Specific Process Knowledge/Etch/ICP Metal Etcher/Chromium
Feedback to this page: click here
Chromium etch in ICP metal
The Chromium etch was carried out on the following substrate stack: 2" Si wafer with Cr laying in a 6" Si wafer with a 4" recess. The area outside the recess was covered by AZ resist. The work was carried out be Erol Zekovic @Nanotech and BGHE@danchip
Parameter | Cr etch |
---|---|
Cl2 (sccm) | 65 |
O2 (sccm) | 15 |
Pressure (mTorr) | 15 |
Coil power (W) | 300 |
Platen power (W) | 15 |
Temperature (oC) | 50 |
Spacers (mm) | 100 |
Etch rate (nm/min) | ~32 (Date: 2014-08-13) |
Zep520A resist selectivity | NA |
Comment | Was masked by capton tape |
Chromium etch in ICP metal
The Chromium etch has ONLY been carried out on the following substrate stack: The Chromium is sputter deposited onto a 2" quartz wafer and pattered by e-beam with Zep520A resist. This 2" QZ wafer is bonded with crystal bond to a 65mmx65mm quartz plate with the thickness: 6.35mm. This QZ plate is bonded to a Si wafer.
Parameter | Cr etch |
---|---|
Cl2 (sccm) | 65 |
O2 (sccm) | 15 |
Pressure (mTorr) | 15 |
Coil power (W) | 300 |
Platen power (W) | 15 |
Temperature (oC) | 50 (no back side cooling) |
Spacers (mm) | 100 |
Etch rate (nm/min) | ~14 |
Zep520A resist selectivity | ~0.9 |
Comment | . |