Specific Process Knowledge/Etch/ICP Metal Etcher/Examples of End point detection
Feedback to this page: click here
Unless otherwise stated, this page is written by DTU Nanolab internal
Information on how to use the optical endpoint system, please take a look here: Specific Process Knowledge/Etch/DryEtchProcessing/OES
Cr etch with end point on SiO2
Done April 2023 by bghe
The above images shows a view from Spectral View during etching Cr with DUV resist as masking material of a 6" wafer with more than to 50% open area. The standard Cr etch was used. The endpoint was most clearly seen with the parameters:
- Cr5208av (Average A for region Cr_5208)
- Cr_Avg_5208
- normClOverCr (ScalePct(CrOVerCl, 10,20,100)
- smCr over Cl (Avg2(normCrOverCl, 50)