Specific Process Knowledge/Etch/DRIE-Pegasus/Notation
How to read the notation on process recipes
Look at two examples from the etch cycle of Step1 of the Process A shown in bold below:
- The Platen power has the setting 120 >> 140 (1.5s) 45 - it is to be interpreted as:
- In the first 1.5 seconds of the every cycle the platen power has a value that is ramped (indicated by >>) from 120 W initially to 150 W in the end
- During the remainder of the cycle the platen power is kept constant at 45 W.
- The Pressure has the setting 25 (1.5 s) 90 >> 150 - it is to be interpreted as:
- In the first 1.5 seconds the pressure is constant at 25 mtorr.
- During the remainder of the cycle the pressure has i higher value that is ramped from 90 initially to 150 mtorr in the last etch cycle.
Step 1 | Step 2 | |||
Parameter | Etch | Dep | Etch | Dep |
Gas flow (sccm) | SF6 350 (1.5 s) 550 | C4F8 200 | SF6 350 (1.5 s) 550 | C4F8 200 |
Cycle time (secs) | 7.0 | 4.0 | 7.0 | 4.0 |
Pressure (mtorr) | 25 (1.5 s) 90 >> 150 | 25 | 25 (1.5 s) 150 | 25 |
Coil power (W) | 2800 | 2000 | 2800 | 2000 |
Platen power (W) | 120 >> 140 (1.5) 45 | 0 | 140 (1.5) 45 | 0 |
Cycles | 11 (keep fixed) | 44 (vary this) | ||
Common | Temperature 20 degs, HBC 10 torr, Short funnel, with baffle & 5mm spacers |