Critical Point Dryer

Critical point dryer: positioned in cleanroom D-3.

The critical point dryer is used to dry fragile structures that may be damaged in a normal drying procedure. Fragile structures can be thin membranes, or free hanging structures like cantilevers and grippers.


The user manual(s), technical information and contact information can be found in LabManager:

Critical Point Dryer in LabManager


In the critical point dryer, samples are dried in supercritical CO. In the drying process, the sample is first put into Isopropanol for at least 1 hour, and then loaded into the machine. When the machine is started, the isopropanol is exchanged for liquid CO. When there is only CO in the machine, the pressure and temperature in the chamber is raised, above the so called “critical point”. At this point, there is no longer a liquid/gas interfacen, but instead the CO is in a supercritical state. Because of this, the problem with surface tension, which destroys the fragile structures through capillary forces, is avoided.

Equipment performance and process related parameters

Purpose Drying of wafers or chips

To dry fragile structures (example membranes, grippers, cantilevers).

Process parameter range Process Temperature 0 oC to 45 oC
Process pressure

1 atm to (maximum) 95 atm (1400 PSI)

Substrates Batch size
  • 1 to 5 wafers per run. Sizes: 2”, 4" or 6" or
  • Pieces (up to 10x10mm)
Materials allowed

Comparison of samples dried in air and with Critical Point Dryer

This shows a comparison of samples that have been dried in the critical point dryer, and samples that have been dried in the air. Pictures below are shown 2 samples that have been prepared in the same way- but dried with different methods after the final etching step. Cantilevers of SiN, up to 710 µm long and 100 µm wide, are fabricated in the etching step. The thickness of the cantilevers is about 0.5 µm


After drying wafers in the air, the majority of the cantilevers are bended and sticking to the underlying surface. But after the wafers were dried in Critical point dryer, the cantilevers were free hanging.


Dried in air: most cantilevers are bended.


Dried in air: most cantilevers are bended.












Dried with Critical Point Dryer.


Dried with Critical Point Dryer. The cantilevers are free hanging, and have not bended. The dimensions of the cantilevers are 100 x 710 µm, and the thickness is about 0.5 µm.












Dried with Critical Point Dryer.

Dried with the Critical Point Dryer. The cantilevers structures are free hanging after the drying.









An acknowledgment goes to Tom Larsen, Nanoprobes, DTU Nanotech, who provided the pictures.