Specific Process Knowledge/Characterization/XRD

From LabAdviser

Feedback to this page: click here

XRD SmartLab

The Rigaku SmartLab is an advanced XRD for measuring on thin films. All thin films can be measured without fixating the sample, as the system has a so called In-Plane arm.

The XRD SmartLab located in cleanroom F-2


The user manual(s), user APV(s), technical information, and contact information can be found in LabManager:

XRD SmartLab in LabManager

Process information

XRR

With X-Ray Reflectivity measurements, it is possible to obtain information on thickness, density, and both surface and interface roughness on thin films. The technique does not depend on crystal structure and can be used on both amorphous, poly-, and single-crystalline materials. For film thickness measurement, films op to around 100 nm can be measured. You are welcome to try it on thicker films, but please confirm the measurement the first time by use of other equipment. XRR is a special case of a Theta/2Theta measurement. Rigaku gives a good explanation of the principles behind the XRR in this paper.:File:X-ray thin film measurements techniques V X-ray reflectivity measurements.pdf

Software for analysis

The software packages used for data analysis are available on the equipment computer, but we recommend that you install it on your personal computer. To run the software you need a USB dongle with a license on, these can be borrowed from Rebecca and Kristian in room 347-077. We only have 9 dongles available, so when you are done please return the dongle to danchip.

The software can be found on "CleanroomDrive\_Equipment\XRD\Rigaku software\RILauncher", it should be possible to install the software without a dongle. To use the software you have to log in. The user is: Administrator. There is no password.

Equipment performance and process related parameters

Equipment XRD SmartLab
Purpose Crystal structure analysis and thin film thickness measurement
  • Phase ID
  • Crystal Size
  • Crystallinity
  • Quality and degree of orientation
  • 3D orientation
  • Latice strain
  • Composition
  • Twist
  • 3D lattice constant
  • Thickness
  • Roughness
  • Density
X-ray generator

Maximum rated output

3 kW

Rated tube voltage

20 to 45 kV

Rated tube current

2 to 60 mA

Type

Sealed tube

Target

Cu

Focus size

0.4x8 mm (Line/Point)

Goniometer

Scanning mode

incident / receiver coupled or independent

Goniomenter radius

300 mm

Minimum step size

0.0001° (0.36")

Sample stage

  • χ:-5~+95°
  • φ:0~360°
  • Z:-4~+1 mm
  • X,Y:±50 mm for a 100 mm wafer
  • Rx,Ry:-5~+5°

Sample size

Diameter: 150 mm Thickness: 0~21 mm

Optics Incident side
  • Cross Beam Optics(CBO)
  • Ge(220)x2 monochromator
  • In-Plane Parallel Slit Collimator (PSC)
  • Soller slit
  • Variable divergence slit
Receiver side
  • Automatic variable scattering slit
  • Automatic variable receiver slit
  • Parallel slit analysers (PSA)
  • Ge(220)x2 analyser
Substrates Substrate size

up to 150 mm wafers

Allowed materials

All materials