Specific Process Knowledge/Wafer cleaning/RCA

From LabAdviser

RCA cleaning

The RCA clean is used for cleaning the wafers before taking them into the furnaces and a few other equipments (check the cross contamination sheet). It consist of two solutions: RCA1 and RCA2 plus diluted HF.

The RCA1 contains: HO, NHOH and HO (5:1:1). It is used for removed of light organics, particles and metal.
The RCA2 contains: HO, HCl and HO (5:1:1). It is used for removal of heavy metals, alkalies and metal hydroxides.

RCA procedure

  • RCA1: 10 min
  • DI wafer rinsing (dumping three times)
  • HF: 30 sec (avoid it if you have oxide as the top layer)
  • DI wafer rinsing (dumping three times)
  • RCA2: 10 min
  • DI wafer rinsing (dumping three times)
  • HF: 30 sec (avoid it if you have oxide as the top layer)
  • DI wafer rinsing (dumping three times)


Overview of RCA process data

RCA1 RCA2 HF
General description

It is used for removal of light organics, particles and metal.

It is used for removal of light organics, particles and metal.

It is used for removal of oxide generated in RCA1 and RCA2

Chemical solution HO, NHOH(29%) and HO(30%) (5:1:1) HO, HCl(37%) and HO(30%) (5:1:1) 5% HF
Process temperature 70-80 oC 70-80 oC Room temperature
Process time

10 min.

10 min.

30 sec.

Life time of the chemical solutions Can only be heated one time. When hot: it lasts for ~1½h Can only be heated one time. When hot: it lasts for ~1½h ~2 months
Allowed materials
  • Silicon
  • Silicon oxide
  • Silicon nitride
  • Quartz wafers
  • Silicon
  • Silicon oxide
  • Silicon nitride
  • Quartz wafers
  • Silicon
  • Silicon oxide
  • Silicon nitride
Batch size

1-25 4" or 6" wafers at a time

1-25 4" or 6" wafers at a time

1-25 4" or 6" wafers at a time

Size of substrate

4"-6" wafers

4"-6" wafers

4"-6" wafers