Specific Process Knowledge/Characterization/Profiler

From LabAdviser

Dektak 8 stylus profiler

Dektak8: positioned in cleanroom 4 - glass cage no. 3

Dektak 8 stylus profiler is a product of Veeco Instruments. It is usable for profiling surfaces of samples with structures in the micrometer range. It can measure in a specific point found with help from high magnification video cameras or it can be programmed to measure several points defined with respect to some deskew points. It can also be used for stress measurements of thin films by measuring the wafer bow.

To get some product information and some tips and tricks from the vendor take a look at Veeco's homepage [1]

A rough overview of the performance of Dektak 8 stylus profiler

Purpose Profiler for measuring micro structures.
  • Single point profiles
  • Wafer mapping
  • Stress measurements by measuring wafer bow
  • Surface roughness on a line scan
Performance Scan range xy

Line scan x: 50 µm to 200 mm

Scan range z

50 Å to 262 µm

Resolution xy

down to 0.067 µm

Resolution z

1Å, 10Å or 20Å

Max. scan depth as a function of trench width W

1.2(W[µm]-5µm)

Hardware settings Tip radius
  • 5 µm 45o cone
Substrates Substrate size
  • up to 8"
Substrate material allowed
  • In principle all materials

Tencor stylus profiler

Tencor: positioned in cleanroom 4 - glass cage no. 3

Tencor P1 stylus profiler is a product of KLA-Tencor Corporation. It is usable for profiling surfaces of samples with structures in the micrometer range. It can measure in a specific point found with help from a high magnification video camera. It can also be used for stress measurements of thin films by measuring the wafer bow.

It is a very old system and should be treated with care.

A rough overview of the performance of Tencor P1 stylus profiler

Purpose Profiler for measuring micro structures.
  • Single point profiles
  • Stress measurements by measuring wafer bow
  • Surface roughness on a line scan
Performance Scan range xy

Line scan x: Full substrate size

. Scan range z

<100 Å to ~0.3 mm

. Resolution xy

up to 5900 data points per profile

. Resolution z

1Å or 25Å

. Max. scan depth as a function of trench width W

0.87(W[µm]-5µm)

Hardware settings Tip radius
  • 5 µm 60o cone
Substrates Substrate size
  • up to 4"
. Substrate material allowed
  • In principle all materials



Comparing the profilers

Take a look at the topographic measurement page.