Specific Process Knowledge/Etch/OES
Optical endpoint detection on the dry etch tools at DTU Nanolab
Several dry etch tools at DTU Nanolab are equipped with an endpoint detection system. Out of those systems only one is not of the type optical endpoint detection. The instruments are:
- ICP Metal Etch
- III-V ICP
- Pegasus 1
- Pegasus 4
The section below describes the principle behind the optical endpoint detection system.
Optical Emission Spectroscopy
As an example, let's take the etching of silicon by fluorine in one of the dry etchers. The fluorine is supplied to the system as SF6 that is decomposed by the plasma into smaller and smaller molecules until the radical F* is formed.
fdgsdfg
Monitored species | Wavelength (nm) | Monitored species | Wavelength (nm) | |
Al | 308.2, 309.3, 396.1 | In | 325.6 | |
AlCl | 261.4 | N | 674.0 | |
As | 235.0 | N2 | 315.9, 337.1 | |
C2 | 516.5 | NO | 247.9, 288.5, 289.3, 303.5, 304.3, 319.8, 320.7, 337.7, 338.6 | |
CF2 | 251.9 | O | 777.2, 844.7 | |
Cl | 741.4 | OH | 281.1, 306.4, 308.9 | |
CN | 289.8, 304.2, 387.0 | S | 469.5 | |
CO | 292.5, 302.8, 313.8, 325.3, 482.5, 483.5, 519.8 | Si | 288.2 | |
F | 703.7, 712.8 | SiCl | 287.1 | |
Ga | 417.2 | SiF | 440.1, 777.0 | |
H | 486.1, 656.5 |