Specific Process Knowledge/Thin film deposition/ALD2 (PEALD)/TiN deposition using ALD2
Feedback to this page: click here
General information
As already mentioned in the general ALD2 (PEALD) page, oxides and nitrides cannot be deposited at the same time. Since deposition of aluminum nitride requires the plasma source, and since the highest quality titanium nitride is deposited using plasma, then it has been decided to run all nitride deposition processes without the thermal lid. Hence the chamber volume is bigger than with the thermal lid which causes that the substrate temperature is around 50 °C lower than the setpoint.
Result from acceptance test
Deposition of thermal TiN using TMA and NH3 precursors
Recipe name: "TiN thermal with lid" or "TiN thermal without lid".
TiCl4 | NH3 | |
---|---|---|
Nitrogen flow | 100 sccm | 100 sccm |
Pulse time | 0.1 s | 2.0 s |
Purge time | 2.0 s | 6.0 s |
DTU Nanolab has done a deposition with 2000 cycles at 450 °C, where the average growth rate was measured to be 0.0173 nm/cycle.
A number of depositions at 500 °C has been made by Evgeniy Shkondin, DTU Photonics. The number of cycles has been changed, and thickness as function of the number of cycles has been plotted as shown in the figure below.
Unlike other ALD materials, the thickness of a deposited TiN layer cannot be measured using the ellipsometer. Hence one has to cleave the wafer and measure the thickness in the SEM. Thin layers (less than 20 nm) are difficult to focus on due to charging and vibrations, so they may not be measured properly.
The roughness of TiN deposited at 45 °C has been measured to be 0.7415 nm as shown in the figure below.
The resistance of the TiN is depends strongly on the thickness. For very thin layers, a high resistance is measured, and for thicker layers the resistance decreases. The resistance of TiN deposited on glass at 500 °C as function of the thickness can be seen in the figure below (Evgeniy Shkondin, DTU Photonics, May 2017).
TiN with NH3 on trenches
The test was done with 4000 cycles at 450 °C where the average growth rate was measured to be 0.0232 nm/cycle.
TiCl4 | NH3 | |
---|---|---|
Nitrogen flow | 100 sccm | 100 sccm |
Pulse time | 0.5 s | 5.0 s |
Purge time | 2.0 s | 6.0 s |
The SEM images below show a homogeneous coverage all over the trenches.
Pernille Voss Larsen, Mikkel Dysseholm Mar and Tanja Amport, DTU Danchip, 2016-2017.