batch | 1 |
Process information
The Aligner: MA6-2 has an i-line notch filter installed. This results in an exposure light peak at 365nm. Compared to exposure on the KS Aligner, the optimal dose should be very similar. The 500W Hg-Xe lamp also enables exposure in the DUV range around 240nm. This functionality is not established yet, partly due to safety concerns.
Exposure dose
Information on UV exposure dose
Alignment mark design and locations
Alignment mark design and locations.
Light intensity and uniformity after lamp ignition
The lamp requires about 15 minutes of on-time after lamp ignition, before the light output has stabilized, and users using the constant power mode should always wait for 15 minutes after ignition, before starting processing.
If users use the constant intensity mode, this waiting time is not necessary, since the internal light sensor automatically adjusts the output to match the setpoint value.
The non-uniformity of the light, in CI mode, is unaffected by the stabilization time.
Alignment
Top Side Alignment:
- TSA microscope standard objectives: 5X, and 10X (20X available)
- TSA microscope special objectives: 11.25X offset (for smaller separation)
- Minimum distance between TSA microscope objectives: 33 mm (8 mm for special objectives)
- Alignment of smaller separations is possible using the "scan microscope" function
- Maximum distance between TSA microscope objectives: 160 mm
- TSA microscope travel range: X +/- 25mm; Y +20mm / -75mm (towards flat)
BackSide Alignment:
- Minimum distance between BSA microscope objectives: 15 mm
- Maximum distance between BSA microscope objectives: 100 mm
- BSA microscope travel range: X +50mm / -16mm; Y +50mm / -20mm (towards flat)
- BSA chuck view ranges:
- 2": X +/- 8-22mm; Y +/- 0-6mm
- 4": X +/- 14-46mm; Y +/- 0-10mm
- 6": X +/- 14-69mm; Y +/- 0-10mm (OBS: Max. separation of BSA microscopes is 100mm)
Microscope field of view (W x H, splitfield):
- TSA 5X
- Oculars: 1.3mm x 2.6mm (Ø2.6mm full field)
- Camera: 350µm x 500µm (700µm x 500µm full field)
- TSA 10X
- Oculars: 0.6mm x 1.3mm (Ø1.3mm full field)
- Camera: 150µm x 250µm (350µm x 250µm full field)
- TSA special
- Oculars: 0.55mm x 1.1mm (Ø1.1mm full field)
- Camera: 150µm x 200µm (300µm x 200µm full field)
- BSA camera
- Low: 1.5mm x 2mm (3mm x 2mm full field)
- High: 450µm x 650µm (950µm x 650µm full field)
Quality Control (QC)
| Quality Control (QC) for Aligner: MA6-2 | ||||||||||||||||
Power supply and/or lamp will be adjusted if intensity is outside the limit. |