Specific Process Knowledge/Lithography/Resist/UVresist/exposureDoseMasklessAligners

From LabAdviser

Exposure dose for maskless aligners

The exposure doses listed below are for generic good exposure results, and can be a compromise between getting good lines, as well as good dots, in both clear field and dark field exposures. The optimal dose and achievable resolution for any given specific project, could be different from the listed values.

All doses are for standard silicon wafers, unless otherwise stated. Development is done using 2.38% TMAH (AZ 726 MIF).


Aligner: Maskless 01

The Aligner: Maskless 01 has a 365 nm laser light source with a FWHM of ~1 nm. The test exposures on this tool were made using the pneumatic autofocus system.

Quality control
Please note that the values listed in the table are not routinely tested as part of quality control, except for the resist AZ 5214E.

Date Operator Film thickness Exposure mode Dose Defoc Resolution Comments
AZ MIR 701 2025-10-16 jehem 1.5 µm Fast 325 mJ/cm2 -1 1.5 µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle

Tested using dehydration reducing measures
AZ 5214E 2025-08-15 taran 1.5 µm Fast 110 mJ/cm2 -1 1.75 µm See QC data for latest values

Development: 60 s puddle
AZ 5214E
Image reversal
2023-04-05 jehem 2.2 µm Fast 20 mJ/cm2
possibly incorrect data
-2
possibly incorrect data
2.0 µm Not under regular quality control

Reversal bake: 60 s @ 110°C
Flood exposure: 500 mJ/cm2
Development: 60 s puddle
AZ nLOF 2020 2023-07-06 jehem 2.0 µm Fast 180 mJ/cm2 0 1.5 µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ 4562 2023-04-05 jehem 10 µm Fast 1050 mJ/cm2
possibly incorrect data
-1
possibly incorrect data
≤5 µm Not under regular quality control

Priming: HMDS
Rehydration after SB: none
Exposure: Single exposure
Degassing after exposure: none
Development: 5 x 60 s multi puddle

New resists version in 2023
As of 2023-03-21 we no longer have any of the old versions of the resists: AZ 5214E, AZ 4562, AZ MiR 701. This table contains only information about the new resist versions.

Dehydration reducing measures used for testing AZ MiR 701:
The CDA used for the pneumatic autofocus will dehydrate the resist. To reduce this effect, the writehead is parked far away from the write area while setting up the job for at least a few minutes, before starting the exposure.


Aligner: Maskless 02

The Aligner: Maskless 02 has a 375 nm laser light source with a FWHM of ~1 nm. The test exposures on this tool were made using the optical autofocus system, unless otherwise specified.

Quality control
Please note that the values listed in the table are not routinely tested as part of quality control, except for the resist AZ 5214E.

Date Operator Film thickness Exposure mode Dose Defoc Resolution Comments
AZ MIR 701 2023-08-21 taran 1.5 µm Quality 325 mJ/cm2 1 1 µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle

Tested using dehydration reducing measures
AZ 5214E 2024-12-09 taran 1.5 µm Quality 100 mJ/cm2 2 1.5 µm See QC data for latest values

Development: 60 s puddle
AZ 5214E
Image reversal
2023-04-17 jehem 2.2 µm Quality 35 mJ/cm2 0 1.5 µm Not under regular quality control

Reversal bake: 60 s @ 110°C
Flood exposure: 500 mJ/cm2
Development: 60 s puddle
AZ nLOF 2020 2023-08-21 taran 2.0 µm Quality 450 mJ/cm2 0 1.5 µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ 4562 2023-04-19 jehem 10 µm Quality 1150 mJ/cm2 0 ≤5 µm Not under regular quality control

Priming: HMDS
Rehydration after SB: none
Exposure: Single exposure
Degassing after exposure: none
Development: 5 x 60 s multi puddle

New writehead
As of 2023-03-21 the MLA2 has a new writehead installed, converting it from a write mode 1 tool to a write mode 2 tool. This makes all previous dose/defocus settings obsolete. This table contains only the dose/defocus values for the new writehead.

New resists version in 2023
As of 2023-03-21 we no longer have any of the old versions of the resists: AZ 5214E, AZ 4562, AZ MiR 701. This table contains only information about the new resist versions.

Dehydration reducing measures used for testing AZ MiR 701:
The CDA used for the pneumatic autofocus will dehydrate the resist. To reduce this effect, the writehead is parked far away from the write area while setting up the job for at least a few minutes, before starting the exposure.


Aligner: Maskless 03

The Aligner: Maskless 03 has a 405 nm laser light source with a FWHM of ~1 nm. The test exposures on this tool were made using the pneumatic autofocus system.

Quality control
Please note that the values listed in the table are not routinely tested as part of quality control, except for the resist AZ 5214E.

Date Operator Film thickness Exposure mode Dose Defoc Resolution Comments
AZ MIR 701 2025-10-20 jehem 1.5 µm Quality 175 mJ/cm2 -1 1 µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle

Tested using dehydration reducing measures
AZ 5214E 2025-09-02 taran 1.5 µm Quality 75 mJ/cm2 1 1 µm See QC data for latest values

Development: 60 s puddle
AZ 5214E
Image reversal
2023-06-30 jehem 2.2 µm Quality 50 mJ/cm2 0 1 µm Not under regular quality control

Reversal bake: 60 s @ 110°C
Flood exposure: 500 mJ/cm2
Development: 60 s puddle
AZ nLOF 2020 2023-06-30 jehem 2 µm Quality 9000 mJ/cm2 2 1 µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle

Due to the high dose required on this tool, it is recommended to process nLOF on tools with a more appropriate exposure light source
AZ 4562 2023-06-30 jehem 10 µm Quality 550 mJ/cm2 3 ≤5 µm Not under regular quality control

Priming: HMDS
Rehydration after SB: none
Exposure: Single exposure
Degassing after exposure: none
Development: 5 x 60 s multi puddle

New resists version in 2023
As of 2023-03-21 we no longer have any of the old versions of the resists: AZ 5214E, AZ 4562, AZ MiR 701. This table contains only information about the new resist versions.

Dehydration reducing measures used for testing AZ MiR 701:
The CDA used for the pneumatic autofocus will dehydrate the resist. To reduce this effect, the writehead is parked far away from the write area while setting up the job for at least a few minutes, before starting the exposure.