Jump to content

Specific Process Knowledge/Lithography/Resist/UVresist/exposureDoseMaskAligners

From LabAdviser

Exposure dose for mask aligners

The exposure doses listed below are for generic good exposure results, and can be a compromise between getting good lines, as well as good dots, in both clear field and dark field exposures. The optimal dose for any given specific project, could be different from the listed values.

All doses are for standard silicon wafers, unless otherwise stated. Development is done using 2.38% TMAH (AZ 726 MIF).


Aligner: MA6-1

The Aligner: MA6-1 has an i-line notch filter installed. This produces an exposure light peak around 365 nm with a FWHM of 7 nm.

Quality control
Please note that the values listed in the table are not routinely tested as part of quality control.

Date Operator Film thickness Exposure mode Dose Resolution Comments
AZ MIR 701 Long ago ? 1.5 µm ? 180 mJ/cm2 ? µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ MIR 701 Long ago ? 2.0 µm ? 200 mJ/cm2 ? µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ MIR 701 Long ago ? 4.0 µm ? 400 mJ/cm2 ? µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ 5214E Long ago ? 1.5 µm ? 72 mJ/cm2 ? µm Not under regular quality control

Development: 60 s puddle
AZ 5214E Long ago ? 2.2 µm ? 80 mJ/cm2 ? µm Not under regular quality control

Development: 60 s puddle
AZ 5214E Long ago ? 4.2 µm ? 160 mJ/cm2 ? µm Not under regular quality control

Development: 60 s puddle
AZ 5214E
Image reversal
Long ago ? 2.2 µm ? ? mJ/cm2 ? µm Not under regular quality control

Reversal bake: 60 s @ 110°C
Flood exposure: 500 mJ/cm2
Development: 60 s puddle
AZ nLOF 2020 Long ago ? 2.0 µm ? 121 mJ/cm2 ? µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle

Side wall angle: ~15°
For smaller angle (~5°): develop 30 seconds instead
AZ 4562 Long ago ? 10 µm ? 510 mJ/cm2 ≤5 µm Not under regular quality control

Priming: HMDS
Rehydration after SB: none
Exposure: Multiple exposures with 15 s pauses is recommended
Degassing after exposure: none
Development: 5 x 60 s multi puddle

New resists version in 2023
As of 2023-03-21 we no longer have any of the old versions of the resists: AZ 5214E, AZ 4562, AZ MiR 701. This table contains only information about the old resist versions.


Aligner: MA6-2

The Aligner: MA6-2 has an i-line notch filter installed. This produces an exposure light peak around 365 nm with a FWHM of 7 nm.

Quality control
Please note that the values listed in the table are not routinely tested as part of quality control.

Date Operator Film thickness Exposure mode Dose Resolution Comments
AZ MIR 701 2023-09-26 taran 1.5 µm Hard contact 150 mJ/cm2 ? µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ 5214E 2025-09-30 jehem 1.5 µm Hard contact 90 mJ/cm2 1.5 µm Not under regular quality control

Development: 60 s puddle
AZ 5214E
Image reversal
2023-01-11 jehem 2.2 µm Hard contact 22 mJ/cm2 2.0 µm Not under regular quality control

Reversal bake: 60 s @ 110°C
Flood exposure: 500 mJ/cm2
Development: 60 s puddle
AZ nLOF 2020 2023-09-26 taran 2.0 µm Hard contact 121 mJ/cm2 ? µm Not under regular quality control

Post exposure bake: 60 s @ 110°C
Development: 60 s puddle
AZ 4562 2021-12-08 jehem 10 µm Soft contact 550 mJ/cm2 ≤5 µm Not under regular quality control

Priming: HMDS
Rehydration after SB: none
Exposure: Single exposure
Degassing after exposure: none
Development: 5 x 60 s multi puddle

New resists version in 2023
As of 2023-03-21 we no longer have any of the old versions of the resists: AZ 5214E, AZ 4562, AZ MiR 701. This table contains only information about the new resist versions.